A Combined PET/CT Scanner for Clinical Oncology

被引:0
作者
Beyer, Thomas
Townsend, David W. [1 ]
Brun, Tony
Kinahan, Paul E.
Charron, Martin
Roddy, Raymond
Jerin, Jeff
Young, John
Byars, Larry
Nutt, Ronald
机构
[1] Univ Pittsburgh, PET Facil, Dept Radiol, 200 Lothrop St, Pittsburgh, PA 15213 USA
关键词
oncology; PET; CT; dual modality; image fusion; ATTENUATION CORRECTION; FDG PET; CT; TRANSMISSION; TOMOGRAPHY;
D O I
10.2967/jnumed.106.252841a
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The availability of accurately aligned, whole-body anatomical (CT) and functional (PET) images could have a significant impact on diagnosing and staging malignant disease and on identifying and localizing metastases. Computer algorithms to align CT and PET images acquired on different scanners are generally successful for the brain, whereas image alignment in other regions of the body is more problematic. Methods: A combined PET/CT tomograph with the unique capability of acquiring accurately aligned functional and anatomical images for any part of the human body has been designed and built. The PET/CT scanner was developed as a combination of a Siemens Somatom AR.SP spiral CT and a partial-ring, rotating ECAT ART PET scanner. All components are mounted on a common rotational support within a single gantry. The PET and CT components can be operated either separately, or in combined mode. In combined mode, the CT images are used to correct the PET data for scatter and attenuation. Fully quantitative whole-body images are obtained for an axial extent of 100 cm in an imaging time of less than 1 h. When operated in PET mode alone, transmission scans are acquired with dual Cs-137 sources. Results: The scanner is fully operational and the combined device has been operated successfully in a clinical environment. Over 110 patients have been imaged, covering a range of different cancers, including lung, esophageal, head and neck, melanoma, lymphoma, pancreas, and renal cell. The aligned PET and CT images are used both for diagnosing and staging disease and for evaluating response to therapy. We report the first performance measurements from the scanner and present some illustrative clinical studies acquired in cancer patients. Conclusion: A combined PET and CT scanner is a practical and effective approach to acquiring co-registered anatomical and functional images in a single scanning session.
引用
收藏
页码:168S / 177S
页数:10
相关论文
共 28 条
[1]   ECAT ART - A continuously rotating PET camera: Performance characteristics, initial clinical studies, and installation considerations in a nuclear medicine department [J].
Bailey, DL ;
Young, H ;
Bloomfield, PM ;
Meikle, SR ;
Glass, D ;
Myers, MJ ;
Spinks, TJ ;
Watson, CC ;
Luk, P ;
Peters, AM ;
Jones, T .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1997, 24 (01) :6-15
[2]  
Charron M, 1999, J NUCL MED, V40, p256P
[3]  
Comtat C, 1999, IEEE MED IM
[4]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[5]   ATTENUATION CORRECTION IN PET USING SINGLE-PHOTON TRANSMISSION MEASUREMENT [J].
DEKEMP, RA ;
NAHMIAS, C .
MEDICAL PHYSICS, 1994, 21 (06) :771-778
[6]  
Eubank WB, 1998, AJR, V171, P1101
[7]   A PROTON PROFILE MONITOR USING AN AMORPHOUS-SILICON PHOTODIODE-ARRAY [J].
HASEGAWA, K ;
SUGIURA, H ;
MOCHIKI, K ;
TAKAHASHI, H ;
HAYAKAWA, Y ;
INADA, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :569-572
[8]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[9]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[10]   SINGLES TRANSMISSION IN VOLUME-IMAGING PET WITH A CS-137 SOURCE [J].
KARP, JS ;
MUEHLLEHNER, G ;
QU, H ;
YAN, XH .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :929-944