Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an EasyClick cuff according to the International Protocol in adults

被引:6
|
作者
Iliman, Nevzat
Altunkan, Sekip
Kayaturk, Nur
Altunkan, Erkan
机构
[1] Metropol Med Ctr, Div Hypertens, TR-06400 Ankara, Turkey
[2] Diskapi Yildirim Beyazit Educ & Res Hosp, Dept Internal Med, Ankara, Turkey
关键词
blood pressure measurement; Braun BP 3550; validation; wrist monitors;
D O I
10.1097/01.mbp.0000218006.80827.dc
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Among the electronic blood pressure measuring devices used for self-measurement of blood pressure by patients, models that measure at the wrist are becoming increasingly popular. Validation studies done with these equipments, however, are quite few. The objective of this study was to evaluate the accuracy of the Braun BP 3550 wrist blood pressure device against the mercury sphygmomanometer in adults according to the International Protocol criteria. Methods Fifty-four patients over 30 years of age were studied and classified on the basis of the range of the International Protocol. Blood pressure measurements at the wrist with the Braun BP 3550 were compared with the results obtained by two trained observers using a mercury sphygmomanometer. Nine sequential blood pressure measurements were taken. A total of 33 individuals with randomly distributed arm circumferences were selected for the validation study. During the validation study, 99 measurements were obtained for comparison from 33 individuals. The first phase was performed on 15 individuals and if the device passed this phase, 18 more individuals were selected. Results Mean discrepancies and standard deviations of the device sphygmomanometer were 0.4 +/- 8.1 mmHg for systolic blood pressure and 2.1 +/- 5.8 mmHg for diastolic blood pressure in the study group. The device passed phase 1 in 15 individuals. In phase 2.1, from the total 99 comparisons, 65, 84 and 91 for systolic blood pressure and 62, 91 and 98 for diastolic blood pressure were <5, < 10 and < 15 mmHg, respectively. In phase 2.2, 23 individuals had at least two of the differences within 5 mmHg, and two had no differences within 5 mmHg for systolic blood pressure. In terms of diastolic blood pressure, 23 individuals had at least two of the differences within 5 mmHg and three had no differences within 5 mmHg. Braun BP 3550 passed phases 2.1 and 2.2 in the adult participants. Conclusion The Braun BP 3550 wrist blood pressure device passed according to the International Protocol criteria and can be recommended for use in adults. Deviation from the average value, however, was greater in patients with high systolic blood pressure as compared with those with lower blood pressure.
引用
收藏
页码:45 / 49
页数:5
相关论文
共 50 条
  • [31] Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Natalizi, Anna
    Gerosa, Pietro Francesco
    Frigerioa, Mauro
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2008, 13 (03) : 177 - 182
  • [32] Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
    O'Brien, E
    Pickering, T
    Asmar, R
    Myers, M
    Parati, G
    Staessen, J
    Mengden, T
    Imai, Y
    Waeber, B
    Palatini, P
    Atkins, N
    Gerin, W
    BLOOD PRESSURE MONITORING, 2002, 7 (01) : 3 - 17
  • [33] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults
    O'Brien, Eoin
    Atkins, Neil
    Stergiou, George
    Karpettas, Nikos
    Parati, Gianfranco
    Asmar, Roland
    Imai, Yutaka
    Wang, Jiguang
    Mengden, Thomas
    Shennan, Andrew
    BLOOD PRESSURE MONITORING, 2010, 15 (01) : 23 - 38
  • [34] Validation of the Andon KD-575 automated blood pressure monitor according to the European Society of Hypertension International Protocol in adults
    Pan, Feng
    Wan, Yi
    Chen, Wan
    Li, Ni
    Tong, Wei
    Xu, Yongyong
    BLOOD PRESSURE MONITORING, 2010, 15 (03) : 167 - 170
  • [35] Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 50 - 53
  • [36] Validation of the Parama-Tech PS-501 Device for Office Blood Pressure Measurement According to the International Protocol
    Kikuya, Masahiro
    Ohkubo, Takayoshi
    Satoh, Michihiro
    Hashimoto, Takanao
    Hirose, Takuo
    Metoki, Hirohito
    Obara, Taku
    Inoue, Ryusuke
    Asayama, Kei
    Totsune, Kazuhito
    Imai, Yutaka
    CLINICAL AND EXPERIMENTAL HYPERTENSION, 2012, 34 (01) : 71 - 73
  • [37] STRIDE BP international initiative for accurate blood pressure measurement: Systematic review of published validation studies of blood pressure measuring devices
    Stergiou, George S.
    O'Brien, Eoin
    Myers, Martin
    Palatini, Paolo
    Parati, Gianfranco
    Kollias, Anastasios
    Birmpas, Dimos
    Kyriakoulis, Konstantinos
    Bountzona, Ioanna
    Stambolliu, Emelina
    Anagnostopoulos, Ioannis
    Karpettas, Nikos
    Menti, Ariadni
    JOURNAL OF CLINICAL HYPERTENSION, 2019, 21 (11): : 1616 - 1622
  • [38] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [39] Validation of the blood pressure measurement device Erkameter 125 PRO according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Zimmermann, Erik
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2016, 21 (06) : 356 - 360
  • [40] Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol
    Ragazzo, Fabio
    Saladini, Francesca
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2010, 15 (01) : 59 - 62