Calibration Method for Angular Positioning Deviation of a High-Precision Rotary Table Based on the Laser Tracer Multi-Station Measurement System

被引:16
作者
Chen, Hongfang [1 ]
Jiang, Bo [1 ]
Lin, Hu [2 ]
Zhang, Shuang [1 ]
Shi, Zhaoyao [1 ]
Song, Huixu [1 ]
Sun, Yanqiang [1 ]
机构
[1] Beijing Univ Technol, Beijing Engn Res Ctr Precis Measurement Technol &, Beijing 100124, Peoples R China
[2] Natl Inst Metrol, Beijing 102200, Peoples R China
来源
APPLIED SCIENCES-BASEL | 2019年 / 9卷 / 16期
基金
中国国家自然科学基金;
关键词
angular positioning deviation; calibration method; laser tracer multistation; measurement system; rotary table; MACHINE-TOOLS; VERIFICATION;
D O I
10.3390/app9163417
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This paper proposes a calibration method for angular positioning deviation of a high-precision rotary table based on the laser tracer multi-station measurement system. The algorithm error of the calibration method for angular positioning deviation of a high-precision rotary table based on the laser tracer multi-station measurement system was mainly discussed. During the experiments, the laser tracer was fixed on the work surface of the rotary table, and the rotary was fixed on the work surface of the coordinate measurement machine (CMM). The rotary table was rotated with the same angular interval. In this case, an optimization method for calculating the coordinates of a laser tracer station by using Levenberg-Marquardt algorithm and singular value decomposition transform was proposed. Then, the angular positioning deviation of the rotary table was calibrated by an established geometric relationship model between the coordinates of laser tracer stations and the rotation angle of the rotary table. The angular positioning deviation of the high-precision rotary table was as low as +/- 0.9 '', and the error of the calibration method was +/- 0.4 ''. The experimental results proved the feasibility of the proposed calibration method. The calibration method proposed in this paper is suitable for the case that the rotary table is not linked with the CMM, especially for large high-precision rotary tables.
引用
收藏
页数:15
相关论文
共 20 条
[1]   Angular calibration method of precision rotating platform based on laser tracker [J].
Bai P.-J. ;
Xue N. ;
Liu S.-T. ;
Song T. ;
Li J.-H. .
Huanan Ligong Daxue Xuebao/Journal of South China University of Technology (Natural Science), 2016, 44 (01) :100-107
[2]   Uncertainty modeling of the spatial coordinate error correction system of the CMM based on laser tracer multi-station measurement [J].
Chen, Hongfang ;
Jiang, Bo ;
Shi, Zhaoyao ;
Sun, Yanqiang ;
Song, Huixu ;
Tang, Liang .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2019, 30 (02)
[3]   Identification and verification of location errors of rotary axes on five-axis machine tools by using a touch-trigger probe and a sphere [J].
Chen, Yu-Ta ;
More, Pruthvikumar ;
Liu, Chien-Sheng .
INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2019, 100 (9-12) :2653-2667
[4]  
Fan B.X., 2017, MEASUREMENT PRINCIPL
[5]   Method for measuring location errors using a touch trigger probe on four-axis machine tools [J].
Jeong, Ji Hun ;
Khim, Gyungho ;
Oh, Jeong Seok ;
Chung, Sung-Chong .
INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2018, 99 (1-4) :1003-1012
[6]  
Ma C.F., 2017, OPTIMIZATION CALCULA
[7]   Verification of Articulated Arm Coordinate Measuring Machines Accuracy Using LaserTracer System as Standard of Length [J].
Ostrowska, K. ;
Gaska, A. ;
Kupiec, R. ;
Sladek, J. ;
Gromczak, K. .
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2016, 31 (04) :241-256
[8]  
[皮世威 Pi Shiwei], 2017, [仪器仪表学报, Chinese Journal of Scientific Instrument], V38, P2484
[9]   SVD principle analysis and fault diagnosis for bearings based on the correlation coefficient [J].
Qiao, Zijian ;
Pan, Zhengrong .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2015, 26 (08)
[10]   Error mapping of CMMs and machine tools by a single tracking interferometer [J].
Schwenke, H ;
Franke, M ;
Hannaford, J .
CIRP ANNALS-MANUFACTURING TECHNOLOGY, 2005, 54 (01) :475-478