Comparison of the x-ray attenuation properties of breast calcifications, aluminium, hydroxyapatite and calcium oxalate

被引:32
|
作者
Warren, L. M. [1 ,2 ]
Mackenzie, A. [1 ,2 ]
Dance, D. R. [1 ,2 ]
Young, K. C. [1 ,2 ]
机构
[1] Royal Surrey Cty Hosp NHS Fdn Trust, Natl Coordinating Ctr Phys Mammog, Guildford GU2 7XX, Surrey, England
[2] Univ Surrey, Fac Engn & Phys Sci, Dept Phys, Guildford GU2 7XH, Surrey, England
基金
英国工程与自然科学研究理事会;
关键词
MICRO-CALCIFICATIONS; DIGITAL MAMMOGRAPHY; MICROCALCIFICATIONS; SIMULATION; FIELD; CONTRAST; VISIBILITY; LESIONS;
D O I
10.1088/0031-9155/58/7/N103
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aluminium is often used as a substitute material for calcifications in phantom measurements inmammography. Additionally, calcium oxalate, hydroxyapatite and aluminium are used in simulation studies. This assumes that these materials have similar attenuation properties to calcification, and this assumption is examined in this work. Sliced mastectomy samples containing calcification were imaged at x5 magnification using a digital specimen cabinet. Images of the individual calcifications were extracted, and the diameter and contrast of each calculated. The thicknesses of aluminium required to achieve the same contrast as each calcification when imaged under the same conditions were calculated using measurements of the contrast of aluminium foils. As hydroxyapatite and calcium oxalate are also used to simulate calcifications, the equivalent aluminium thicknesses of these materials were also calculated using tabulated attenuation coefficients. On average the equivalent aluminium thickness was 0.85 times the calcification diameter. For calcium oxalate and hydroxyapatite, the equivalent aluminium thicknesses were 1.01 and 2.19 times the thickness of these materials respectively. Aluminium and calcium oxalate are suitable substitute materials for calcifications. Hydroxyapatite is much more attenuating than the calcifications and aluminium. Using solid hydroxyapatite as a substitute for calcification of the same size would lead to excessive contrast in the mammographic image.
引用
收藏
页码:N103 / N113
页数:11
相关论文
共 50 条
  • [1] CNR performance of semiconductor materials for X-ray imaging of breast calcifications
    Mendoza, L.
    Avila, C.
    Rodriguez, R.
    Loaiza, L.
    Roque, G.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (11)
  • [2] Classification of the Micromorphology of Breast Calcifications in X-ray Dark-Field Mammography
    Willer, Konstantin
    Scherer, Kai
    Braig, Eva
    Ehn, Sebastian
    Schock, Jonathan
    Wolf, Johannes
    Birnbacher, Lorenz
    Chabior, Michael
    Mayr, Doris
    Grandl, Susanne
    Sztrokay-Gaul, Aniko
    Hellerhoff, Karin
    Reiser, Maximilian
    Pfeiffer, Franz
    Herzen, Julia
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [3] Discrimination analysis of breast calcifications using x-ray dark-field radiography
    Rauch, Thomas
    Rieger, Jens
    Pelzer, Georg
    Horn, Florian
    Erber, Ramona
    Wunderle, Marius
    Emons, Julius
    Nabieva, Naiba
    Fuhrich, Nicole
    Michel, Thilo
    Hartmann, Arndt
    Fasching, Peter A.
    Anton, Gisela
    MEDICAL PHYSICS, 2020, 47 (04) : 1813 - 1826
  • [4] X-ray properties of an anthropomorphic breast phantom for MRI and x-ray imaging
    Freed, Melanie
    Badal, Andreu
    Jennings, Robert J.
    de las Heras, Hugo
    Myers, Kyle J.
    Badano, Aldo
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (12) : 3513 - 3533
  • [5] Characterization of breast calcification types using dual energy x-ray method
    Martini, N.
    Koukou, V.
    Fountos, G.
    Michail, C.
    Bakas, A.
    Kandarakis, I.
    Speller, R.
    Nikiforidis, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (19) : 7741 - 7764
  • [6] Estimation of effective x-ray tissue attenuation differences for volumetric breast density measurement
    Chen, Biao
    Ruth, Chris
    Jing, Zhenxue
    Ren, Baorui
    Smith, Andrew
    Kshirsagar, Ashwini
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [7] Measurement of breast-tissue x-ray attenuation by spectral mammography: solid lesions
    Fredenberg, Erik
    Kilburn-Toppin, Fleur
    Willsher, Paula
    Moa, Elin
    Danielsson, Mats
    Dance, David R.
    Young, Kenneth C.
    Wallis, Matthew G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (07) : 2595 - 2612
  • [8] Improved Diagnostics by Assessing the Micromorphology of Breast Calcifications via X-Ray Dark-Field Radiography
    Scherer, Kai
    Braig, Eva
    Ehn, Sebastian
    Schock, Jonathan
    Wolf, Johannes
    Birnbacher, Lorenz
    Chabior, Michael
    Herzen, Julia
    Mayr, Doris
    Grandl, Susanne
    Sztrokay-Gaul, Aniko
    Hellerhoff, Karin
    Pfeiffer, Franz
    SCIENTIFIC REPORTS, 2016, 6
  • [9] Measurement of breast-tissue x-ray attenuation by spectral mammography: first results on cyst fluid
    Fredenberg, Erik
    Dance, David R.
    Willsher, Paula
    Moa, Elin
    von Tiedemann, Miriam
    Young, Kenneth C.
    Wallis, Matthew G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (24) : 8609 - 8620
  • [10] Measurement of breast-tissue x-ray attenuation by spectral mammography: First results on cyst fluid
    Fredenberg, Erik
    Dance, David R.
    Willsher, Paula
    von Tiedemann, Miriam
    Young, Kenneth C.
    Wallis, Matthew G.
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668