Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels

被引:94
作者
Alfa, Michelle J. [1 ,2 ,3 ]
Fatima, Iram [3 ]
Olson, Nancy [3 ]
机构
[1] Diagnost Serv Manitoba, Clin Microbiol, Winnipeg, MB R2H 2A6, Canada
[2] St Boniface Res Ctr, Microbiol Lab, Winnipeg, MB, Canada
[3] Univ Manitoba, Dept Med Microbiol, Winnipeg, MB, Canada
关键词
Simulated use; Endoscope channel; Cleaning-compliance; Benchmark; Relative light units; DECONTAMINATION; EFFICACY;
D O I
10.1016/j.ajic.2012.03.018
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Compliance with cleaning of flexible endoscope channels cannot be verified using visual inspection. Adenosine triphosphate (ATP) has been suggested as a possible rapid cleaning monitor for flexible endoscope channels. There have not been published validation studies to specify the level of ATP that indicates inadequate cleaning has been achieved. Objective: The objective of this study was to validate the Clean-Trace (3M Inc, St. Paul, MN) ATP water test method for monitoring manual cleaning of flexible endoscopes. Methods: This was a simulated use study using a duodenoscope as the test device. Artificial test soil containing 10(6) colony-forming units of Pseudomonas aeruginosa and Enterococcus faecalis was used to perfuse all channels. The flush sample method for the suction-biopsy (L1) or air-water channel (L2) using 40 and 20 mLs sterile reverse osmosis water, respectively, was validated. Residuals of ATP, protein, hemoglobin, and bioburden were quantitated from channel samples taken from uncleaned, partially cleaned, and fully cleaned duodenoscopes. The benchmarks for clean were as follows: <6.4 mu g/cm(2) protein, <2.2 mu g/cm(2) hemoglobin, and <4-log(10) colony-forming units/cm(2) bioburden. Results: The average ATP in clean channel samples was 27.7 RLUs and 154 RLUs for L1 and L2, respectively (<200 RLUs for all channels). The average protein, hemoglobin, and bioburden benchmarks were achieved if <200 RLUs were detected. If the channel sample was >200 RLUs, the residual organic and bioburden levels would exceed the acceptable benchmarks. Conclusion: Our data validated that flexible endoscopes that have complete manual cleaning will have <200 RLUs by the Clean-Trace ATP test. Copyright (C) 2013 by the Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:245 / 248
页数:4
相关论文
共 15 条
[1]  
*AAMI, 2003, TIR30 AAMI
[2]   Evaluation of ATP Bioluminescence Assays for Potential Use in a Hospital Setting [J].
Aiken, Zoie A. ;
Wilson, Michael ;
Pratten, Jonathan .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2011, 32 (05) :507-509
[3]   Worst-case soiling levels for patient-used flexible endoscopes before and after cleaning [J].
Alfa, MJ ;
Degagne, P ;
Olson, N .
AMERICAN JOURNAL OF INFECTION CONTROL, 1999, 27 (05) :392-401
[4]  
Alfa MJ, 2005, ZENTR STERIL, V13, P387
[5]  
Association for the Advancement of Medical Instrumentation (AAMI) TIR 12, 2004, 12 TIR AAMI
[6]  
AUMERAN C, 2010, ENDOSCOPY, V42, P957
[7]   ESGE-ESGENA guideline for quality assurance in reprocessing: Microbiological surveillance testing in endoscopy [J].
Beilenhoff, U. ;
Neumann, C. S. ;
Rey, J. F. ;
Biering, H. ;
Blum, R. ;
Schmidt, V. .
ENDOSCOPY, 2007, 39 (02) :175-181
[8]   A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure [J].
Bisset, Linda ;
Cossart, Yvonne E. ;
Selby, Warwick ;
West, Richard ;
Catterson, Denise ;
O'Hara, Kate ;
Vickery, Karen .
AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (05) :274-280
[9]   The effectiveness of existing and modified cleaning regimens in a Welsh hospital [J].
Griffith, C. J. ;
Obee, P. ;
Cooper, R. A. ;
Burton, N. F. ;
Lewis, M. .
JOURNAL OF HOSPITAL INFECTION, 2007, 66 (04) :352-359
[10]  
Hansen Dorothea, 2004, Ger Med Sci, V2, pDoc04