Evaluation of stability and capsular bag opacification of a new foldable adjustable intraocular lens

被引:18
作者
MacLean, Kyle D. [1 ]
Werner, Liliana [1 ]
Kramer, Gregory D. [1 ]
Farukhi, M. Aabid [1 ]
Gardiner, Gareth L. [1 ]
Kahook, Malik Y. [2 ]
Mamalis, Nick [1 ]
机构
[1] Univ Utah, John A Moran Eye Ctr, Dept Ophthalmol & Visual Sci, Salt Lake City, UT 84132 USA
[2] Univ Colorado, Sch Med, Dept Ophthalmol, Aurora, CO USA
关键词
cataract surgery; intraocular lens; posterior capsule opacification; ZEALAND ALBINO RABBITS; CATARACT-SURGERY; FOLLOW-UP; REGENERATION; IMPLANTATION; PREVENTION; EXTRACTION; OUTCOMES;
D O I
10.1111/ceo.12526
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
BackgroundIncorrect lens power remains one of the most common reasons for intraocular lens explantation/exchange. We evaluated stability and effect on capsular bag opacification of a new foldable modular lens system, with a base component and an optic component in rabbit eyes. MethodsBilateral phacoemulsification was performed on six rabbits; one eye received the test lens (Harmoni Modular Lens, ClarVista Medical), whereas the other received a single-piece hydrophobic acrylic control lens (SA60AT, Alcon). Slit-lamp examinations were performed at postoperative weeks 1, 2, 3, 4 and 6. All rabbits were sacrificed at week 6. After enucleation, the eyes were examined grossly from the Miyake-Apple view. Following, all globes were sectioned and processed for histopathological examination. ResultsThe modular lens was stable in all six rabbits. Uveal biocompatibility was similar to or better in the test eyes when compared with control eyes up to 6 weeks postoperatively. Overall, capsular bag opacification was found to be significantly lower in the test eyes when compared with control eyes. At the 6-week gross examination, central posterior capsule opacification was scored as 0.580.73 in the test group, and as 3.00 +/- 1.26 in the control group (paired two samples for means t-test; two-tail P=0.005). ConclusionsThe Harmoni Modular system remains stable within the capsular bag after implantation. Due to the design of the base component featuring long loops, along with peripheral anterior and posterior square edges, the lens appeared to prevent overall capsular bag opacification in the rabbit model.
引用
收藏
页码:648 / 654
页数:7
相关论文
共 50 条
  • [21] Dead bag syndrome-in the capsular bag subluxated intraocular lens centration and refixation: A new technique
    Thakur, Anchal
    Bansal, Muskaan
    Challa, Divya
    Malhotra, Chintan
    Jain, Arun K.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (10) : 3412 - 3414
  • [22] Nd:YAG capsulotomy in posterior capsular opacification after implantation of a multifocal intraocular lens
    Dick, B
    Schwenn, O
    Eisenmann, D
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1997, 211 (06) : 363 - 368
  • [23] EFFECT OF INTRAOCULAR-LENS OPTIC DESIGN ON POSTERIOR CAPSULAR OPACIFICATION
    BORN, CP
    RYAN, DK
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (02) : 188 - 192
  • [24] Liquid-like layer coated intraocular lens for posterior capsular opacification prevention
    Wu, Qianni
    Liu, Dong
    Chen, Wan
    Chen, Hui
    Yang, Cheng
    Li, Xiangling
    Yang, Chengduan
    Lin, Haotian
    Chen, Shuyi
    Hu, Ning
    Chen, Weirong
    Xie, Xi
    APPLIED MATERIALS TODAY, 2021, 23
  • [25] EVALUATION OF VITRECTOMY AND REIMPLANTATION FOLLOWING LATE DISLOCATION OF THE INTRAOCULAR LENS-CAPSULAR BAG COMPLEX A 3-Year Follow-up Study
    Bande, Manuel F.
    Garcia-Garces, Irina
    Paniagua, Laura
    Ruiz-Oliva, Francisco
    Pineiro, Antonio
    Jose Blanco-Teijeiro, Maria
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2017, 37 (05): : 925 - 929
  • [26] Effect of total lens epithelial cell destruction on intraocular lens fixation in the human capsular bag
    Spalton, David J.
    Russell, Sarah L.
    Evans-Gowing, Richard
    Eldred, Julie A.
    Wormstone, I. Michael
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (02) : 306 - 312
  • [27] Rotation versus non-rotation of intraocular lens for prevention of posterior capsular opacification
    Joshi, Rajesh S.
    Chavan, Shrutika A.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (09) : 1428 - 1432
  • [28] Postoperative posterior capsular striae and the posterior capsular opacification in patients implanted with two types of intraocular lens material
    Joshi, Rajesh Subhash
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2017, 65 (06) : 466 - 471
  • [29] Hydrophobic versus double-square-edged hydrophilic foldable acrylic intraocular lens: Effect on posterior capsule opacification
    Iwase, Takeshi
    Nishi, Yutarou
    Oveson, Brian C.
    Jo, Young-Joon
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (06) : 1060 - 1068
  • [30] Reopening the capsular bag by removing the capsular proliferative membrane to enable secondary in-the-bag intraocular lens implantation in pediatric aphakic eyes
    Zhou, Yue
    Zhang, Yu
    Chen, Hui
    Bai, Bingyu
    Wang, Qiwei
    Lin, Zhuoling
    Li, Jing
    Zhang, Xiulan
    Chen, Wan
    Chen, Weirong
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2024, 50 (03) : 250 - 256