Priori mask guided image reconstruction (p-MGIR) for ultra-low dose cone-beam computed tomography

被引:6
|
作者
Park, Justin C. [1 ]
Zhang, Hao [2 ]
Chen, Yunmei [2 ]
Fan, Qiyong [3 ]
Kahler, Darren L. [1 ]
Liu, Chihray [1 ]
Lu, Bo [1 ]
机构
[1] Univ Florida, Dept Radiat Oncol, Gainesville, FL 32610 USA
[2] Univ Florida, Dept Math, Gainesville, FL 32611 USA
[3] Univ Nebraska Med Ctr, Dept Radiat Oncol, Omaha, NE 68105 USA
关键词
p-MGIR; 3DCBCT; iterative image reconstruction; compressed sensing; on-line IGRT; COMPRESSED SENSING ABOCS; CT RECONSTRUCTION; CANCER-RISKS; GPU; PROJECTION; ALGORITHM; NOISE;
D O I
10.1088/0031-9155/60/21/8505
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently, the compressed sensing (CS) based iterative reconstruction method has received attention because of its ability to reconstruct cone beam computed tomography (CBCT) images with good quality using sparsely sampled or noisy projections, thus enabling dose reduction. However, some challenges remain. In particular, there is always a tradeoff between image resolution and noise/streak artifact reduction based on the amount of regularization weighting that is applied uniformly across the CBCT volume. The purpose of this study is to develop a novel low-dose CBCT reconstruction algorithm framework called priori mask guided image reconstruction (p-MGIR) that allows reconstruction of high-quality low-dose CBCT images while preserving the image resolution. In p-MGIR, the unknown CBCT volume was mathematically modeled as a combination of two regions: (1) where anatomical structures are complex, and (2) where intensities are relatively uniform. The priori mask, which is the key concept of the p-MGIR algorithm, was defined as the matrix that distinguishes between the two separate CBCT regions where the resolution needs to be preserved and where streak or noise needs to be suppressed. We then alternately updated each part of image by solving two sub-minimization problems iteratively, where one minimization was focused on preserving the edge information of the first part while the other concentrated on the removal of noise/artifacts from the latter part. To evaluate the performance of the p-MGIR algorithm, a numerical head-and-neck phantom, a Catphan 600 physical phantom, and a clinical head-and-neck cancer case were used for analysis. The results were compared with the standard Feldkamp-Davis-Kress as well as conventional CS-based algorithms. Examination of the p-MGIR algorithm showed that high-quality low-dose CBCT images can be reconstructed without compromising the image resolution. For both phantom and the patient cases, the p-MGIR is able to achieve a clinically-reasonable image with 60 projections. Therefore, a clinically-viable, high-resolution head-and-neck CBCT image can be obtained while cutting the dose by 83%. Moreover, the image quality obtained using p-MGIR is better than the quality obtained using other algorithms. In this work, we propose a novel low-dose CBCT reconstruction algorithm called p-MGIR. It can be potentially used as a CBCT reconstruction algorithm with low dose scan requests
引用
收藏
页码:8505 / 8524
页数:20
相关论文
共 50 条
  • [21] Deep learning based ultra-low dose fan-beam computed tomography image enhancement algorithm: Feasibility study in image quality for radiotherapy
    Jiang, Hua
    Qin, Songbing
    Jia, Lecheng
    Wei, Ziquan
    Xiong, Weiqi
    Xu, Wentao
    Gong, Wei
    Zhang, Wei
    Yu, Liqin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024,
  • [22] Nonlinear ill-posed problem in low-dose dental cone-beam computed tomography&DAG;
    Park, Hyoung Suk
    Hyun, Chang Min
    Seo, Jin Keun
    IMA JOURNAL OF APPLIED MATHEMATICS, 2023, 89 (01) : 231 - 253
  • [23] A rapid, accurate image simulation strategy for mega-voltage cone-beam computed tomography
    Shi, Mengying
    Myronakis, Marios
    Jacobson, Matthew
    Lehmann, Mathias
    Ferguson, Dianne
    Baturin, Paul
    Huber, Pascal
    Fueglistaller, Rony
    Harris, Thomas
    Lozano, Ingrid Valencia
    Williams, Christopher
    Morf, Daniel
    Berbeco, Ross, I
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (13)
  • [24] The use of deep learning methods in low-dose computed tomography image reconstruction: a systematic review
    Zhang, Minghan
    Gu, Sai
    Shi, Yuhui
    COMPLEX & INTELLIGENT SYSTEMS, 2022, 8 (06) : 5545 - 5561
  • [25] A rapid algebraic 3D volume image reconstruction technique for cone beam computed tomography
    Al-masni, Mohammed A.
    Al-antari, Mugahed A.
    Metwally, Mohamed K.
    Kadah, Yasser M.
    Han, Seung-Moo
    Kim, Tae-Seong
    BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2017, 37 (04) : 619 - 629
  • [26] Filtered-backprojection reconstruction for a cone-beam computed tomography scanner with independent source and detector rotations
    Rit, Simon
    Clackdoyle, Rolf
    Keuschnigg, Peter
    Steininger, Philipp
    MEDICAL PHYSICS, 2016, 43 (05) : 2344 - 2352
  • [27] Performance evaluation of local ROI algorithms for exact ROI reconstruction in spiral cone-beam computed tomography
    Sourbelle, K
    Lauritsch, G
    Tam, KC
    Noo, F
    Kalender, WA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 697 - 702
  • [28] A Computationally Efficient Reconstruction Algorithm for Circular Cone-Beam Computed Tomography Using Shallow Neural Networks
    Lagerwerf, Marinus J.
    Pelt, Daniel M.
    Palenstijn, Willem Jan
    Batenburg, Kees Joost
    JOURNAL OF IMAGING, 2020, 6 (12)
  • [29] An approximate reconstruction method for helical cone-beam differential phase-contrast computed tomography images
    Li, Jing
    Zhu, Peiping
    Sun, Yi
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08) : 2347 - 2356
  • [30] Sparse-View Image Reconstruction in Cone-Beam Computed Tomography with Variance-Reduced Stochastic Gradient Descent and Locally-Adaptive Proximal Operation
    Karimi, Davood
    Ward, Rabab K.
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2017, 37 (03) : 420 - 440