Body composition assessment in adults with cystic fibrosis: comparison of dual-energy X-ray absorptiometry with skinfolds and bioelectrical impedance analysis

被引:36
|
作者
King, S
Wilson, J
Kotsimbos, T
Bailey, M
Nyulasi, I
机构
[1] Monash Univ, Alfred Hosp, Dept Med, Melbourne, Vic 3004, Australia
[2] Alfred Hosp, Dept Nutr, Melbourne, Vic, Australia
[3] Alfred Hosp, Dept Allergy Immunol & Resp Med, Melbourne, Vic, Australia
[4] Monash Univ, Dept Epidemiol & Prevent Med, Melbourne, Vic 3004, Australia
[5] Monash Univ, Med Ctr, Dept Med, Clayton, Vic 3168, Australia
基金
英国医学研究理事会;
关键词
cystic fibrosis; nutritional assessment; body composition; dual-energy X-ray absorptiometry; bioelectrical impedance analysis;
D O I
10.1016/j.nut.2005.04.005
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: We compared body composition measurement in adults with cystic fibrosis (CF) by using non-invasive methods (skinfold thicknesses and bioelectrical impedance analysis [BIA]) with dual-energy X-ray absorptiometry (DXA). Methods: Seventy-six adults with CF (mean age 29.9 +/- 7.9 y, mean body mass index 21.5 +/- 2.5 kg/m(2)) were studied. Body composition was measured to calculate fat-free mass (FFM) using DXA, the sum of four skinfold thicknesses, and BIA (predictive equations of Lukaski and of Segal). Results: Mean FFM values +/- standard deviation measured using DXA were 54.8 +/- 7.3 kg in men and 41.2 +/- 3.9 kg in women. Mean FFM values measured using BIA/Lukaski were 51.5 +/- 7.8 kg in men and 40.4 +/- 4.9 kg in women (P < 0.0005 for men, not significant for women for comparison with DXA). Mean FFM values measured using BIA/Segal were 54.2 +/- 7.5 kg for men and 44.1 +/- 5.9 kg for women (not significant for men, P < 0.0005 for women for comparison with DXA). Mean FFM values measured using skinfolds were significantly higher than those for FFM with DXA (57.2 +/- 7.2 kg in men, 43.3 +/- 4.3 kg in women, P < 0.0005 for comparison with DXA). The 95% limits of agreement with FFM using DXA were, for men and women, respectively, -8.3 to 1.7 kg and -6.4 to 4.8 kg for BIA/Lukaski, -4.8 to 3.6 kg and -3.1 to 8.9 kg for BIA/Segal, and -2.8 to 7.3 kg and -1.5 to 5.7 kg for skinfolds. Conclusion: This study suggests that skinfold thickness measurements and BIA will incorrectly estimate FFM in many adults with CF compared with DXA measurements of FFM. These methods have limited application in the assessment of body composition in individual adult patients with CF. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:1087 / 1094
页数:8
相关论文
共 50 条
  • [31] BODY-COMPOSITION FOLLOWING HEMODIALYSIS - STUDIES USING DUAL-ENERGY X-RAY ABSORPTIOMETRY AND BIOELECTRICAL-IMPEDANCE ANALYSIS
    FORMICA, C
    ATKINSON, MG
    NYULASI, I
    MCKAY, J
    HEALE, W
    SEEMAN, E
    OSTEOPOROSIS INTERNATIONAL, 1993, 3 (04) : 192 - 197
  • [32] Body composition assessment of undernourished older subjects by dual-energy x-ray absorptiometry and bioelectric impedance analysis
    Vilaca, K. H. C.
    Paula, F. J. A.
    Ferriolli, E.
    Lima, N. K. C.
    Marchini, J. S.
    Moriguti, J. C.
    JOURNAL OF NUTRITION HEALTH & AGING, 2011, 15 (06): : 439 - 443
  • [33] Body composition assessment of undernourished older subjects by dual-energy x-ray absorptiometry and bioelectric impedance analysis
    Karla Helena Coelho Vilaça
    F. J. A. Paula
    E. Ferriolli
    N. K. C. Lima
    J. S. Marchini
    J. C. Moriguti
    The journal of nutrition, health & aging, 2011, 15 : 439 - 443
  • [34] Validation of rapid 4-component body composition assessment with the use of dual-energy X-ray absorptiometry and bioelectrical impedance analysis
    Ng, Bennett K.
    Liu, Yong L.
    Wang, Wei
    Kelly, Thomas L.
    Wilson, Kevin E.
    Schoeller, Dale A.
    Heymsfield, Steven B.
    Shepherd, John A.
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 2018, 108 (04): : 708 - 715
  • [35] Differences in air displacement plethysmography, bioelectrical impedance analysis and dual-energy X-ray absorptiometry for estimating body composition in Chinese children and adolescents
    Huang, Yiwen
    Wang, Xi
    Cheng, Hong
    Dong, Hongbo
    Shan, Xinying
    Zhao, Xiaoyuan
    Wang, Xia
    Xie, Xianghui
    Mi, Jie
    JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2023, 59 (03) : 470 - 479
  • [36] Body composition analysis of pigs by dual-energy x-ray absorptiometry
    Mitchell, AD
    Conway, JM
    Potts, WJE
    JOURNAL OF ANIMAL SCIENCE, 1996, 74 (11) : 2663 - 2671
  • [37] MEASUREMENT OF BODY-COMPOSITION IN THE ELDERLY - DUAL-ENERGY X-RAY ABSORPTIOMETRY, UNDERWATER WEIGHING, BIOELECTRICAL-IMPEDANCE ANALYSIS, AND ANTHROPOMETRY
    BRODOWICZ, GR
    MANSFIELD, RA
    MCCLUNG, MR
    ALTHOFF, SA
    GERONTOLOGY, 1994, 40 (06) : 332 - 339
  • [38] Body composition assessment by dual-energy X-ray absorptiometry (DXA)
    Andreoli, A.
    Scalzo, G.
    Masala, S.
    Tarantino, U.
    Guglielmi, G.
    RADIOLOGIA MEDICA, 2009, 114 (02): : 286 - 300
  • [39] Assessment of body composition in Indian adults: comparison between dual-energy X-ray absorptiometry and isotope dilution technique
    Kulkarni, Bharati
    Kuper, Hannah
    Taylor, Amy
    Wells, Jonathan C.
    Radhakrishna, K. V.
    Kinra, Sanjay
    Ben-Shlomo, Yoav
    Smith, George Davey
    Ebrahim, Shah
    Kurpad, A. V.
    Byrne, Nuala M.
    Hills, Andrew P.
    BRITISH JOURNAL OF NUTRITION, 2014, 112 (07) : 1147 - 1153
  • [40] Comparison of Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry for Estimating Bone Mineral Content
    Stone, Tori M.
    Wingo, Jonathan E.
    Nickerson, Brett S.
    Esco, Michael R.
    INTERNATIONAL JOURNAL OF SPORT NUTRITION AND EXERCISE METABOLISM, 2018, 28 (05) : 542 - 546