Biomechanical comparison of femoral fixation devices for anterior cruciate ligament reconstruction using a novel testing method

被引:8
|
作者
Ehrensberger, Mark [1 ,2 ]
Hohman, Donald W., Jr. [2 ]
Duncan, Kurt [2 ]
Howard, Craig [2 ]
Bisson, Leslie [2 ]
机构
[1] SUNY Buffalo, Dept Biomed Engn, Buffalo, NY 14260 USA
[2] SUNY Buffalo, Dept Orthopaed Surg, Buffalo, NY 14260 USA
关键词
Anterior cruciate ligament (ACL); Biomechanics; Femoral anchor; Graft fixation; HAMSTRING TENDON GRAFT; INTERFERENCE SCREW FIXATION; IN-VITRO; PORCINE MODEL; PULL-OUT; STRENGTH; AUGMENTATION; TITANIUM; SITE;
D O I
10.1016/j.clinbiomech.2012.12.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: A novel biomechanical test method was implemented to compare the mechanical performance of two femoral fixation anchors (AperFix(r), Cayenne Medical, Scottsdale, AZ, USA or the AppianFx(r), KFx Medical, Carlsbad, CA, USA) that were utilized in anterior cruciate ligament reconstruction. Methods: Anterior cruciate ligament reconstructions were performed in 20 porcine femurs by using bovine extensor tendon grafts secured with 9 mm femoral anchors (AperFix(r) or AppianFx(r)). 10 specimens were tested for each anchor type. Infrared position sensors determined the repair construct displacements during conditioning (20 cycles at 5-50 N at 0.25 Hz), cyclic loading (1500 cycles at 50-200 N at 1 Hz), and ultimate loading (150 mm/min). Outcomes included tendon elongation, anchor displacement, stiffness, maximum load, yield load, and load at 5mm of anchor displacement. It was hypothesized that there would be no differences in the outcomes of these two devices. Independent measure t-tests compared the performance of the devices (p<0.05). Findings: The performance of the two anchors was comparable during the cyclic loading. During ultimate loading, a statistically higher yield load (p<0.01) and a load at 5 mm of anchor displacement (p<0.01) were demonstrated for the AppianFx(r) as compared to AperFix(r). Maximum load and stiffness were not significantly different. Interpretation: Given the good clinical track record of the AperFix(r), the comparable, and in some cases superior, the biomechanical data presented here for the AppianFx(r) are encouraging for their clinical implementation. This study also introduced a novel test method that directly tracks the relevant construct displacements during cyclic and ultimate loading tests of the anterior cruciate ligament reconstructions. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:193 / 198
页数:6
相关论文
共 50 条
  • [1] Biomechanical comparison of anterior cruciate ligament reconstruction fixation methods and implications on clinical outcomes
    McDermott, Emily
    DeFoor, Mikalyn T.
    Blaber, Olivia K.
    Aman, Zachary S.
    DePhillipo, Nicholas N.
    Dekker, Travis J.
    ANNALS OF JOINT, 2023, 8
  • [2] Three Femoral Fixation Devices for Anterior Cruciate Ligament Reconstruction: Comparison of Fixation on the Lateral Cortex Versus the Anterior Cortex
    Conner, Chad S.
    Perez, Brandon A.
    Morris, Randal P.
    Buckner, James W.
    Buford, William L., Jr.
    Ivey, Frank M.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2010, 26 (06) : 796 - 807
  • [3] Comparison of Outcomes of Two Femoral Fixation Devices in Hamstring Anterior Cruciate Ligament Reconstruction
    Angel Sanchez-Carrasco, Miguel
    Francisco Abellan, Juan
    Qudsi-Sinclair, Salima
    Ruiz-Merino, Guadalupe
    Javier Carrillo-Julia, Francisco
    Bo-Rueda, David
    INDIAN JOURNAL OF ORTHOPAEDICS, 2017, 51 (05) : 487 - 492
  • [4] Biomechanical Evaluation of Anterior Cruciate Ligament Femoral Fixation Techniques
    Kleweno, Conor P.
    Jacir, Alberto M.
    Gardner, Thomas R.
    Ahmad, Christopher S.
    Levine, William N.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2009, 37 (02) : 339 - 345
  • [5] Evaluation and comparison of clinical results of femoral fixation devices in arthroscopic anterior cruciate ligament reconstruction
    Aydin, Deniz
    Ozcan, Mert
    KNEE, 2016, 23 (02) : 227 - 232
  • [6] Femoral Cortical Suspension Devices for Soft Tissue Anterior Cruciate Ligament Reconstruction A Comparative Biomechanical Study
    Petre, Benjamin M.
    Smith, Sean D.
    Jansson, Kyle S.
    de Meijer, Peter-Paul
    Hackett, Thomas R.
    LaPrade, Robert F.
    Wijdicks, Coen A.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2013, 41 (02) : 416 - 422
  • [7] Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
    Jin Cheng
    Siddhartha Venkata Paluvadi
    SungJae Lee
    SeungJin Yoo
    Eun-Kyoo Song
    Jong-Keun Seon
    International Orthopaedics, 2018, 42 : 1291 - 1296
  • [8] Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
    Cheng, Jin
    Paluvadi, Siddhartha Venkata
    Lee, SungJae
    Yoo, SeungJin
    Song, Eun-Kyoo
    Seon, Jong-Keun
    INTERNATIONAL ORTHOPAEDICS, 2018, 42 (06) : 1291 - 1296
  • [9] Biomechanical comparison of anterior cruciate ligament reconstruction fixation methods and implications on clinical outcomes
    McDermott, Emily
    DeFoor, Mikalyn T.
    Blaber, Olivia K.
    Aman, Zachary S.
    DePhillipo, Nicholas N.
    Dekker, Travis J.
    ANNALS OF JOINT, 2023, 8
  • [10] Reverse Anterior Cruciate Ligament Reconstruction Fixation: A Biomechanical Comparison Study of Tibial Cross-Pin and Femoral Interference Screw Fixation
    Lawley, Richard J.
    Klein, Samuel E.
    Chudik, Steven C.
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2017, 33 (03) : 625 - 632