Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling

被引:64
作者
Clasen, S
Schmidt, D
Boss, A
Dietz, K
Kröber, SM
Claussen, CD
Pereira, PL
机构
[1] Univ Tubingen, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Univ Tubingen, Dept Med Biometry, Tubingen, Germany
[3] Univ Tubingen, Inst Pathol, D-7400 Tubingen, Germany
关键词
D O I
10.1148/radiol.2382050571
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the size and geometry of thermally induced coagulation by Using multipolar radiofrequency (RF) ablation and to determine a mathematic model to predict coagulation volume. Materials and Methods: Multipolar RF ablations (n = 80) were pet-formed in ex vivo bovine livers by using three internally cooled bipolar same shaft. Applicators were placed in a trianglilar array (spacing, 2-5 cm) and were activated in multipolar mode (power output, 75-225 W). The size and geometry of the coagulation zone, together with ablation time, were assessed. Mathematic functions were fitted. and the goodness or fit was assessed by using r(2). Results: Coagulation volome, short-axis diameter, and ablation time were dependent on power output an([ applicator distance. The maximum zone of coagulation (volume, 324 cm(3); short-axis diameter. 8.4 cm: ablation time, 193 min) was induced with a power output or 75 W at an applicator distance of 5 cm. Coagulation volume and ablation time decreased as power output increased. Power outputs Of 100-125 W at applicator distances of 2-4 cm led to a reasonable compromise between coagulation volume and ablation time. At 2 cm (100 W), coagulation volume, 3 short-axis diameter, and ablation time were 66 cm, 4.5 cm, and 19 min, respectively; at 3 cm (100 W), go cm(3),13 5.2 3 cm, and 22 min, respectively; at 4 cm (100 W) 132 cm 6.1 cm, and 27 min, respectively; at 2 cm (125 W), 56 cm 3 4.2 cm. and 9 min, respectively: at 3 cot (125 W), 73 cm 4.9 cm, and 12 min. respectively; and at. 4 cm (125 W) 103 cm(3), 5.5 cm, and 16 min, respectively. At applicator distances of 4 ern (>125 W) and 5 cm (>100 W), the zones of coagulation were not confluent. Coagulation on volume (r(2) = 0.80) and RF ablation time (r(2) = 0.93) were determined by using the mathematic model. Conclusions: Multipolar RF ablation with three bipolar applicators may produce large volumes of confluent coagulation ex vivo. A compromise is necessary between prolonged RF ablations at lower power outputs, which produce larger volumes of coagulation. and faster RF ablations at higher power outputs, which produce smaller volumes of coagulation. (c) RSNA, 2006.
引用
收藏
页码:881 / 890
页数:10
相关论文
共 31 条
  • [1] Bipolar saline-enhanced electrode for radiofrequency ablation:: Results of experimental study of in vivo porcine liver
    Burdío, F
    Güemes, A
    Burdío, JM
    Navarro, A
    Sousa, R
    Castiella, T
    Cruz, I
    Burzaco, O
    Lozano, R
    [J]. RADIOLOGY, 2003, 229 (02) : 447 - 456
  • [2] Large liver tumors: Protocol for radiofrequency ablation and its clinical application in 110 patients - Mathematic model, overlapping mode, and electrode placement process
    Chen, MH
    Yang, W
    Yan, K
    Zou, MW
    Solbiati, L
    Liu, LB
    Dai, Y
    [J]. RADIOLOGY, 2004, 232 (01) : 260 - 271
  • [3] Radiofrequency liver ablation: Experimental comparative study of water-cooled versus expandable systems
    de Baere, T
    Denys, A
    Wood, BJ
    Lassau, N
    Kardache, M
    Vilgrain, V
    Menu, Y
    Roche, A
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (01) : 187 - 192
  • [4] Desinger K, 1996, MINIM INVASIVE MED, V7, P92
  • [5] Radiofrequency thermal ablation: Computer analysis of the size of the thermal injury created by overlapping ablations
    Dodd, GD
    Frank, MS
    Aribandi, M
    Chopra, S
    Chintapalli, KN
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (04) : 777 - 782
  • [6] Tumor ablation with radio-frequency energy
    Gazelle, GS
    Goldberg, SN
    Solbiati, L
    Livraghi, T
    [J]. RADIOLOGY, 2000, 217 (03) : 633 - 646
  • [7] Percutaneous radiofrequency tissue ablation: Optimization of pulsed radiofrequency technique to increase coagulation necrosis
    Goldberg, SN
    Stein, MC
    Gazelle, GS
    Sheiman, RG
    Kruskal, JB
    Clouse, ME
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (07) : 907 - 916
  • [8] GOLDBERG SN, 1995, ACAD RADIOL, V2, P670
  • [9] Large-volume tissue ablation with radio frequency by using a clustered, internally cooled electrode technique: Laboratory and clinical experience in liver metastases
    Goldberg, SN
    Solbiati, L
    Hahn, PF
    Cosman, E
    Conrad, JE
    Fogle, R
    Gazelle, GS
    [J]. RADIOLOGY, 1998, 209 (02) : 371 - 379
  • [10] Image-Guided Tumor Ablation: Standardization of terminology and reporting criteria
    Goldberg, SN
    Grassi, CJ
    Cardella, JF
    Charboneau, JW
    Dodd, GD
    Dupuy, DE
    Gervais, D
    Gillams, AR
    Kane, RA
    Lee, FT
    Livraghi, T
    McGahan, J
    Phillips, DA
    Rhim, H
    Silverman, SG