Multicentric evaluation of eight glucose and four ketone blood meters

被引:12
作者
Guimont, Marie-Christine [1 ]
Desjobert, Helene [2 ]
Fonfrede, Michele [3 ]
Vitoux, Dominique [4 ]
Benoist, Jean-Francois [5 ]
Launay, Jean-Marie [6 ]
Peoc'h, Katell [1 ,6 ]
Lefevre, Guillaume [7 ]
机构
[1] Hop Beaujon, AP HP, Serv Biochim Clin, F-92118 Clichy, France
[2] Hop Bichat Claude Bernard, AP HP, Serv Biochim Paris, F-75018 Clichy, France
[3] Hop La Pitie Salpetriere, AP HP, Serv Biochim, F-75013 Paris, France
[4] Hop St Louis, AP HP, Serv Biochim & Biol Mol, F-75010 Paris, France
[5] Hop Robert Debre, AP HP, Serv Biochim, F-75019 Paris, France
[6] Hop Lariboisiere, AP HP, Serv Biochim & Biol Mol, F-75010 Paris, France
[7] Hop Tenon, AP HP, Serv Biochim & Hormonol, F-75020 Paris, France
关键词
Point-of-care; Glucose meter; Ketone meter; ISO; 15197; norm; Diabetes; Glycemia; ACCURACY EVALUATION; MONITORING SYSTEMS; CARE; ERRORS;
D O I
10.1016/j.clinbiochem.2015.07.032
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: High precision meters for blood glycemia are mandatory for monitoring glucose status in patients, avoiding both hypo-and hyper-glycemia. Health care providers routinely used in both out-and inpatients point-of-care measurements of glucose and ketone. These measurements, frequently used for medical decisions, are known to be less accurate than those performed in laboratories. Our aim was to evaluate, within the frame of an Assistance Publique-Hopitaux de Paris (AP-HP) multicentric study, the performances of eight glucose and four ketone meters, either connected or non-connected to a laboratory software. Design and methods: Glucose meter accuracy, precision, correlation with plasma glucose determined in central laboratories and hematocrit interferences were determined according to the ISO 15197: 2003 norm. The same norm was applied for the determination of accuracy, precision and recovery of ketone meters for B-hydroxybutyrate measurements. Results and conclusion: Among those meters, seven were considered as acceptable for glucose measurement and two for ketone measurement. Since all meters do not fit clinically relevant criteria, meters' performances have to be evaluated before use in clinical practice. (C) 2015 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1310 / 1316
页数:7
相关论文
共 23 条
[1]  
[Anonymous], 15197 ISO
[2]  
[Anonymous], DIABETES CARE
[3]   How to use near-patient capillary ketone meters [J].
Bashford, James ;
Acerini, Carlo L. .
ARCHIVES OF DISEASE IN CHILDHOOD-EDUCATION AND PRACTICE EDITION, 2012, 97 (06) :217-221
[4]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[5]  
Boren Suzanne Austin, 2010, J Diabetes Sci Technol, V4, P84
[6]  
Boyd JC, 2001, CLIN CHEM, V47, P209
[7]  
Ceriotti Ferruccio, 2015, J Diabetes Sci Technol, V9, P268, DOI 10.1177/1932296814563351
[8]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[9]   System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197 [J].
Freckmann, Guido ;
Baumstark, Annette ;
Jendrike, Nina ;
Zschornack, Eva ;
Kocher, Serge ;
Tshiananga, Jacques ;
Heister, Frank ;
Haug, Cornelia .
DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 (03) :221-231
[10]   Measurement of blood acetoacetate and β-hydroxybutyrate in an automatic analyser [J].
Galán, A ;
Hernández, JM ;
Jimenez, O .
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY, 2001, 23 (03) :69-76