Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements

被引:120
作者
Ascenti, Giorgio [1 ]
Mileto, Achille [1 ]
Krauss, Bernhard [2 ]
Gaeta, Michele [1 ]
Blandino, Alfredo [1 ]
Scribano, Emanuele [1 ]
Settineri, Nicola [1 ]
Mazziotti, Silvio [1 ]
机构
[1] Policlin G Martino, Dipartimento Sci Radiol, I-98125 Messina, Italy
[2] Siemens AG, Healthcare Sect, Forchheim, Germany
关键词
Dual-energy CT; Renal masses; Standard enhancement measurements; Iodine quantification; Diagnostic accuracy; CYST PSEUDOENHANCEMENT; PHANTOM; LESIONS; BENIGN; SINGLE;
D O I
10.1007/s00330-013-2811-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare the diagnostic accuracy of iodine quantification and standard enhancement measurements in distinguishing enhancing from nonenhancing renal masses. The Institutional Review Board approved this retrospective study conducted from data found in institutional patient databases and archives. Seventy-two renal masses were characterised as enhancing or nonenhancing using standard enhancement measurements (in HU) and iodine quantification (in mg/ml). Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of standard enhancement measurements and iodine quantification were calculated from chi (2) tests of contingency with histopathology or imaging follow-up as the reference standard. Difference in accuracy was assessed by means of McNemar analysis. Sensitivity, specificity, PPV, NPV and diagnostic accuracy for standard enhancement measurements and iodine quantification were 77.7 %, 100 %, 100 %, 81.8 %, 89 % and 100 %, 94.4 %, 94.7, 100 % and 97 %, respectively. The McNemar analysis showed that the accuracy of iodine quantification was significantly better (P < 0.001) than that of standard enhancement measurements. Compared with standard enhancement measurements, whole-tumour iodine quantification is more accurate in distinguishing enhancing from nonenhancing renal masses.
引用
收藏
页码:2288 / 2295
页数:8
相关论文
共 32 条
[1]   Pseudoenhancement of simulated renal cysts in a phantom using different multidetector CT scanners [J].
Abdulla, C ;
Kalra, MK ;
Saini, S ;
Maher, MM ;
Ahmad, A ;
Halpern, E ;
Silverman, SG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (06) :1473-1476
[2]   Single-phase dual-energy CT urography in the evaluation of haematuria [J].
Ascenti, G. ;
Mileto, A. ;
Gaeta, M. ;
Blandino, A. ;
Mazziotti, S. ;
Scribano, E. .
CLINICAL RADIOLOGY, 2013, 68 (02) :E87-E94
[3]   Dual-Source Dual-Energy CT Evaluation of Complex Cystic Renal Masses [J].
Ascenti, Giorgio ;
Mazziotti, Silvio ;
Mileto, Achille ;
Racchiusa, Sergio ;
Donato, Rocco ;
Settineri, Nicola ;
Gaeta, Michele .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) :1026-1034
[4]   Dual-Energy CT for Detection of Endoleaks After Endovascular Abdominal Aneurysm Repair: Usefulness of Colored Iodine Overlay [J].
Ascenti, Giorgio ;
Mazziotti, Silvio ;
Lamberto, Salvatore ;
Bottari, Antonio ;
Caloggero, Simona ;
Racchiusa, Sergio ;
Mileto, Achille ;
Scribano, Emanuele .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) :1408-1414
[5]   Multiphasic renal CT: Comparison of renal mass enhancement during the corticomedullary and nephrographic phases [J].
Birnbaum, BA ;
Jacobs, JE ;
Ramchandani, P .
RADIOLOGY, 1996, 200 (03) :753-758
[6]   Renal cyst pseudoenhancement: Evaluation with an anthropomorphic body CT phantom [J].
Birnbaum, BA ;
Maki, DD ;
Chakraborty, DP ;
Jacobs, JE ;
Babb, JS .
RADIOLOGY, 2002, 225 (01) :83-90
[7]   Renal cyst pseudoenhancement: Influence of multidetector CT reconstruction algorithm and scanner type in phantom model [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 244 (03) :767-775
[8]   Multi-detector row CT attenuation measurements: Assessment of intra- and interscanner variability with an anthropomorphic body CT phantom [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 242 (01) :109-119
[9]   Problems in the detection and characterization of small renal masses [J].
Bosniak, MA ;
Rofsky, NM .
RADIOLOGY, 1996, 200 (01) :286-287
[10]   Dual-energy CT iodine overlay technique for characterization of renal masses as cyst or solid: a phantom feasibility study [J].
Brown, C. L. ;
Hartman, R. P. ;
Dzyubak, O. P. ;
Takahashi, N. ;
Kawashima, A. ;
McCollough, C. H. ;
Bruesewitz, M. R. ;
Primak, A. M. ;
Fletcher, J. G. .
EUROPEAN RADIOLOGY, 2009, 19 (05) :1289-1295