SiPM-based dual-ended-readout DOI-TOF PET module based on mean-time method

被引:15
|
作者
Han, Y. B. [1 ]
Kang, H. G. [1 ,2 ]
Song, S. H. [1 ]
Ko, G. B. [3 ]
Lee, J. S. [3 ]
Hong, S. J. [1 ,4 ]
机构
[1] Eulji Univ, Dept Senior Healthcare, Grad Sch, 771 Gyeryong Ro, Daejeon 34824, South Korea
[2] Natl Inst Radiol Sci, Dept Radiat Measurement & Dose Assessment, Inage Ku, 4-9-1 Anagawa, Chiba 2638555, Japan
[3] Seoul Natl Univ, Coll Med, Dept Nucl Med, 101 Daehak Ro, Seoul 03080, South Korea
[4] Eulji Univ, Dept Radiol Sci, 553 Sanseong Daero, Seongnam Si 13135, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
Gamma camera; SPECT; PET PET/CT; coronary CT angiography (CTA); Scintigraphy and whole-body imaging; OF-FLIGHT; DETECTOR; DEPTH; PERFORMANCE; RESOLUTION;
D O I
10.1088/1748-0221/14/02/P02023
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Positron emission tomography (PET) with high resolution and high sensitivity is desirable for detecting cancers and neurological diseases. In this work, a depth-of-interaction (DOI)-time-of-flight (TOF) PET has been attempted to achieve both high spatial and high timing resolution. Dual-ended readout is a simple technique that can provide excellent timing and DOI resolutions and consistent signal arrival times, regardless of the DOI position along the scintillation crystal, as a mean-time method is used. A dual-ended readout DOI-TOF PET module consisting of a 6 x 6 array of 2 x 2 x 20 mm(3) saw-cut cerium-doped lutetium-yttrium oxyorthosilicate (LYSO) crystals is constructed. Both ends of the LYSO crystal array are optically coupled to a multi-pixel photon counter (MPPC) with 4 x 4 channels. The sixteen MPPC outputs are reduced to four position signals using a charge division circuit (CDC) board, and the timing signal is extracted from the common cathode of the MPPC. The four position signals from the MPPC are digitized by a DRS4-based high-speed waveform digitizer with a sampling rate of 5 GSa/s. A Na-22 source is placed in front of a reference detector and at the side of the dual-ended readout DOI-TOF PET module in five steps of 2 mm, 6 mm, 10 mm, 14 mm, and 18 mm to measure the DOI, coincidence timing resolutions (CTRs), and mean-times. The full-width-half-maximums (FWHMs) of DOI resolutions and CTRs varied from 3.0 mm to 3.8 mm, with an average of 3.5 mm, and from 333 ps to 367 ps, with an average of 349 ps, respectively. The average of the slopes of the mean-time versus DOI position, for the 36 crystals, was -0.60 +/- 1.68 ps/nm, which was consistent with the null value. The dual-ended readout DOI-TOF PET module based on the mean-time method produced both good DOI and CTRs, and consistent signal arrival times. The found solution would be the most advantageous in the small aperture PET systems, such as those for brain and breast imaging.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Crystal surface and reflector optimization for the SiPM-based dual-ended readout TOF-DOI PET detector
    Kang, Han Gyu
    Yamaya, Taiga
    Han, Young Been
    Song, Seong Hyun
    Ko, Guen Bae
    Lee, Jae Sung
    Hong, Seong Jong
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (06)
  • [2] A Dual-Ended Readout Detector Using a Meantime Method for SiPM TOF-DOI PET
    Kang, Han Gyu
    Ko, Guen Bae
    Rhee, June Tak
    Kim, Kyeong Min
    Lee, Jae Sung
    Hong, Seong Jong
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 1935 - 1943
  • [3] A systematic study on the strip-line readout method for SiPM-based TOF PET
    Kim, Heejong
    Hua, Yuexuan
    Lyu, Xudong
    Xu, Feng
    Wang, Yunbo
    Lyu, Jine
    Eclov, Neville
    Chen, Chin-Tu
    Xie, Qingguo
    Kao, Chien-Min
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [4] SiPM-based PET Module with Depth of Interaction
    Pennazio, F.
    Ambrosi, G.
    Bisogni, M. G.
    Cerello, P.
    Corsi, F.
    Del Guerra, A.
    Ionica, M.
    Marino, N.
    Marzocca, C.
    Morrocchi, M.
    Peroni, C.
    Pirrone, G.
    Santoni, C.
    Wheadon, R.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3786 - 3789
  • [5] A SiPM-based TOF-PET detector with high speed digital DRS4 readout
    Ronzhin, A.
    Albrow, M.
    Los, S.
    Martens, M.
    Murat, P.
    Ramberg, E.
    Kim, H.
    Chen, C. -T.
    Kao, C. -M.
    Niessen, K.
    Zatserklyaniy, A.
    Mazzillo, M.
    Carbone, B.
    Condorelli, G.
    Fallica, G.
    Piana, A.
    Sanfilippo, D.
    Valvo, G.
    Ritt, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 703 : 109 - 113
  • [6] Simulations of the 4DMPET SiPM-Based PET Module
    Pennazio, Francesco
    Barrio, John
    Bisogni, Maria Giuseppina
    Cerello, Piergiorgio
    De Luca, Giulia
    Del Guerra, Alberto
    Lacasta, Carlos
    Llosa, Gabriela
    Magazzu, Guido
    Moehrs, Sascha
    Peroni, Cristiana
    Wheadon, Richard
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2316 - 2320
  • [7] SiPM-based PET detector module for a 4π span scanner
    D-Perez-Benito
    Chil, R.
    Udias, J. M.
    Desco, M.
    Vaquero, J. J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 936 : 18 - 21
  • [8] Ultra Precise Timing with SiPM-Based TOF PET Scintillation Detectors
    Seifert, Stefan
    Vinke, Ruud
    van Dam, Herman T.
    Loehner, Herbert
    Dendooven, Peter
    Beekman, Freek J.
    Schaart, Dennis R.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2329 - +
  • [9] A High Bandwidth Preamplifier for SiPM-Based TOF PET Scintillation Detectors
    Seifert, Stefan
    Schaart, Dennis R.
    van Dam, Herman T.
    Huizenga, Jan
    Vinke, Ruud
    Dendooven, Peter
    Lohner, Herbert
    Beekman, Freek J.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 891 - +
  • [10] Development of capacitive multiplexing circuit for SiPM-based time-of-flight (TOF) PET detector
    Choe, Hyeok-Jun
    Choi, Yong
    Hu, Wei
    Yan, Jianhua
    Jung, Jin Ho
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (07): : N120 - N133