Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels

被引:92
|
作者
Alfa, Michelle J. [1 ,2 ,3 ]
Fatima, Iram [3 ]
Olson, Nancy [3 ]
机构
[1] Diagnost Serv Manitoba, Clin Microbiol, Winnipeg, MB R2H 2A6, Canada
[2] St Boniface Res Ctr, Microbiol Lab, Winnipeg, MB, Canada
[3] Univ Manitoba, Dept Med Microbiol, Winnipeg, MB, Canada
关键词
Simulated use; Endoscope channel; Cleaning-compliance; Benchmark; Relative light units; DECONTAMINATION; EFFICACY;
D O I
10.1016/j.ajic.2012.03.018
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Compliance with cleaning of flexible endoscope channels cannot be verified using visual inspection. Adenosine triphosphate (ATP) has been suggested as a possible rapid cleaning monitor for flexible endoscope channels. There have not been published validation studies to specify the level of ATP that indicates inadequate cleaning has been achieved. Objective: The objective of this study was to validate the Clean-Trace (3M Inc, St. Paul, MN) ATP water test method for monitoring manual cleaning of flexible endoscopes. Methods: This was a simulated use study using a duodenoscope as the test device. Artificial test soil containing 10(6) colony-forming units of Pseudomonas aeruginosa and Enterococcus faecalis was used to perfuse all channels. The flush sample method for the suction-biopsy (L1) or air-water channel (L2) using 40 and 20 mLs sterile reverse osmosis water, respectively, was validated. Residuals of ATP, protein, hemoglobin, and bioburden were quantitated from channel samples taken from uncleaned, partially cleaned, and fully cleaned duodenoscopes. The benchmarks for clean were as follows: <6.4 mu g/cm(2) protein, <2.2 mu g/cm(2) hemoglobin, and <4-log(10) colony-forming units/cm(2) bioburden. Results: The average ATP in clean channel samples was 27.7 RLUs and 154 RLUs for L1 and L2, respectively (<200 RLUs for all channels). The average protein, hemoglobin, and bioburden benchmarks were achieved if <200 RLUs were detected. If the channel sample was >200 RLUs, the residual organic and bioburden levels would exceed the acceptable benchmarks. Conclusion: Our data validated that flexible endoscopes that have complete manual cleaning will have <200 RLUs by the Clean-Trace ATP test. Copyright (C) 2013 by the Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 9 条
  • [1] The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels
    Alfa, Michelle J.
    Fatima, Iram
    Olson, Nancy
    AMERICAN JOURNAL OF INFECTION CONTROL, 2013, 41 (03) : 249 - 253
  • [2] Development and validation of rapid use scope test strips to determine the efficacy of manual cleaning for flexible endoscope channels
    Alfa, Michelle J.
    Olson, Nancy
    DeGagne, Pat
    Simner, Patricia J.
    AMERICAN JOURNAL OF INFECTION CONTROL, 2012, 40 (09) : 860 - 865
  • [3] Use of adenosine triphosphate to audit reprocessing of flexible endoscopes with an elevator mechanism
    Quan, Erik
    Mahmood, Rizwan
    Naik, Amar
    Sargon, Peter
    Shastri, Nikhil
    Venu, Mukund
    Parada, Jorge P.
    Gupta, Neil
    AMERICAN JOURNAL OF INFECTION CONTROL, 2018, 46 (11) : 1272 - 1277
  • [4] The validity of adenosine triphosphate measurement in detecting endoscope contamination
    McCafferty, C. E.
    Abi-Hanna, D.
    Aghajani, M. J.
    Micali, G. T.
    Lockart, I
    Vickery, K.
    Gosbell, I. B.
    Jensen, S. O.
    JOURNAL OF HOSPITAL INFECTION, 2018, 100 (03) : E142 - E145
  • [5] Physical and composition characteristics of clinical secretions compared with test soils used for validation of flexible endoscope cleaning
    Alfa, M. J.
    Olson, N.
    JOURNAL OF HOSPITAL INFECTION, 2016, 93 (01) : 83 - 88
  • [6] Simulated-use evaluation of rapid ChannelCheckTM cleaning test for optimal detection of organic residues in flexible endoscope channels
    Kulkarni, K.
    Gavette, M.
    Alfa, M. J.
    JOURNAL OF HOSPITAL INFECTION, 2024, 152 : 66 - 72
  • [7] Methodological proposal for validation of the disinfecting efficacy of an automated flexible endoscope reprocessor
    Graziano, Kazuko Uchikawa
    Auler Pereira, Marta Elisa
    Koda, Elaine
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2016, 24
  • [8] A new buildup biofilm model that mimics accumulation of material in flexible endoscope channels
    Luciano, Cristiana da Costa
    Olson, Nancy
    DeGagne, Patricia
    Franca, Rodrigo
    Veiga Tipple, Anaclara Ferreira
    Alfa, Michelle
    JOURNAL OF MICROBIOLOGICAL METHODS, 2016, 127 : 224 - 229
  • [9] On-site comparison of an enzymatic detergent and a non-enzymatic detergent-disinfectant for routine manual cleaning of flexible endoscopes
    Gonzalez, Jonathan Alfageme
    Vanzieleghem, Thomas
    Dumazy, Axelle
    Meuris, Christelle
    Mutsers, Jacques
    Christiaens, Genevieve
    Leclercq, Philippe
    Loly, Jean-Philippe
    Louis, Edouard
    Gast, Pierrette
    ENDOSCOPY INTERNATIONAL OPEN, 2019, 7 (04) : E412 - E420