Biomechanical comparison of femoral fixation devices for anterior cruciate ligament reconstruction using a novel testing method

被引:8
|
作者
Ehrensberger, Mark [1 ,2 ]
Hohman, Donald W., Jr. [2 ]
Duncan, Kurt [2 ]
Howard, Craig [2 ]
Bisson, Leslie [2 ]
机构
[1] SUNY Buffalo, Dept Biomed Engn, Buffalo, NY 14260 USA
[2] SUNY Buffalo, Dept Orthopaed Surg, Buffalo, NY 14260 USA
关键词
Anterior cruciate ligament (ACL); Biomechanics; Femoral anchor; Graft fixation; HAMSTRING TENDON GRAFT; INTERFERENCE SCREW FIXATION; IN-VITRO; PORCINE MODEL; PULL-OUT; STRENGTH; AUGMENTATION; TITANIUM; SITE;
D O I
10.1016/j.clinbiomech.2012.12.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A novel biomechanical test method was implemented to compare the mechanical performance of two femoral fixation anchors (AperFix(r), Cayenne Medical, Scottsdale, AZ, USA or the AppianFx(r), KFx Medical, Carlsbad, CA, USA) that were utilized in anterior cruciate ligament reconstruction. Methods: Anterior cruciate ligament reconstructions were performed in 20 porcine femurs by using bovine extensor tendon grafts secured with 9 mm femoral anchors (AperFix(r) or AppianFx(r)). 10 specimens were tested for each anchor type. Infrared position sensors determined the repair construct displacements during conditioning (20 cycles at 5-50 N at 0.25 Hz), cyclic loading (1500 cycles at 50-200 N at 1 Hz), and ultimate loading (150 mm/min). Outcomes included tendon elongation, anchor displacement, stiffness, maximum load, yield load, and load at 5mm of anchor displacement. It was hypothesized that there would be no differences in the outcomes of these two devices. Independent measure t-tests compared the performance of the devices (p<0.05). Findings: The performance of the two anchors was comparable during the cyclic loading. During ultimate loading, a statistically higher yield load (p<0.01) and a load at 5 mm of anchor displacement (p<0.01) were demonstrated for the AppianFx(r) as compared to AperFix(r). Maximum load and stiffness were not significantly different. Interpretation: Given the good clinical track record of the AperFix(r), the comparable, and in some cases superior, the biomechanical data presented here for the AppianFx(r) are encouraging for their clinical implementation. This study also introduced a novel test method that directly tracks the relevant construct displacements during cyclic and ultimate loading tests of the anterior cruciate ligament reconstructions. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:193 / 198
页数:6
相关论文
共 50 条
  • [21] Biomechanical Evaluation of a 1-Stage Revision Anterior Cruciate Ligament Reconstruction Technique Using a Structural Bone Void Filler for Femoral Fixation
    Vaughn, Zackary D.
    Schmidt, Josh
    Lindsey, Derek P.
    Dragoo, Jason L.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2009, 25 (09) : 1011 - 1018
  • [22] In vitro comparison between cortical and cortico-cancellous femoral suspension devices for anterior cruciate ligament reconstruction: implications for mobilization
    Rodriguez, Cristina
    Eduardo Garcia, Tomas
    Montes, Susana
    Rodriguez, Luis
    Maestro, Antonio
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (08) : 2324 - 2329
  • [23] Anterior cruciate ligament reconstruction: Assessment of the hamstring autograft femoral fixation using the EndoButton CL®
    Plaweski, S.
    Rossi, J.
    Merloz, P.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2009, 95 (08) : 606 - 613
  • [24] Biomechanical analysis of press-fit fixation of anterior cruciate ligament transplants
    Jagodzinski, M.
    Ettinger, M.
    Haasper, C.
    Hankemeier, S.
    Breitmeier, D.
    Hurschler, C.
    Krettek, C.
    UNFALLCHIRURG, 2010, 113 (07): : 532 - 539
  • [25] A Biomechanical Comparison of 2 Femoral Fixation Techniques for Anterior Cruciate Ligament Reconstruction in Skeletally Immature Patients: Over-the-Top Fixation Versus Transphyseal Technique
    Lertwanich, Pisit
    Kato, Yuki
    Martins, Cesar A. Q.
    Maeyama, Akira
    Ingham, Sheila J. M.
    Kramer, Scott
    Linde-Rosen, Monica
    Smolinski, Patrick
    Fu, Freddie H.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2011, 27 (05) : 672 - 680
  • [26] Tibial tunnel widening after anterior cruciate ligament reconstructions with hamstring tendons using Rigidfix femoral fixation and Intrafix tibial fixation
    Choi, Nam-Hong
    Lee, Jung-Hoon
    Son, Kyung-Mo
    Victoroff, Brian N.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2010, 18 (01) : 92 - 97
  • [27] Anterior cruciate ligament reconstruction: graft fixation
    Wagner, M.
    Scheffler, S. U.
    Weiler, A.
    ARTHROSKOPIE, 2005, 18 (01) : 27 - 35
  • [28] Biomechanical considerations for graft choice in anterior cruciate ligament reconstruction
    Banovetz, Mark T.
    Kennedy, Nicholas I.
    LaPrade, Robert F.
    Engebretsen, Lars
    Moatshe, Gilbert
    ANNALS OF JOINT, 2023, 8
  • [29] Biomechanical considerations for graft choice in anterior cruciate ligament reconstruction
    Banoyetz, Mark T.
    Kennedy, Nicholas, I
    LaPrade, Robert F.
    Engebretsen, Lars
    Moatshe, Gilbert
    ANNALS OF JOINT, 2023, 8
  • [30] Lateral Femoral Cortical Breach During Anterior Cruciate Ligament Reconstruction: A Biomechanical Analysis
    Hammond, Kyle E.
    Dierckman, Brian D.
    Potini, Vishnu C.
    Xerogeanes, John W.
    Labib, Sameh A.
    Hutton, William C.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2012, 28 (03) : 365 - 371