Design of an Out-Folded Patch Antenna With a Zeroth-Order Resonance for Non-Invasive Continuous Glucose Monitoring

被引:9
|
作者
Nguyen, Mai Anh [1 ]
Bien, Franklin [1 ]
Byun, Gangil [1 ]
机构
[1] Ulsan Natl Inst Sci & Technol UNIST, Ulsan 44919, South Korea
关键词
Glucose; Transmission line measurements; Geometry; Antennas; Skin; Resonant frequency; Antenna measurements; Composite right-; left-handed (CRLH) transmission line (TL); continuous glucose monitoring (CGM); folded patch antenna; noninvasive sensing; zeroth-order resonance (ZOR); BLOOD-GLUCOSE; SENSOR;
D O I
10.1109/TAP.2022.3177517
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This article proposes an out-folded patch antenna with a zeroth-order resonance (ZOR) that is robust to a high-dielectric condition of a human body for noninvasive continuous glucose monitoring (CGM). The proposed antenna has two faces of a tapered section and a rectangular patch separated by a gap, and those faces are connected through vias at two corners near the gap. Two identical antennas are fabricated on a flexible printed circuit board (FPCB), and the properties of ZOR are compared with those of the fundamental mode to demonstrate the feasibility. The ZOR of the proposed antenna exhibits a negligible frequency shift, and its quality factor is 2.9 times higher than the fundamental mode, when attached to the skin. The evaluation is further extended to a human experiment using the two antennas for real-time estimation of coupling strength according to the change in glucose concentration. The trend of measured coupling strength well describes that of the glucose concentration obtained from a glucose meter with a mean absolute relative difference of 12.23%. The results demonstrate that the proposed concept has a great potential to improve the accuracy of noninvasive estimation in microwave spectrum.
引用
收藏
页码:8932 / 8940
页数:9
相关论文
共 50 条
  • [21] Compact Continuous Non-Invasive Blood Glucose Monitoring using Bluetooth
    Sreenivas, Chaithanya
    Laha, Soumyasanta
    2019 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS 2019), 2019,
  • [22] Differential Continuous Wave Photoacoustic Spectroscopy for Non-Invasive Glucose Monitoring
    Tanaka, Yujiro
    Tajima, Takuro
    Seyama, Michiko
    Waki, Kayo
    IEEE SENSORS JOURNAL, 2020, 20 (08) : 4453 - 4458
  • [23] A Review of Non-Invasive Optical Systems for Continuous Blood Glucose Monitoring
    Alsunaidi, Bushra
    Althobaiti, Murad
    Tamal, Mahbubunnabi
    Albaker, Waleed
    Al-Naib, Ibraheem
    SENSORS, 2021, 21 (20)
  • [24] A Zeroth-Order Slot-Loaded Cap-Shaped Patch Antenna With Omnidirectional Radiation Characteristic for UHF RFID Tag Design
    Ooi, Shin-Yi
    Chee, Pei-Song
    Lim, Eng-Hock
    Low, Jen-Hahn
    Bong, Fwee-Leong
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2023, 71 (01) : 131 - 139
  • [25] Continuous non-invasive glucose monitoring in type I diabetic patients
    Hememann, L
    Berger, M
    Bocker, D
    Christiansen, JS
    Essenpreis, M
    Gries, FA
    Heise, T
    Koschinsky, T
    Orskov, H
    Patterson, MS
    SchmelzEisen, G
    DIABETES, 1996, 45 : 813 - 813
  • [26] Design Flexible Microstrip Antenna for Non-Invasive Blood Glucose Detection
    Rahayu, Yusnita
    Arifa, Fildza
    Kholik, Dede Abdul
    Anhar
    Masdar, Huriatul
    Praludi, Teguh
    2022 INTERNATIONAL CONFERENCE ON RADAR, ANTENNA, MICROWAVE, ELECTRONICS, AND TELECOMMUNICATIONS (ICRAMET): EMERGING SCIENCE AND INDUSTRIAL INNOVATION IN ELECTRONICS AND TELECOMMUNICATION, 2022, : 179 - 182
  • [27] Non-invasive, intelligent, and continuous monitoring of human blood glucose with UWB dual-antenna and cascade CNN
    Wang, Zengxiang
    Xiao, Xia
    Pang, Yanwei
    Su, Wenling
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2023, 34 (09)
  • [28] A Glucose Sensing Contact Lens: A Non-Invasive Technique for Continuous Physiological Glucose Monitoring
    Ramachandram Badugu
    Joseph R. Lakowicz
    Chris D. Geddes
    Journal of Fluorescence, 2003, 13 : 371 - 374
  • [29] A glucose sensing contact lens: A non-invasive technique for continuous physiological glucose monitoring
    Badugu, R
    Lakowicz, JR
    Geddes, CD
    JOURNAL OF FLUORESCENCE, 2003, 13 (05) : 371 - 374
  • [30] Design of Continuous Non-Invasive Blood Glucose Monitoring Sensor Based on a Microwave Split Ring Resonator
    Choi, Heungjae
    Nylon, Jack
    Luzio, Stephen
    Beutler, Jan
    Porch, Adrian
    2014 IEEE MTT-S INTERNATIONAL MICROWAVE WORKSHOP SERIES ON RF AND WIRELESS TECHNOLOGIES FOR BIOMEDICAL AND HEALTHCARE APPLICATIONS (IMWS-BIO), 2014,