Optimization of the mechanical collimation for minibeam generation in proton minibeam radiation therapy

被引:35
作者
Guardiola, Consuelo
Peucelle, Cecile
Prezado, Yolanda [1 ]
机构
[1] CNRS, IMNC, UMR 8165, Paris 7, 15 Rue Georges Clemenceau, F-91405 Orsay, France
关键词
Monte Carlo simulations; proton minibeam radiation therapy; spatial fractionation of the dose; GLIOMA-BEARING RATS; MICROBEAM; SIMULATION; BRAIN; RADIOTHERAPY; BEAM; GLIOSARCOMA; IRRADIATION;
D O I
10.1002/mp.12131
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The dose tolerances of normal tissues continue to be the main barrier in radiation therapy. To lower it, a novel concept based on a combination of proton therapy and the use of arrays of parallel and thin beams has been recently proposed: proton minibeam radiation therapy (pMBRT). It allies the inherent advantages of protons with the remarkable normal tissue preservation observed when irradiated with submillimetric spatially fractionated beams. Due to multiple Coulomb scattering, the tumor receives a homogeneous dose distribution, while normal tissues in the beam path benefit from the spatial fractionation of the dose. This promising technique has already been implemented at a clinical center (Proton therapy Center of Orsay) by means of a first prototype of a multislit collimator. The main goal of this work was to optimize the minibeam generation by means of a mechanical collimation. Methods: Monte Carlo simulations (GATE V7.1) were used to evaluate the influence of the collimator material (brass, nickel, iron, tungsten), thickness, phantom-to-collimator distance (PCD), among other parameters, on the dose distributions. Maximization of the peak-to-valley dose ratios (PVDR) in normal tissues along with minimization of full width at half maximum, penumbras and neutron contamination were used as figures of merit. As a starting point for the optimization, the collimator employed in our previous works was used. It consisted in 400 m x 2 cm slits with a center-to-center distance (c-t-c) of 3200 m. As the main targets of pMBRT will be neurological cases, 100 MeV energy proton minibeams were considered. This energy range would allow treating tumors located at the center of the brain (the worst scenario). Results: Tungsten and brass are the most advantageous materials among those considered. A tungsten collimator provides the highest PVDR and lowest penumbra. Although the neutron yield generated in the tungsten collimator is 3 times higher than that of the other materials, the biologic neutron doses at the patient position amount to less than 0.05% and 0.7% of the peak and valley doses, respectively. In addition, shorter PCD than the one currently used (7 cm) leads to thinner beams (enhancing the dose-volume effects), accompanied, however, by an increase of neutron dose at the phantom surface. Finally, no gain in dose distributions is obtained by using nonparallel slits. Conclusions: The collimator design and irradiation configuration have been optimized to minimize the angular spread, deliver the highest PVDR and the lowest valley possible in the normal tissues in pMBRT. We have also confirmed that even though the neutron yield generated in the multislit collimator is higher with respect to the one produced by the collimators used in conventional proton therapy, the increase of biological neutron dose in the patient will remain low (less than 1%). (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:1470 / 1478
页数:9
相关论文
共 53 条
  • [1] GEANT4-a simulation toolkit
    Agostinelli, S
    Allison, J
    Amako, K
    Apostolakis, J
    Araujo, H
    Arce, P
    Asai, M
    Axen, D
    Banerjee, S
    Barrand, G
    Behner, F
    Bellagamba, L
    Boudreau, J
    Broglia, L
    Brunengo, A
    Burkhardt, H
    Chauvie, S
    Chuma, J
    Chytracek, R
    Cooperman, G
    Cosmo, G
    Degtyarenko, P
    Dell'Acqua, A
    Depaola, G
    Dietrich, D
    Enami, R
    Feliciello, A
    Ferguson, C
    Fesefeldt, H
    Folger, G
    Foppiano, F
    Forti, A
    Garelli, S
    Giani, S
    Giannitrapani, R
    Gibin, D
    Cadenas, JJG
    González, I
    Abril, GG
    Greeniaus, G
    Greiner, W
    Grichine, V
    Grossheim, A
    Guatelli, S
    Gumplinger, P
    Hamatsu, R
    Hashimoto, K
    Hasui, H
    Heikkinen, A
    Howard, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) : 250 - 303
  • [2] Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film
    Amanda Garcia-Garduno, Olivia
    Angel Celis, Miguel
    Manuel Larraga-Gutierrez, Jose
    Moreno-Jimenez, Sergio
    Martinez-Davalos, Arnulfo
    Rodriguez-Villafuerte, Mercedes
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (03): : 90 - 98
  • [3] American Society for Radiation Oncology, 2014, MOD POL PROT BEAM TH
  • [4] [Anonymous], 2017, SCANNING MICROPROBE
  • [5] [Anonymous], 1993, J ICRU
  • [6] High dose bystander effects in spatially fractionated radiation therapy
    Asur, Rajalakshmi
    Butterworth, Karl T.
    Penagaricano, Jose A.
    Prise, Kevin M.
    Griffin, Robert J.
    [J]. CANCER LETTERS, 2015, 356 (01) : 52 - 57
  • [7] Cancer and Radiation Therapy: Current Advances and Future Directions
    Baskar, Rajamanickam
    Lee, Kuo Ann
    Yeo, Richard
    Yeoh, Kheng-Wei
    [J]. INTERNATIONAL JOURNAL OF MEDICAL SCIENCES, 2012, 9 (03): : 193 - 199
  • [8] Air kerma to Hp(3) conversion coefficients for a new cylinder phantom for photon reference radiation qualities
    Behrens, R.
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 151 (03) : 450 - 455
  • [9] Effects of microbeam radiation therapy on normal and tumoral blood vessels
    Bouchet, Audrey
    Serduc, Raphaeel
    Laissue, Jean Albert
    Djonov, Valentin
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (06): : 634 - 641
  • [10] Synchrotron microbeam radiation therapy induces hypoxia in intracerebral gliosarcoma but not in the normal brain
    Bouchet, Audrey
    Lemasson, Benjamin
    Christen, Thomas
    Potez, Marine
    Rome, Claire
    Coquery, Nicolas
    Le Clec'h, Celine
    Moisan, Anaick
    Braeuer-Krisch, Elke
    Leduc, Geraldine
    Remy, Chantal
    Laissue, Jean A.
    Barbier, Emmanuel L.
    Brun, Emmanuel
    Serduc, Raphael
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 108 (01) : 143 - 148