Late type III endoleak from fabric tears of a zenith stent graft: report of a case

被引:16
作者
Banno, Hiroshi [1 ]
Morimae, Hirofumi [1 ]
Ihara, Tsutomu [1 ]
Kobayashi, Masayoshi [1 ]
Yamamoto, Kiyohito [1 ]
Komori, Kimihiro [1 ]
机构
[1] Nagoya Univ, Grad Sch Med, Div Vasc Surg, Dept Surg,Showa Ku, Nagoya, Aichi 4668550, Japan
关键词
Endoleak; Fabric tears; Stent graft; EVAR; AORTIC-ANEURYSM; MANAGEMENT; DIAGNOSIS; REPAIR; CLASSIFICATION;
D O I
10.1007/s00595-012-0320-8
中图分类号
R61 [外科手术学];
学科分类号
摘要
We report a case of late type III B endoleak from a fabric tear in the main body of a Zenith bifurcated stent-graft approximately 19 months after implantation. A follow-up computed tomography (CT) scan showed gradual growth in the size of the aneurysm, but no apparent endoleak was recognized. However, the CT scan had demonstrated a mosaic pattern in the aneurysm sac, which could indicate an endoleak from a fabric tear. The defect was repaired by implantation of body extensions, but the patient died 5 h after the procedure.
引用
收藏
页码:1206 / 1209
页数:4
相关论文
共 15 条
[1]   Late type III endoleak and graft failure of an Ancure stent-graft [J].
Brown, Katherine E. ;
Heyer, Kamaldeep S. ;
Matsumura, Jon S. ;
Eskandari, Mark K. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (10) :1506-1508
[2]  
Cao P, 2010, J CARDIOVASC SURG, V51, P53
[3]   Successful treatment of an abdominal aortic aneurysm by endovascular graft placement through a previously placed prosthetic graft: Report of a case [J].
Hori, Daijiro ;
Yuri, Koichi ;
Nemoto, Kazunari ;
Yamaguchi, Atsushi ;
Adachi, Hideo .
SURGERY TODAY, 2010, 40 (12) :1169-1172
[4]   Diagnosis of Type III Endoleak and Endovascular Treatment with Aortouniiliac Stent-Graft [J].
Juszkat, Robert ;
Staniszewski, Ryszard ;
Zarzecka, Anna ;
Majewski, Waclaw .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (01) :125-129
[5]   Late type III endoleak from graft erosion of an Excluder stent graft: A case report [J].
Lee, W. Anthony ;
Huber, Thomas S. ;
Seeger, James M. .
JOURNAL OF VASCULAR SURGERY, 2006, 44 (01) :183-185
[6]  
Parodi J C, 1991, Ann Vasc Surg, V5, P491, DOI 10.1007/BF02015271
[7]   Imaging techniques for detection and management of endoleaks after endovascular aortic aneurysm repair [J].
Stavropoulos, S. William ;
Charagundla, Sridhar R. .
RADIOLOGY, 2007, 243 (03) :641-655
[8]   Fabric tears as a new cause of type III endoleak with Ancure endograft [J].
Teutelink, A ;
van der Laan, MJ ;
Milner, R ;
Blankensteijn, JD .
JOURNAL OF VASCULAR SURGERY, 2003, 38 (04) :843-846
[9]   First report of a late type III endoleak from fabric tears of a Zenith stent graft [J].
Wanhainen, Anders ;
Nyman, Rickard ;
Eriksson, Mats-Ola ;
Bjorck, Martin .
JOURNAL OF VASCULAR SURGERY, 2008, 48 (03) :723-726
[10]  