Dosimetry comparison between TG-43 and Monte Carlo calculations using the Freiburg flap for skin high-dose-rate brachytherapy

被引:18
|
作者
Vijande, Javier [1 ,2 ]
Ballester, Facundo [1 ]
Ouhib, Zoubir [3 ]
Granero, Domingo [4 ]
Carmen Pujades-Claumarchirant, M. [5 ]
Perez-Calatayud, Jose [5 ]
机构
[1] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[2] Univ Valencia, CSIC, IFIC, Paterna, Valencia, Spain
[3] Boca Raton Reg Hosp, Lynn Canc Inst, Boca Raton, FL USA
[4] Hosp Gen Univ, ERESA, Dept Radiat Phys, Valencia, Spain
[5] La Fe Univ Hosp, Dept Radiat Oncol, Valencia, Spain
关键词
Ir-192; Brachytherapy; Dosimetry; Penelope2008; Freiburg flap; PARAMETERS; IR-192; APPLICATORS;
D O I
10.1016/j.brachy.2011.11.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of this work was to evaluate whether the delivered dose to the skin surface and at the prescription depth when using a Freiburg flap applicator is in agreement with the one predicted by the treatment planning system (TPS) using the TG-43 dose-calculation formalism. METHODS AND MATERIALS: Monte Carlo (MC) simulations and radiochromic film measurements have been performed to obtain dose distributions with the source located at the center of one of the spheres and between two spheres. Primary and scatter dose contributions were evaluated to understand the role played by the scatter component. A standard treatment plan was generated using MC- and TG-43-based TPS applying the superposition principle. RESULTS: The MC model has been validated by performing additional simulations in the same conditions but transforming air and Freiburg flap materials into water to match TG-43 parameters. Both dose distributions differ less than 1%. Scatter defect compared with TG-43 data is up to 15% when the source is located at the center of the sphere and up to 25% when the source is between two spheres. Maximum deviations between TPS- and MC-based distributions are of 5%. CONCLUSIONS: The deviations in the TG-43-based dose distributions for a standard treatment plan with respect to the MC dose distribution calculated taking into account the composition and shape of the applicator and the surrounding air are lower than 5%. Therefore, this study supports the validity of the TPS used in clinical practice. (C) 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:528 / 535
页数:8
相关论文
共 50 条
  • [21] Monte Carlo characterization of 169Yb as a high-dose-rate source for brachytherapy application by FLUKA code
    Anjomrouz, Marzieh
    Sadeghi, Mahdi
    Haddadi, Asghar
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 196 - 205
  • [22] Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    Daskalov, GM
    Loffler, E
    Williamson, JF
    MEDICAL PHYSICS, 1998, 25 (11) : 2200 - 2208
  • [23] Monte Carlo Evaluation for the Effect of Positional Inaccuracy of Source on Patient's Dose in High-Dose-Rate Brachytherapy
    Lee, Kyung Ja
    An, So Hyun
    Lee, Rena
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (06) : 2361 - 2365
  • [24] Comparison of high-dose-rate intracavitary brachytherapy dosimetry with and without anesthesia in patients with cervical carcinoma
    Sharma, Daya N.
    Chaudhari, Pritee
    Sharma, Seema
    Gupta, Leena
    Jagadesan, Pandjatcharam
    Rath, Goura K.
    Julka, Pramod K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 60 - 66
  • [25] A Monte Carlo Dosimetric Parameters of the 60Co High Dose Rate Brachytherapy and Investigation of TG43 Dose Accuracy in Different Media Using GATE v8.2 Code
    Ait-Mlouk, Lahcen
    Khalis, Mohammed
    Ouabi, Hmad
    Asnaoui, Hassan
    Elboukhari, Said
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (SUPPL 1) : S68 - S79
  • [26] Application of Monte Carlo calculations for validation of a treatment planning system in high dose rate brachytherapy
    Naseri, Alireza
    Mesbahi, Asghar
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2009, 14 (06) : 200 - 204
  • [27] Monte Carlo simulated correction factors for high dose rate brachytherapy postal dosimetry audit methodology
    Chelminski, Krzysztof
    Dimitriadis, Alexis
    Abdulrahim, Roua
    Kazantsev, Pavel
    Granizo-Roman, Evelyn
    Kalinowski, Jonathan
    Enger, Shirin Abbasi
    Azangwe, Godfrey
    Carrara, Mauro
    Swamidas, Jamema
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 32
  • [28] Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
    Zehtabian, M.
    Faghihi, R.
    Sina, S.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2010, 7 (04): : 217 - 222
  • [29] A Monte Carlo Dosimetric Parameters of the 60Co High Dose Rate Brachytherapy and Investigation of TG43 Dose Accuracy in Different Media Using GATE v8.2 Code
    Lahcen Ait-Mlouk
    Mohammed Khalis
    Hmad Ouabi
    Hassan Asnaoui
    Said Elboukhari
    Moscow University Physics Bulletin, 2021, 76 : S68 - S79
  • [30] In vivo Measurement of Bladder and Rectum Doses of High-Dose-Rate Brachytherapy in Gynaecologic Tumors and Their Comparison with the Computer Calculations
    Algul, Emriye
    Ozbay, Ismail
    Aslay, Isik
    Ozturk, Nural
    TURKIYE KLINIKLERI TIP BILIMLERI DERGISI, 2010, 30 (01): : 263 - 269