High-resolution energy and fast timing measurements for nuclear spectroscopy using digital signal processing

被引:0
作者
Tandel, S. K. [1 ]
Negi, D. [1 ]
Wahid, S. G. [1 ]
Suman, S. [1 ]
Patel, A. [1 ]
Roy, P. [1 ]
Hemalatha, M. [2 ]
Biswas, D. C. [3 ]
Venaruzzo, M. [4 ]
Tintori, C. [4 ]
机构
[1] Univ Mumbai, UM DAE Ctr Excellence Basic Sci, Sch Phys Sci, Mumbai 400098, Maharashtra, India
[2] Univ Mumbai, Dept Phys, Mumbai 400098, Maharashtra, India
[3] Bhabha Atom Res Ctr, Nucl Phys Div, Mumbai 400085, Maharashtra, India
[4] CAEN SpA, Via Vetraia 11, I-55049 Viareggio, LU, Italy
关键词
Digital signal processing (DSP); Gamma detectors (scintillators; CZT; HPGe; HgI etc); Spectrometers; TIME; PERFORMANCE; ANALOG;
D O I
10.1088/1748-0221/15/08/P08013
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Simultaneous high-resolution energy and picosecond timing measurements have been established with an 8-channel, 14-bit, 500 MS/s digitizer coupled to semiconductor and scintillation radiation detectors, resulting in performance equivalent to that obtained with analog electronics. By interpolating between the 2 nanosecond time stamps associated with the sampling frequency, timing up to few tens of picoseconds is achieved. Energy resolution and linearity are determined to be excellent. Coincidence and precise lifetime measurements performed indicate the utility in complex gamma-ray spectroscopic studies.
引用
收藏
页数:10
相关论文
共 21 条
  • [1] Energy resolution and throughput of a new real time digital pulse processing system for x-ray and gamma ray semiconductor detectors
    Abbene, L.
    Gerardi, G.
    Raso, G.
    Basile, S.
    Brai, M.
    Principato, F.
    [J]. JOURNAL OF INSTRUMENTATION, 2013, 8
  • [2] Fission fragment velocity distribution measurement using time of flight technique
    Biswas, D. C.
    Vind, R. P.
    Kumar, Nishant
    Gupta, Y. K.
    Jangale, R., V
    Inkar, A. L.
    Kinage, L. A.
    Joshi, B. N.
    Mukhopadhyay, S.
    Prajapati, G. K.
    Dubey, Shradha
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 901 : 76 - 83
  • [3] The validation of a digital coincidence counting system
    Buckman, SM
    Keightley, JD
    Smith, D
    Woods, MJ
    [J]. APPLIED RADIATION AND ISOTOPES, 1998, 49 (9-11) : 1135 - 1140
  • [4] Digital pulse height correction in HgI2 γ-ray detectors
    Chaudhuri, S. K.
    Lohstroh, A.
    Nakhostin, M.
    Sellin, P. J.
    [J]. JOURNAL OF INSTRUMENTATION, 2012, 7
  • [5] FAST TIMING WITH HAMAMATSU R2083Q PHOTOMULTIPLIERS
    DEVRIES, J
    ZECCA, A
    BRUSA, RS
    GRISENTI, RG
    OSS, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 275 (01) : 194 - 196
  • [6] Digital pulse processing for NaI(T1) detectors
    Di Fulvio, A.
    Shin, T. H.
    Hamel, M. C.
    Pozzi, S. A.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 806 : 169 - 174
  • [7] On nuclear spectrometry pulses digital shaping and processing
    Imperiale, C
    Imperiale, A
    [J]. MEASUREMENT, 2001, 30 (01) : 49 - 73
  • [8] DIGITAL SYNTHESIS OF PULSE SHAPES IN REAL-TIME FOR HIGH-RESOLUTION RADIATION SPECTROSCOPY
    JORDANOV, VT
    KNOLL, GF
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) : 337 - 345
  • [9] Comparison of analog and digital signal processing systems using pulsers
    Koskelo, MJ
    Koskelo, IJ
    Sielaff, B
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) : 373 - 378
  • [10] Performance analysis of gamma ray spectrometric parameters on digital signal and analog signal processing based MCA systems using NaI(Tl) detector
    Kukreti, B. M.
    Sharma, G. K.
    [J]. APPLIED RADIATION AND ISOTOPES, 2012, 70 (05) : 901 - 905