Force and Position Control System for Freehand Ultrasound

被引:86
作者
Gilbertson, Matthew W. [1 ]
Anthony, Brian W. [1 ]
机构
[1] MIT, Dept Mech Engn, Mfg & Prod Lab, Cambridge, MA 02139 USA
关键词
Ergonomics; force control; medical robotics; pre-compression; pressure; ultrasound (US); HUMAN HAND; ELASTOGRAPHY; ELASTICITY; TRACKING; TISSUE; VIBRATION; SURGERY; DEVICE; PROBE;
D O I
10.1109/TRO.2015.2429051
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
A hand-held force-controlled ultrasound probe has been developed for medical imaging applications. The probe-patient contact force can be held constant to improve image stability, swept through a range, or cycled. The mechanical portion of the device consists of a ball screw linear actuator driven by a servo motor, along with a load cell, accelerometer, and limit switches. The performance of the system was assessed in terms of the frequency response to simulated sonographer hand motion and in hand-held image feature tracking during simulated patient motion. The system was found to attenuate contact force variation by 97% at 0.1 Hz, 83% at 1 Hz, and 33% 10 Hz, a range that spans the typical human hand tremor frequency spectrum. In studies with 15 human operators, the device applied the target contact force with ten times less variation than in conventional ultrasound imaging. An ergonomic, human-in-the-loop, imaging-workflow enhancing control scheme, which combines both force-and position-control, permits smooth making and breaking of probe-patient contact, and helps the operator keep the probe centered within its range of motion. By controlling ultrasound probe contact force and consequently the amount of tissue deformation, the system enhances the repeatability, usability, and diagnostic capabilities of ultrasound imaging.
引用
收藏
页码:835 / 849
页数:15
相关论文
共 55 条
[1]  
Anthony B., 2012, Patent No. [8 328 725, 8328725]
[2]  
Anthony B., 2012, Patent No. [US Patents 8 382 671, 8382671]
[3]  
Anthony B., 2012, Patent No. [US Patents 8 333 704, 8333704]
[4]   Effects of Precompression on Elasticity Imaging of the Breast Development of a Clinically Useful Semiquantitative Method of Precompression Assessment [J].
Barr, Richard G. ;
Zhang, Zheng .
JOURNAL OF ULTRASOUND IN MEDICINE, 2012, 31 (06) :895-902
[5]  
Bhatia Kunwar S S, 2013, Cancer Imaging, V13, P253, DOI 10.1102/1470-7330.2013.0026
[6]   Robotics revolutionizing surgery: the intuitive surgical "Da Vinci" system [J].
Broeders, IAMJ ;
Ruurda, J .
INDUSTRIAL ROBOT, 2001, 28 (05) :387-391
[7]   A system for simultaneously measuring contact force, ultrasound, and position information for use in force-based correction of freehand scanning [J].
Burcher, MR ;
Noble, JA ;
Han, LH ;
Gooding, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (08) :1330-1342
[8]  
Chadwick S. G., 2012, P TRANSP RES BOARD A, P1
[9]   A needle tracking device for ultrasound guided percutaneous procedures [J].
Chan, C ;
Lam, F ;
Rohling, R .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (11) :1469-1483
[10]  
Degoulange E, 1998, 1998 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS - PROCEEDINGS, VOLS 1-3, P959, DOI 10.1109/IROS.1998.727423