mine water analysis;
arsenic;
voltammetry;
in situ;
matrix removal;
chelating resin;
D O I:
暂无
中图分类号:
TD [矿业工程];
学科分类号:
0819 ;
摘要:
This study presents a matrix-removal system for metals that interfere with the determination of arsenic with a voltammetric instrument capable of on-site use. A three-step sample pre-treatment system was developed to remove Fe, Cu and Zn: filtration, adjustment,of pH (pH 5.5) and column-extraction of interfering metals using chelating resin. This set-up allowed the direct and accurate determination of As in mine waters containing up to 1500 mu g L-1 and 500 mu g L-1 Zn and confirmed interferences at higher concentrations. Column optimisation and method adaptation for rapid on-site use are discussed.