Platelet storage solution effects on the accuracy of laboratory tests for platelet function: a multi-laboratory study

被引:66
作者
VandenBroeke, T
Dumont, LJ
Hunter, S
Nixon, J
Murphy, S
Roger, J
Herschel, L
AuBuchon, JP
Gulliksson, H
Dengler, T
Hornsey, V
Prowse, C
机构
[1] GambroBCT Inc, Lakewood, CO 80215 USA
[2] Amer Red Cross, Penn Jersey Reg, Philadelphia, PA USA
[3] Dartmouth Hitchcock Med Ctr, Lebanon, NH 03766 USA
[4] Huddinge Univ Hosp, Stockholm, Sweden
[5] German Red Cross, Baden Baden, Germany
[6] Scottish Natl Blood Transfus Serv, Edinburgh, Midlothian, Scotland
关键词
diluent; extent of shape change; hypotonic shock response; in vitro platelet function;
D O I
10.1111/j.0042-9007.2004.00412.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives Extent of shape change (ESC) and hypotonic shock response (HSR) have been widely used to characterize the in vitro function of platelets and have been shown to correlate with in vivo viability. These assays have been routinely performed using platelet-poor plasma (PPP) as the test sample diluent. Because of the increasing popularity of storing platelets in synthetic media, it is important to understand the effects of using these synthetic media as test diluents for ESC and HSR measurements. The objective of this study was to determine the effect of using platelet storage solutions vs. plasma for the in vitro testing of ESC and HSR. Materials and Methods Six laboratories participated in this study. Platelets were prepared by apheresis, the platelet-rich plasma (PRP) method, or derived from buffy-coats. Each platelet preparation was divided, half being stored in plasma and the other half in storage solution. ESC and HSR testing were performed in duplicate on days 1 and 5, using each of three diluents: autologous plasma; fresh-frozen plasma; or storage solution. Results For both ESC and HSR, dilutions made in each of the three diluents yielded significantly different results. Dilutions made in storage solutions were more than 30% lower for ESC and HSR than those made in autologous plasma (P<0.0001). Dilutions made in thawed fresh-frozen plasma were more than 16% lower for ESC and HSR than those made in liquid autologous plasma (P<0.0005). Conclusions ESC and HSR test results are significantly affected by the test diluent. Platelets should be diluted in plasma (preferably autologous) for the in vitro testing of ESC and HSR, regardless of the media in which they are stored.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 15 条
[1]   WBC content of platelet concentrates prepared by the buffy coat method using different processing procedures and storage solutions [J].
de Wildt-Eggen, J ;
Schrijver, JG ;
Bins, M .
TRANSFUSION, 2001, 41 (11) :1378-1383
[2]   Platelet surface P-selectin measurements in platelet preparations: An international collaborative study [J].
Dumont, LJ ;
VandenBroeke, T ;
Ault, KA .
TRANSFUSION MEDICINE REVIEWS, 1999, 13 (01) :31-42
[3]   Seven-day storage of single-donor platelets: recovery and survival in an autologous transfusion study [J].
Dumont, LJ ;
AuBuchon, JP ;
Whitley, P ;
Herschel, LH ;
Johnson, A ;
McNeil, D ;
Sawyer, S ;
Roger, JC .
TRANSFUSION, 2002, 42 (07) :847-854
[4]  
Gorlin J, 2002, STANDARDS BLOOD BANK
[5]   Storage of platelets in additive solutions:: a pilot in vitro study of the effects of potassium and magnesium [J].
Gulliksson, H ;
AuBuchon, JP ;
Vesterinen, M ;
Sandgren, P ;
Larsson, S ;
Pickard, CA ;
Herschel, I ;
Roger, J ;
Tracy, JE ;
Langweiler, M .
VOX SANGUINIS, 2002, 82 (03) :131-136
[6]  
HANDIN R I, 1970, Transfusion (Bethesda), V10, P305
[7]   A multi-laboratory evaluation of in vitro platelet assays: the tests for extent of shape change and response to hypotonic shock [J].
Holme, S ;
Moroff, G ;
Murphy, S .
TRANSFUSION, 1998, 38 (01) :31-40
[8]  
HOLME S, 1978, BLOOD, V52, P425
[9]   PLATELET RESPONSE TO HYPOTONIC SHOCK - ITS VALUE AS AN INDICATOR OF PLATELET VIABILITY AFTER STORAGE [J].
KIM, BK ;
BALDINI, MG .
TRANSFUSION, 1974, 14 (02) :130-138
[10]   P2 receptor subtypes in the cardiovascular system [J].
Kunapuli, SP ;
Daniel, JL .
BIOCHEMICAL JOURNAL, 1998, 336 :513-523