摘要
An algorithm is described for reducing ghost artifacts in echo planar imaging (EPI) using phase corrections derived from images reconstructed using only even or odd k-space lines. The N/2 ghost, that arises principally from time-reversal of alternate k-space lines, was significantly reduced by this algorithm without the need for a calibration scan, In images obtained in eight subjects undergoing EPI for auditory functional MRI (fMRI) experiments, N/2 ghost intensity was reduced from 10.3% +/- 2,1% (range: 7.9-14.1%) to 4.5% +/- 0.2% (range: 4.1-4.9%) of parent image intensity, corresponding to a percent reduction in ghost intensity of 54% +/- 9% (range: 43-65%), and the algorithm restored this intensity to the parent image. It provided a significant improvement in image appearance, and increased the correlation coefficients related to neural activation in functional MRI studies. The algorithm provided reduction of artifacts from all polynomial orders of spatial phase errors in both spatial directions, The algorithm did not eliminate N/2 ghost intensity contributed by field inhomogeneities, susceptibility, or chemical shift.
引用
收藏
页码:89 / 100
页数:12
相关论文
共 24 条
  • [1] ALSOP DC, 1995, RADIOLOGY, V197, P388
  • [2] PROCESSING STRATEGIES FOR TIME-COURSE DATA SETS IN FUNCTIONAL MRI OF THE HUMAN BRAIN
    BANDETTINI, PA
    JESMANOWICZ, A
    WONG, EC
    HYDE, JS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) : 161 - 173
  • [3] TEMPORAL AND SPATIAL-ANALYSIS OF FIELDS GENERATED BY EDDY CURRENTS IN SUPERCONDUCTING MAGNETS - OPTIMIZATION OF CORRECTIONS AND QUANTITATIVE CHARACTERIZATION OF MAGNET GRADIENT SYSTEMS
    BOESCH, C
    GRUETTER, R
    MARTIN, E
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) : 268 - 284
  • [4] IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING
    BRUDER, H
    FISCHER, H
    REINFELDER, HE
    SCHMITT, F
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) : 311 - 323
  • [5] BUONOCORE MH, 1994, RADIOLOGY, V193, P136
  • [6] BUONOCORE MH, 1985, P SMRM 4 ANN M LOND, P576
  • [7] INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    THOMPSON, RM
    JACK, CR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) : 67 - 72
  • [8] ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING
    FARZANEH, F
    RIEDERER, SJ
    PELC, NJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) : 123 - 139
  • [9] PHASE ERRORS IN MULTISHOT ECHO-PLANAR IMAGING
    FEINBERG, DA
    OSHIO, K
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) : 535 - 539
  • [10] ECHO-PLANAR IMAGING WITH ASYMMETRIC GRADIENT MODULATION AND INNER-VOLUME EXCITATION
    FEINBERG, DA
    TURNER, R
    JAKAB, PD
    VONKIENLIN, M
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) : 162 - 169