Validation of velocity map imaging conditions over larger areas

被引:15
|
作者
Reid, Mike [1 ,2 ,3 ]
Koehler, Sven P. K. [1 ,2 ,3 ]
机构
[1] Univ Manchester, Sch Chem, Manchester M13 9PL, Lancs, England
[2] Univ Manchester, Photon Sci Inst, Manchester M13 9PL, Lancs, England
[3] Univ Manchester, Dalton Cumbrian Facil, Moor Row CA24 3HA, Whitehaven, England
关键词
DYNAMICS; PHOTODISSOCIATION; PHOTOELECTRON; SPECTROSCOPY; PHOTOLYSIS; ELECTRONS; IONS; O-2; NM;
D O I
10.1063/1.4798646
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have established through simulations and experiments the area over which Velocity Map Imaging (VMI) conditions prevail. We designed a VMI setup in which we can vary the ionization position perpendicular to the center axis of the time-of-flight spectrometer. We show that weak extraction conditions are far superior over standard three-plate setups if the aim is to increase the ionization volume without distorting VMI conditions. This is important for a number of crossed molecular beam experiments that already utilize weak extraction conditions, but to a greater extent for surface studies where fragments are desorbed or scattered off a surface in all directions. Our results on the dissociation of NO2 at 226 nm show that ionization of the fragments can occur up to +/- 5.5 mm away from the center axis of the time-of-flight spectrometer without affecting resolution or arrival position. (C) 2013 American Institute of Physics. [http://dx.doi.org/10.1063/1.4798646]
引用
收藏
页数:6
相关论文
共 50 条
  • [1] A large aperture magnification lens for velocity map imaging
    Zhang, Yongwei
    Yang, Chung-Hsin
    Wu, Shiou-Min
    van Roij, Andre
    van der Zande, Wim J.
    Parker, David H.
    Yang, Xueming
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (01)
  • [2] Velocity-map imaging at low extraction fields
    Horke, Daniel A.
    Roberts, Gareth M.
    Lecointre, Julien
    Verlet, Jan R. R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (06)
  • [3] Collinear Velocity-map Photoelectron Imaging Spectrometer for Cluster Anions
    Wu, Xia
    Qin, Zheng-bo
    Xie, Hua
    Wu, Xiao-hu
    Cong, Ran
    Tang, Zi-chao
    CHINESE JOURNAL OF CHEMICAL PHYSICS, 2010, 23 (04) : 373 - 380
  • [4] Time-stretched multi-hit 3D velocity map imaging of photoelectrons
    Goudreau, E. Scott
    Boguslavskiy, Andrey E.
    Moffatt, Douglas J.
    Makhija, Varun
    Hemsworth, Michael
    Lausten, Rune
    Marceau, Claude
    Wilkinson, Iain
    Stolow, Albert
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2023, 94 (06)
  • [5] A velocity map imaging study of the photodissociation of the (A)over-tilde state of ammonia
    Rodriguez, Javier D.
    Gonzalez, Marta G.
    Rubio-Lago, Luis
    Banares, Luis
    PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 2014, 16 (02) : 406 - 413
  • [6] Near-ambient pressure velocity map imaging
    Chien, Tzu-En
    Hohmann, Lea
    Harding, Dan J.
    JOURNAL OF CHEMICAL PHYSICS, 2022, 157 (03)
  • [7] DAVIS: A direct algorithm for velocity-map imaging system
    Harrison, G. R.
    Vaughan, J. C.
    Hidle, B.
    Laurent, G. M.
    JOURNAL OF CHEMICAL PHYSICS, 2018, 148 (19)
  • [8] Time-resolved velocity map ion imaging study of NH3 photodissociation
    Wells, Kym L.
    Perriam, Gareth
    Stavros, Vasilios G.
    JOURNAL OF CHEMICAL PHYSICS, 2009, 130 (07)
  • [9] Velocity map photoelectron-photoion coincidence imaging on a single detector
    Lehmann, C. Stefan
    Ram, N. Bhargava
    Janssen, Maurice H. M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (09)
  • [10] Account: An introduction to velocity-map imaging mass spectrometry (VMImMS)
    Bull, James N.
    Lee, Jason W. L.
    Gardiner, Sara H.
    Valiance, Claire
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2014, 20 (02) : 117 - 129