Parallel Validation of Three Molecular Devices for Simultaneous Detection and Identification of Influenza A and B and Respiratory Syncytial Viruses

被引:48
|
作者
Ling, Lifen [1 ,2 ]
Kaplan, Samuel E. [2 ]
Lopez, Juan C. [2 ]
Stiles, Jeffrey [2 ]
Lu, Xuedong [1 ]
Tang, Yi-Wei [2 ,3 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 8, Dept Lab Med, Shenzhen, Peoples R China
[2] Mem Sloan Kettering Canc Ctr, Dept Lab Med, 1275 York Ave, New York, NY 10021 USA
[3] Cornell Univ, Weill Med Coll, Dept Pathol & Lab Med, New York, NY 10021 USA
关键词
influenza viruses A and B; respiratory syncytial virus; molecular testing; hands-on time; turnaround time; PERFORMANCE EVALUATION; DIAGNOSTIC-ACCURACY; CLINICAL-EVALUATION; MULTIPLEX PCR; VIRAL PANEL; RSV ASSAY; FLU A/B; INFECTION; TIME; MANAGEMENT;
D O I
10.1128/JCM.01691-17
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Rapid identification of respiratory pathogens, such as influenza virus A (FluA), influenza virus B (FluB), and respiratory syncytial virus (RSV), reduces unnecessary antimicrobial use and enhances infection control practice. We performed a comparative evaluation of three molecular methods: (i) the Aries Flu A/B & RSV, (ii) the Xpert Xpress Flu/RSV, and (iii) the Cobas Flu A/B & RSV assays. The clinical performances of the three methods were evaluated using 200 remnant nasopharyngeal swab (NPS) specimens against a combined reference standard. The limits of detection (LODs) were determined using FluA, FluB, and RSV control strains with known titers. The 95% LODs were between 1.702 and 0.0003 50% tissue culture infective dose (TCID50), with no significant differences revealed among the three assays. Perfect qualitative detection agreement was obtained in the reproducibility study. The Cobas assay failed at the first run on 13 clinical specimens, resulting in an invalid rate of 6.5%. The sensitivities and specificities for all assays were 96.0 to 100.0% and 99.3 to 100% for all three viruses. For on-demand single-specimen and batched 12-specimen workflows, the test turnaround times were 115.5 and 128.8 min for the Aries assay (12 sample capacity), 34.2 and 44.2 min for the Xpress assay (16 sample capacity), and 21.0 and 254.4 min for the Cobas assay (one instrument), respectively. In summary, the Aries, Xpress, and Cobas Liat assays demonstrated excellent sensitivities and specificities for simultaneous detection and identification of FluA, FluB, and RSV from NPS specimens in cancer patients. Test turnaround time was significantly shorter on the Xpress when instrument scalability is unlimited.
引用
收藏
页数:6
相关论文
共 44 条
  • [1] Performance evaluation of a rapid molecular diagnostic, MultiCode based, sample-to-answer assay for the simultaneous detection of Influenza A, B and respiratory syncytial viruses
    Voermans, Jolanda J. C.
    Seven-Deniz, S.
    Fraaij, P. L. A.
    van der Eijk, Annemiek A.
    Koopmans, M. P. G.
    Pas, Suzan D.
    JOURNAL OF CLINICAL VIROLOGY, 2016, 85 : 65 - 70
  • [2] Semiquantitative one-step RT-PCR for simultaneous identification of human influenza and respiratory syncytial viruses
    Chi, X. Sherry
    Li, Fenglan
    Tam, John S.
    Rappaport, Ruth
    Cheng, Sheau-Mei
    JOURNAL OF VIROLOGICAL METHODS, 2007, 139 (01) : 90 - 92
  • [3] Detection of Influenza A and B Viruses and Respiratory Syncytial Virus by Use of Clinical Laboratory Improvement Amendments of 1988 (CLIA)-Waived Point-of-Care Assays: a Paradigm Shift to Molecular Tests
    Azar, Marwan M.
    Landry, Marie L.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2018, 56 (07)
  • [4] Performance of a Novel Point-of-Care Molecular Assay for Detection of Influenza A and B Viruses and Respiratory Syncytial Virus (Enigma MiniLab) in Children with Acute Respiratory Infection
    Douthwaite, Sam T.
    Walker, Charlotte
    Adams, Elisabeth J.
    Mak, Catherine
    Ortiz, Andres Vecino
    Martinez-Alier, Nuria
    Goldenberg, Simon D.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2016, 54 (01) : 212 - 215
  • [5] Diagnostic Accuracy of Xpert Xpress Flu/RSV for the Detection of Influenza and Respiratory Syncytial Viruses
    Liu, Ye-Ling
    Xie, Tian-Ao
    Lin, Geng-Ling
    Deng, Wei
    Lin, Qin-Rong
    Pan, Zhi-Yong
    Fan, Shu-Jin
    Li, Zhen-Xing
    Ouyang, Shi
    Zhu, Guo-Dong
    Ji, Tian-Xing
    Wu, Li-Juan
    Xia, Yong
    Guo, Xu-Guang
    JAPANESE JOURNAL OF INFECTIOUS DISEASES, 2022, 75 (02) : 183 - 191
  • [6] Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
    Otto, Caitlin C.
    Kaplan, Samuel E.
    Stiles, Jeffrey
    Mikhlina, Albina
    Lee, Cindy
    Babady, N. Esther
    Tang, Yi-Wei
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2017, (119):
  • [7] The Enigma ML FluAB-RSV assay: a fully automated molecular test for the rapid detection of influenza A, B and respiratory syncytial viruses in respiratory specimens
    Goldenberg, Simon D.
    Edgeworth, Jonathan D.
    EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2015, 15 (01) : 23 - 32
  • [8] Real-Time Detection of Influenza A, Influenza B, and Respiratory Syncytial Virus A and B in Respiratory Specimens by Use of Nanoparticle Probes
    Jannetto, Paul J.
    Buchan, Blake W.
    Vaughan, Kimberly A.
    Ledford, Joellen S.
    Anderson, Dennis K.
    Henley, Donald C.
    Quigley, Neil B.
    Ledeboer, Nathan A.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (11) : 3997 - 4002
  • [9] Rapid Molecular Tests for Influenza, Respiratory Syncytial Virus, and Other Respiratory Viruses: A Systematic Review of Diagnostic Accuracy and Clinical Impact Studies
    Vos, Laura M.
    Bruning, Andrea H. L.
    Reitsma, Johannes B.
    Schuurman, Rob
    Riezebos-Brilman, Annelies
    Hoepelman, Andy I. M.
    Oosterheert, Jan Jelrik
    CLINICAL INFECTIOUS DISEASES, 2019, 69 (07) : 1243 - 1253
  • [10] Comparative Evaluation of Two Automated Multiplex RT-PCR Tests for Rapid Detection of Influenza and Respiratory Syncytial Viruses
    Hur, Kyu-Hwa
    Sung, Heungsup
    Kim, Mi-Na
    CLINICAL LABORATORY, 2021, 67 (06) : 1472 - 1476