Monte Carlo Evaluation for the Effect of Positional Inaccuracy of Source on Patient's Dose in High-Dose-Rate Brachytherapy

被引:2
作者
Lee, Kyung Ja [1 ]
An, So Hyun [1 ]
Lee, Rena [1 ]
机构
[1] Ewha Womans Univ, Dept Radiat Oncol, Sch Med, Seoul 120750, South Korea
关键词
Monte Carlo models; Brachytherapy; Treatment planning; VERIFICATION;
D O I
10.3938/jkps.55.2361
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Although the potential inaccuracies of source position in high-dose-rate (HDR) brachytherapy could affect the dose to the target and normal tissues, the dose variation to the target and critical organs due to the source position error have not been evaluated. Therefore, the effect of the positional inaccuracy on the dose is evaluated for uterine cervical cancer patients. Ten patients who were eligible for intracavitary brachytherapy were involved in this study. Conventional treatment plans (PLATO planning system) for a Fletcher-type metal applicator (one tandem and two ovoids) were performed to deliver 100 % of the prescribed dose to point A. The planned source dwell positions were shifted by 2.5 mm, and the doses to point A, point B, the external cervical os, the bladder, and the rectum were calculated with both MCNPX and PLATO planning system. As a result, point A has a dose difference within 5 % in all patients. The bladder and the rectum had dose differences within 10 %. PLATO and MCNPX had dose differences within 25 % at all reference points. The results show that the positional inaccuracy of the source more affect organ more than points A and B.
引用
收藏
页码:2361 / 2365
页数:5
相关论文
共 16 条
[1]   MCPI©:: A sub-minute Monte Carlo dose calculation engine for prostate implants [J].
Chibani, O ;
Williamson, JF .
MEDICAL PHYSICS, 2005, 32 (12) :3688-3698
[2]  
DEWERD L, 1994, MED PHYS, V22, P435
[3]   Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera [J].
Duan, J ;
Macey, DJ ;
Pareek, PN ;
Brezovich, IA .
MEDICAL PHYSICS, 2001, 28 (02) :167-173
[4]   High dose-rate brachytherapy source position quality assurance using radiochromic film [J].
Evans, M. D. C. ;
Devic, S. ;
Podgorsak, E. B. .
MEDICAL DOSIMETRY, 2007, 32 (01) :13-15
[5]  
GEORGI D, 1998, MED PHYS, V25, P2200
[6]  
GLENN G, 1979, MED PHYS, V6, P49
[7]   Real-time verification of HDR brachytherapy source location: implementation of detector redundancy [J].
Nakano, T ;
Suchowerska, N ;
McKenzie, DR ;
Bilek, MM .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (02) :319-327
[8]  
NATH R, 1979, 56 AAPM AM I PHYS
[9]  
*NUCL, 2008, APPL GUID 2008
[10]  
*NUCL, 2008, PLATO BRACH REM AFT