Biomechanical comparison of femoral fixation devices for anterior cruciate ligament reconstruction using a novel testing method

被引:8
|
作者
Ehrensberger, Mark [1 ,2 ]
Hohman, Donald W., Jr. [2 ]
Duncan, Kurt [2 ]
Howard, Craig [2 ]
Bisson, Leslie [2 ]
机构
[1] SUNY Buffalo, Dept Biomed Engn, Buffalo, NY 14260 USA
[2] SUNY Buffalo, Dept Orthopaed Surg, Buffalo, NY 14260 USA
关键词
Anterior cruciate ligament (ACL); Biomechanics; Femoral anchor; Graft fixation; HAMSTRING TENDON GRAFT; INTERFERENCE SCREW FIXATION; IN-VITRO; PORCINE MODEL; PULL-OUT; STRENGTH; AUGMENTATION; TITANIUM; SITE;
D O I
10.1016/j.clinbiomech.2012.12.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A novel biomechanical test method was implemented to compare the mechanical performance of two femoral fixation anchors (AperFix(r), Cayenne Medical, Scottsdale, AZ, USA or the AppianFx(r), KFx Medical, Carlsbad, CA, USA) that were utilized in anterior cruciate ligament reconstruction. Methods: Anterior cruciate ligament reconstructions were performed in 20 porcine femurs by using bovine extensor tendon grafts secured with 9 mm femoral anchors (AperFix(r) or AppianFx(r)). 10 specimens were tested for each anchor type. Infrared position sensors determined the repair construct displacements during conditioning (20 cycles at 5-50 N at 0.25 Hz), cyclic loading (1500 cycles at 50-200 N at 1 Hz), and ultimate loading (150 mm/min). Outcomes included tendon elongation, anchor displacement, stiffness, maximum load, yield load, and load at 5mm of anchor displacement. It was hypothesized that there would be no differences in the outcomes of these two devices. Independent measure t-tests compared the performance of the devices (p<0.05). Findings: The performance of the two anchors was comparable during the cyclic loading. During ultimate loading, a statistically higher yield load (p<0.01) and a load at 5 mm of anchor displacement (p<0.01) were demonstrated for the AppianFx(r) as compared to AperFix(r). Maximum load and stiffness were not significantly different. Interpretation: Given the good clinical track record of the AperFix(r), the comparable, and in some cases superior, the biomechanical data presented here for the AppianFx(r) are encouraging for their clinical implementation. This study also introduced a novel test method that directly tracks the relevant construct displacements during cyclic and ultimate loading tests of the anterior cruciate ligament reconstructions. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:193 / 198
页数:6
相关论文
共 50 条
  • [41] Biomechanical study of a computer simulated reconstruction of the anterior cruciate ligament (ACL)
    Veselko, M
    Godler, I
    COMPUTERS IN BIOLOGY AND MEDICINE, 2000, 30 (06) : 299 - 309
  • [42] Clinical comparison of two suspensory fixation devices for anatomic double-bundle anterior cruciate ligament reconstruction
    Kondo, Eiji
    Yasuda, Kazunori
    Miyatake, Shin
    Kitamura, Nobuto
    Tohyama, Harukazu
    Yagi, Tomonori
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (07) : 1261 - 1267
  • [43] In vitro biomechanical testing of anterior cruciate ligament reconstruction: Traditional versus physiologically relevant load analysis
    Trump, Mark
    Palathinkal, Darren M.
    Beaupre, Lauren
    Otto, Dave
    Leung, Paul
    Amirfazli, A.
    KNEE, 2011, 18 (03) : 193 - 201
  • [44] Hybrid femoral fixation of soft-tissue grafts in anterior cruciate ligament reconstruction using the EndoButton CL and bioabsorbable interference screws: A biomechanical study
    Oh, Young Ho
    Namkoong, Suk
    Strauss, Eric J.
    Ishak, Charbel
    Jazrawi, Laith M.
    Rosen, Jeffrey
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (11) : 1218 - 1224
  • [45] Comparative biomechanical study of the Ligament Plate® and other fixation devices in ACL reconstruction
    Sim, Jae Ang
    Kwak, Ji Hoon
    Yang, Sang Hoon
    Lee, Beom Koo
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (05) : 1269 - 1274
  • [46] Graft fixation alternatives in anterior cruciate ligament reconstruction
    Saccomanni B.
    MUSCULOSKELETAL SURGERY, 2011, 95 (3) : 183 - 191
  • [47] Biomechanical Effects of Additional Anterolateral Structure Reconstruction With Different Femoral Attachment Sites on Anterior Cruciate Ligament Reconstruction
    Katakura, Mai
    Koga, Hideyuki
    Nakamura, Tomomasa
    Araki, Daisuke
    Nagai, Kanto
    Nishida, Kyohei
    Kuroda, Ryosuke
    Muneta, Takeshi
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2019, 47 (14) : 3373 - 3380
  • [48] Onlay Reconstruction of the Posterior Cruciate Ligament: Biomechanical Comparison of Unicortical and Bicortical Tibial Fixation
    de Albuquerque, Joao Bourbon, II
    Pfeiffer, Ferris
    Stannard, James P.
    Cook, James L.
    Kfuri, Mauricio
    JOURNAL OF KNEE SURGERY, 2019, 32 (10) : 972 - 978
  • [49] Pull-out strength of four tibial fixation devices used in anterior cruciate ligament reconstruction
    Chivot, M.
    Harrosh, S.
    Kelberine, F.
    Pithioux, M.
    Argenson, J. -N.
    Ollivier, M.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2018, 104 (02) : 203 - 207
  • [50] Comparison of femoral fixation methods for anterior cruciate ligament reconstruction with patellar tendon graft: a mechanical analysis in porcine knees
    Giuseppe Milano
    Pier Damiano Mulas
    Fabio Ziranu
    Laura Deriu
    Carlo Fabbriciani
    Knee Surgery, Sports Traumatology, Arthroscopy, 2007, 15 : 733 - 738