Comparison of the x-ray attenuation properties of breast calcifications, aluminium, hydroxyapatite and calcium oxalate

被引:32
|
作者
Warren, L. M. [1 ,2 ]
Mackenzie, A. [1 ,2 ]
Dance, D. R. [1 ,2 ]
Young, K. C. [1 ,2 ]
机构
[1] Royal Surrey Cty Hosp NHS Fdn Trust, Natl Coordinating Ctr Phys Mammog, Guildford GU2 7XX, Surrey, England
[2] Univ Surrey, Fac Engn & Phys Sci, Dept Phys, Guildford GU2 7XH, Surrey, England
基金
英国工程与自然科学研究理事会;
关键词
MICRO-CALCIFICATIONS; DIGITAL MAMMOGRAPHY; MICROCALCIFICATIONS; SIMULATION; FIELD; CONTRAST; VISIBILITY; LESIONS;
D O I
10.1088/0031-9155/58/7/N103
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aluminium is often used as a substitute material for calcifications in phantom measurements inmammography. Additionally, calcium oxalate, hydroxyapatite and aluminium are used in simulation studies. This assumes that these materials have similar attenuation properties to calcification, and this assumption is examined in this work. Sliced mastectomy samples containing calcification were imaged at x5 magnification using a digital specimen cabinet. Images of the individual calcifications were extracted, and the diameter and contrast of each calculated. The thicknesses of aluminium required to achieve the same contrast as each calcification when imaged under the same conditions were calculated using measurements of the contrast of aluminium foils. As hydroxyapatite and calcium oxalate are also used to simulate calcifications, the equivalent aluminium thicknesses of these materials were also calculated using tabulated attenuation coefficients. On average the equivalent aluminium thickness was 0.85 times the calcification diameter. For calcium oxalate and hydroxyapatite, the equivalent aluminium thicknesses were 1.01 and 2.19 times the thickness of these materials respectively. Aluminium and calcium oxalate are suitable substitute materials for calcifications. Hydroxyapatite is much more attenuating than the calcifications and aluminium. Using solid hydroxyapatite as a substitute for calcification of the same size would lead to excessive contrast in the mammographic image.
引用
收藏
页码:N103 / N113
页数:11
相关论文
共 50 条
  • [31] Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography
    Delogu, Pasquale
    Di Trapani, Vittorio
    Brombal, Luca
    Mettivier, Giovanni
    Taibi, Angelo
    Oliva, Piernicola
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [32] Deep learning for x-ray scatter correction in dedicated breast CT
    Pautasso, Juan J.
    Caballo, Marco
    Mikerov, Mikhail
    Boone, John M.
    Michielsen, Koen
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2023, 50 (04) : 2022 - 2036
  • [33] X-ray scatter correction method for dedicated breast computed tomography
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2012, 39 (05) : 2896 - 2903
  • [34] Generation and analysis of clinically relevant breast imaging x-ray spectra
    Hernandez, Andrew M.
    Seibert, J. Anthony
    Nosratieh, Anita
    Boone, John M.
    MEDICAL PHYSICS, 2017, 44 (06) : 2148 - 2160
  • [35] X-Ray diffraction analysis of 99.1% recycled aluminium subjected to equal channel angular extrusion
    Rebhi, A.
    Makhlouf, T.
    Njah, N.
    PROCEEDINGS OF THE JMSM 2008 CONFERENCE, 2009, 2 (03): : 1263 - 1270
  • [36] High-Resolution X-Ray Phase-Contrast 3-D Imaging of Breast Tissue Specimens as a Possible Adjunct to Histopathology
    Baran, Patrycja
    Mayo, Sheridan
    McCormack, Mikkaela
    Pacile, Serena
    Tromba, Giuliana
    Dullin, Christian
    Zanconati, Fabrizio
    Arfelli, Fulvia
    Dreossi, Diego
    Fox, Jane
    Prodanovic, Zdenka
    Cholewa, Marian
    Quiney, Harry
    Dimmock, Matthew
    Nesterets, Yakov
    Thompson, Darren
    Brennan, Patrick
    Gureyev, Timur
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (12) : 2642 - 2650
  • [37] Resolution and noise properties of scintillator coated X-ray detectors
    Dubaric, E
    Fröjdh, C
    Nilsson, HE
    Petersson, CS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 466 (01) : 178 - 182
  • [38] Attenuation correction of polychromatic L-shell X-ray fluorescence computed tomography imaging
    Liu, Long
    Zhou, Xiaolin
    Liu, Huanlong
    Ding, Ning
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2019, 56 (05) : 376 - 384
  • [39] Visibility of microcalcification in cone beam breast CT: Effects of x-ray tube voltage and radiation dose
    Lai, Chao-Jen
    Shaw, Chris C.
    Chen, Lingyun
    Altunbas, Mustafa C.
    Liu, Xinming
    Han, Tao
    Wang, Tianpeng
    Yang, Wei T.
    Whitman, Gary J.
    Tu, Shu-Ju
    MEDICAL PHYSICS, 2007, 34 (07) : 2995 - 3004
  • [40] Attenuation correction for x-ray fluorescence computed tomography (XFCT) utilizing transmission CT image
    Zhang, Siyuan
    Li, Liang
    Chen, Zhiqiang
    15TH INTERNATIONAL MEETING ON FULLY THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 2019, 11072