Liquid chromatographic assay of ivermectin in human plasma for application to clinical pharmacokinetic studies

被引:30
作者
Kitzman, D [1 ]
Wei, SY [1 ]
Fleckenstein, L [1 ]
机构
[1] Univ Iowa, Coll Pharm, Iowa City, IA 52242 USA
关键词
ivermectin; antifilarial; antiparasitic; HPLC; liquid chromatography;
D O I
10.1016/j.jpba.2005.08.026
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is a need for an accurate, sensitive and selective high-performance liquid chromatography (HPLC) method for the quantitation of ivermectin in human plasma that separates the parent drug from metabolites. Ivermectin and the internal standard, moxidectin, were extracted from 0.2 ml of human plasma using Oasis HLB solid phase extraction cartridges. After extraction. fluorescent derivatives of ivemectin and moxidectin were made by reaction with trifluoroacetic anhydride and N-methylimidazole. Separation was achieved on a Alltech Ultrasphere C 18 5 L column with a mobile phase composed of tetrahydrofuran-acetonitrile-water (40:38:22 v/v/v). Detection is by fluorescence, with an excitation of 365 min and emission of 475 nm. The retention times of ivermectin and internal standard, moxidectin are approximately 24.5 and 12.5 min, respectively. The assay is linear over the concentration range of 0.2-200 ng/ml of ivermectin in human plasma (r = 0.9992, weighted by 1/concentration). Recoveries of ivermectin are greater than 80% at all concentrations. The analysis Of quality control samples for ivermectin 0.2, 25, and 200 ng/ml demonstrated excellent precision with coefficient of variation of 6.1, 3.6 and 2.3%, respectively (it = 6). The method is accurate with all intra-day (it = 6) and interday (n = 12) mean concentration within 10% of nominal values at all quality control sample concentrations. Storage stability for 30 days at -80 degrees C and after three freeze-thaw cycles are within acceptable limits. The method separates ivermectin from multiple less and more polar unidentified metabolites. This method is robust and suitable for clinical pharmacokinetic studies. The analytical procedure has been applied to a pharmacokinetic study of ivermectin in healthy volunteers and to the analysis of plasma specimens from patients with disseminated strongyloidiasis. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:1013 / 1020
页数:8
相关论文
共 18 条
[1]   Liquid chromatographic assay of moxidectin in human plasma for application to pharmacokinetic studies [J].
Chen, YC ;
Hung, YP ;
Fleckenstein, L .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 29 (05) :917-926
[2]   DETERMINATION OF IVERMECTIN IN HUMAN-PLASMA AND MILK BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION [J].
CHIOU, R ;
STUBBS, RJ ;
BAYNE, WF .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 416 (01) :196-202
[3]  
Chiu S. H. L., 1989, Ivermectin and abamectin., P131
[4]   Determination of ivermectin B1a in animal plasma by liquid chromatography combined with electrospray ionization mass spectrometry [J].
Croubels, S ;
De Baere, S ;
Cherlet, M ;
De Backer, P .
JOURNAL OF MASS SPECTROMETRY, 2002, 37 (08) :840-847
[5]   LIQUID-CHROMATOGRAPHIC DETERMINATION OF IVERMECTIN IN ANIMAL PLASMA WITH TRIFLUOROACETIC-ANHYDRIDE AND N-METHYLIMIDAZOLE AS THE DERIVATIZATION REAGENT [J].
DEMONTIGNY, P ;
SHIM, JSK ;
PIVNICHNY, JV .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (06) :507-511
[6]  
Fink D. W., 1989, Ivermectin and abamectin., P113
[7]   Evolution of a specific fluorogenic derivatization of ivermectin for bioanalytical applications - A review [J].
Fink, DW ;
deMontigny, P ;
Shim, JSK .
ANALYST, 1996, 121 (11) :1533-1537
[8]   Safety, tolerability, and pharmacokinetics of escalating high doses of ivermectin in healthy adult subjects [J].
Guzzo, CA ;
Furtek, CI ;
Porras, AG ;
Chen, C ;
Tipping, R ;
Clineschmidt, CM ;
Sciberras, DG ;
Hsieh, JYK ;
Lasseter, KC .
JOURNAL OF CLINICAL PHARMACOLOGY, 2002, 42 (10) :1122-1133
[9]   DETERMINATION OF 22,23-DIHYDROAVERMECTIN-B1A IN DOG PLASMA USING SOLID-PHASE EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KOJIMA, K ;
YAMAMOTO, K ;
NAKANISHI, Y ;
KATAE, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 413 :326-331
[10]  
KRISHNA DR, 1993, ARZNEIMITTEL-FORSCH, V43-1, P609