Simultaneous Determination of Inorganic Anions and Cations in Water and Biological Samples by Capillary Electrophoresis with a Capacitive Coupled Contactless Conductivity Detector Using Capillary Filling Method

被引:16
|
作者
Yamamoto, Sachio [1 ]
Fujiwara, Haruka [1 ]
Maruyama, Katsushi [1 ]
Tanaka, Yoshihide [1 ]
Kinoshita, Mitsuhiro [1 ,2 ]
Suzuki, Shigeo [1 ,2 ]
机构
[1] Kindai Univ, Fac Pharmaceut Sci, 3-4-1 Kowakae, Higashiosaka, Osaka 5778502, Japan
[2] Kindai Univ, Antiaging Ctr, 3-4-1 Kowakae, Higashiosaka, Osaka 5778502, Japan
基金
日本学术振兴会;
关键词
Capillary electrophoresis (CE); capacitively coupled contactless conductivity detector ((CD)-D-4); capillary filling method (CMF); water analysis; INJECTION-ANALYSIS MANIFOLD; EXHALED BREATH CONDENSATE; SIMULTANEOUS SEPARATION; STEP WIDTH; RESOLUTION;
D O I
10.2116/analsci.18P422
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical method for concurrent analysis of inorganic anions and cations has been developed using a capillary electrophoresis (CE)-capacitively coupled contactless conductivity detector ((CD)-D-4) system. Although hydrodynamic and electrokinetie injection techniques have been widely used in CE, we employed a capillary filling method (CFM) for the analysis of inorganic ions. The procedure is relatively simple and has the advantage that CMF does not require pressure control and vial exchange. Three anions (chloride, sulfate, nitrate) and five cations (ammonium, potassium, sodium, magnesium, calcium) were successfully separated and detected at ppm levels within 80 s using a 9 mM histidine/15 mM malic acid (pH 3.6) containing 50 mM N-dodecyl-N,N-dimethyl-3-ammonio-1-propanesulfonate as background electrolyte. Applying this analytical condition, the electroosmotic flow is negligible and anions and cations were migrated concurrently to different polarities according to their electrophoretic mobility. Obtained raw data showed stepwise increases in detected conductivity due to the migration of sample components, which expresses as peak profiles by differentiation of electropherograms. The RSD values of the peak area and migration times for the anions and cations were satisfactory and were less than 5.15 and 2.04%, respectively. The developed method was applied for the analysis of inorganic anions and cations in commercial mineral waters, tap water, urine, and exhaled breath condensate. These results indicate that the CE-(CD)-D-4 system with CFM is suitable for the rapid analysis of inorganic anions and cations in various samples.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 50 条
  • [1] Simultaneous Determination of Inorganic Anions and Cations in Water and Biological Samples by Capillary Electrophoresis with a Capacitive Coupled Contactless Conductivity Detector Using Capillary Filling Method
    Sachio Yamamoto
    Haruka Fujiwara
    Katsushi Maruyama
    Yoshihide Tanaka
    Mitsuhiro Kinoshita
    Shigeo Suzuki
    Analytical Sciences, 2019, 35 : 295 - 300
  • [2] Simultaneous separation of inorganic anions and cations using capillary electrophoresis with a movable contactless conductivity detector
    Unterholzner, V
    Macka, M
    Haddad, PR
    Zemann, A
    ANALYST, 2002, 127 (06) : 715 - 718
  • [3] Simultaneous determination of inorganic anions and cations by capillary electrophoresis
    Padarauskas, A
    REVIEWS IN ANALYTICAL CHEMISTRY, 2001, 20 (04): : 271 - 301
  • [4] Determination of inorganic cations in biological fluids using a hybrid capillary electrophoresis device coupled with contactless conductivity detection
    Lobo-Junior, Eulicio Oliveira
    Chagas, Cyro L. S.
    Coltro, Wendell K. T.
    JOURNAL OF SEPARATION SCIENCE, 2018, 41 (16) : 3310 - 3317
  • [5] Simultaneous determination of inorganic cations and anions by dual opposite ends injection coupled with capillary electrophoresis with capacitively contactless conductivity detection
    Tan, F
    Yang, BC
    Guan, YF
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (03) : 313 - 316
  • [6] Simultaneous determination of inorganic anions and cations in waters by capillary electrophoresis
    Padarauskas, A
    Olsauskaité, V
    Paliulionyté, V
    JOURNAL OF CHROMATOGRAPHY A, 1998, 829 (1-2) : 359 - 365
  • [7] Simultaneous Determination of Inorganic Anions and Cations in Cosmetics by Capillary Zone Electrophoresis
    Wang, Zong-Hua
    Yan, Liu-Shui
    Luo, Guo-An
    Kao Teng Hsueh Hsiao Hua Heush Hsueh Pao/ Chemical Journal of Chinese Universities, 2001, 22 (10 SUPPL.): : 40 - 41
  • [8] Simultaneous Determination of Inorganic Anions and Cations in Cosmetics by Capillary Zone Electrophoresis
    Wang Zong-Hua
    Yan Liu-Shui
    Luo Guo-An
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2001, 22 : 37 - 41
  • [9] Simultaneous determination of γ-hydroxybutyric acid, ibotenic acid and psilocybin in saliva samples by capillary electrophoresis coupled with a contactless conductivity detector
    Saar-Reismaa, P.
    Vaher, M.
    Kaljurand, M.
    Kulp, M.
    Mazina-Sinkar, J.
    ANALYTICAL METHODS, 2017, 9 (21) : 3128 - 3133
  • [10] Analysis of inorganic cations in biological samples by the combination of micro-electrodialysis and capillary electrophoresis with capacitively coupled contactless conductivity detection
    Thi Kieu Oanh Doan
    Kuban, Pavel
    Kuban, Petr
    Kiplagat, Isaac K.
    Bocek, Petr
    ELECTROPHORESIS, 2011, 32 (3-4) : 464 - 471