共 50 条
Validation of velocity map imaging conditions over larger areas
被引:15
|作者:
Reid, Mike
[1
,2
,3
]
Koehler, Sven P. K.
[1
,2
,3
]
机构:
[1] Univ Manchester, Sch Chem, Manchester M13 9PL, Lancs, England
[2] Univ Manchester, Photon Sci Inst, Manchester M13 9PL, Lancs, England
[3] Univ Manchester, Dalton Cumbrian Facil, Moor Row CA24 3HA, Whitehaven, England
关键词:
DYNAMICS;
PHOTODISSOCIATION;
PHOTOELECTRON;
SPECTROSCOPY;
PHOTOLYSIS;
ELECTRONS;
IONS;
O-2;
NM;
D O I:
10.1063/1.4798646
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
We have established through simulations and experiments the area over which Velocity Map Imaging (VMI) conditions prevail. We designed a VMI setup in which we can vary the ionization position perpendicular to the center axis of the time-of-flight spectrometer. We show that weak extraction conditions are far superior over standard three-plate setups if the aim is to increase the ionization volume without distorting VMI conditions. This is important for a number of crossed molecular beam experiments that already utilize weak extraction conditions, but to a greater extent for surface studies where fragments are desorbed or scattered off a surface in all directions. Our results on the dissociation of NO2 at 226 nm show that ionization of the fragments can occur up to +/- 5.5 mm away from the center axis of the time-of-flight spectrometer without affecting resolution or arrival position. (C) 2013 American Institute of Physics. [http://dx.doi.org/10.1063/1.4798646]