Compensation System for Biomagnetic Measurements with Optically Pumped Magnetometers inside a Magnetically Shielded Room

被引:55
作者
Jodko-Wladzinska, Anna [1 ]
Wildner, Krzysztof [1 ]
Palko, Tadeusz [1 ]
Wladzinski, Michal [1 ]
机构
[1] Warsaw Univ Technol, Fac Mechatron, Inst Metrol & Biomed Engn, Boboli 8 St, PL-02525 Warsaw, Poland
关键词
optically pumped magnetometer; magnetically shielded room; Helmholtz coils; biomagnetism; MAGNETOENCEPHALOGRAPHY; FIELD; TEMPERATURE; PERFORMANCE;
D O I
10.3390/s20164563
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Magnetography with superconducting quantum interference device (SQUID) sensor arrays is a well-established technique for measuring subtle magnetic fields generated by physiological phenomena in the human body. Unfortunately, the SQUID-based systems have some limitations related to the need to cool them down with liquid helium. The room-temperature alternatives for SQUIDs are optically pumped magnetometers (OPM) operating in spin exchange relaxation-free (SERF) regime, which require a very low ambient magnetic field. The most common two-layer magnetically shielded rooms (MSR) with residual magnetic field of 50 nT may not be sufficiently magnetically attenuated and additional compensation of external magnetic field is required. A cost-efficient compensation system based on square Helmholtz coils was designed and successfully used for preliminary measurements with commercially available zero-field OPM. The presented setup can reduce the static ambient magnetic field inside a magnetically shielded room, which improves the usability of OPMs by providing a proper environment for them to operate, independent of initial conditions in MSR.
引用
收藏
页码:1 / 14
页数:14
相关论文
共 47 条
[41]   Biomagnetism using SQUIDs: status and perspectives [J].
Sternickel, K ;
Braginski, AI .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2006, 19 (03) :S160-S171
[42]   Low-Cost Fetal Magnetocardiography: A Comparison of Superconducting Quantum Interference Device and Optically Pumped Magnetometers [J].
Strand, Sarah ;
Lutter, William ;
Strasburger, Janette F. ;
Shah, Vishal ;
Baffa, Oswaldo ;
Wakai, Ronald T. .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2019, 8 (16)
[43]   Magnetocardiography for fetal arrhythmias [J].
Strasburger, Janette F. ;
Cheulkar, Bageshree ;
Wakai, Ronad T. .
HEART RHYTHM, 2008, 5 (07) :1073-1076
[44]  
Stroink G., 2007, MAGNETISM MED HDB, V2nd, P183
[45]   Measures to reduce the residual field and field gradient inside a magnetically shielded room by a factor of more than 10 [J].
Voigt, Jens ;
Knappe-Grueneberg, Silvia ;
Schnabel, Allard ;
Koerber, Rainer ;
Burghoff, Martin .
METROLOGY AND MEASUREMENT SYSTEMS, 2013, 20 (02) :239-248
[46]   BIOMAGNETISM TOPICAL REVIEW [J].
WILLIAMSON, SJ ;
KAUFMAN, L .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1981, 22 (02) :129-201
[47]   Magnetoencephalography with an atomic magnetometer [J].
Xia, H. ;
Baranga, A. Ben-Amar ;
Hoffman, D. ;
Romalis, M. V. .
APPLIED PHYSICS LETTERS, 2006, 89 (21)