A novel formulation for scratch-based wear modelling in total hip arthroplasty

被引:6
作者
Kruger, Karen M. [1 ,2 ]
Tikekar, Nishant M. [3 ]
Heiner, Anneliese D. [1 ,2 ]
Baer, Thomas E. [1 ]
Lannutti, John J. [3 ]
Callaghan, John J. [1 ,2 ]
Brown, Thomas D. [1 ,2 ]
机构
[1] Univ Iowa, Dept Orthopaed & Rehabil, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
[3] Ohio State Univ, Dept Mat Sci & Engn, Columbus, OH 43210 USA
关键词
wear; total hip arthroplasty; third body; finite element analysis; retrieval; MOLECULAR-WEIGHT POLYETHYLENE; CROSS-LINKED POLYETHYLENE; FINITE-ELEMENT-ANALYSIS; SURFACE DAMAGE; 3RD-BODY WEAR; FEMORAL HEADS; SIMULATOR; OSTEOLYSIS; FIXATION;
D O I
10.1080/10255842.2012.739168
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Damage to the femoral head in total hip arthroplasty often takes the form of discrete scratches, which can lead to dramatic wear acceleration of the polyethylene ( PE) liner. Here, a novel formulation is reported for finite element (FE) analysis of wear acceleration due to scratch damage. A diffused-light photography technique was used to globally locate areas of damage, providing guidance for usage of high-magnification optical profilometry to determine individual scratch morphology. This multiscale image combination allowed comprehensive input of scratch-based damage patterns to an FE Archard wear model, to determine the wear acceleration associated with specific retrieval femoral heads. The wear algorithm imposed correspondingly elevated wear factors on areas of PE incrementally overpassed by individual scratches. Physical validation was provided by agreement with experimental data for custom-ruled scratch patterns. Illustrative wear acceleration results are presented for four retrieval femoral heads.
引用
收藏
页码:1227 / 1236
页数:10
相关论文
共 32 条
[1]   CONTACT AND RUBBING OF FLAT SURFACES [J].
ARCHARD, JF .
JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) :981-988
[2]   A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) :569-576
[3]   Hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions [J].
Bowsher, JG ;
Shelton, JC .
WEAR, 2001, 250 :167-179
[4]   Surface Damage After Multiple Dislocations of a 38-mm-Diameter, Metal-on-Metal Hip Prosthesis [J].
Bowsher, John G. ;
Donaldson, Thomas K. ;
Williams, Paul A. ;
Clarke, Ian C. .
JOURNAL OF ARTHROPLASTY, 2008, 23 (07) :1090-1096
[5]   Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[6]   Local head roughening as a factor contributing to variability of total hip wear: A finite element analysis [J].
Brown, TD ;
Stewart, KJ ;
Nieman, JC ;
Pedersen, DR ;
Callaghan, JJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2002, 124 (06) :691-698
[7]   Clinical Biomechanics of Third Body Acceleration of Total Hip Wear [J].
Brown, Thomas D. ;
Lundberg, Hannah J. ;
Pedersen, Douglas R. ;
Callaghan, John J. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (07) :1885-1897
[9]  
D'Antonio J.A., 2011, CLIN ORTHOP RELAT R, V470, P1696
[10]   THE ROLE OF COUNTERFACE IMPERFECTIONS IN THE WEAR OF POLYETHYLENE [J].
DOWSON, D ;
TAHERI, S ;
WALLBRIDGE, NC .
WEAR, 1987, 119 (03) :277-293