A comparison of two headless compression screws for operative treatment of scaphoid fractures

被引:26
|
作者
Grewal, Ruby [1 ]
Assini, Joseph [1 ]
Sauder, David [1 ]
Ferreira, Louis [1 ]
Johnson, Jim [1 ]
Faber, Kenneth [1 ]
机构
[1] St Josephs Hlth Care, Hand & Upper Limb Ctr, London, ON N6A 4L6, Canada
来源
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH | 2011年 / 6卷
关键词
CAST IMMOBILIZATION; FIXATION; WAIST;
D O I
10.1186/1749-799X-6-27
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The purpose of this study was to compare the interfragmentary compression force across a simulated scaphoid fracture by two commonly used compression screw systems; the Acutrak 2 Standard and the 3.0 mm Synthes headless compression screw. Methods: Sixteen (8 pairs; 6 female, 2 male) cadaver scaphoids were randomly assigned to receive either the Acutrak 2 or Synthes screw with the contralateral scaphoid designated to receive the opposite. Guide wires were inserted under fluoroscopic control. Following transverse osteotomy, the distal and proximal fragments were placed on either side of a custom load cell, to measure interfragmentary compression. Screws were placed under fluoroscopic control using the manufacturer's recommended surgical technique. Compressive forces were measured during screw insertion. Recording continued for an additional 60s in order to measure any loss of compression after installation was complete. The peak and final interfragmentary compression were recorded and paired t-tests performed. Results: The mean peak compression generated by the Acutrak 2 Standard was greater than that produced by the Synthes compression screw (103.9 +/- 33.2 N vs. 88.7 +/- 38.6 N respectively, p = 0.13). The mean final interfragmentary compression generated by the Acutrak 2 screw (68.6 +/- 36.4 N) was significantly greater (p = 0.04) than the Synthes screw (37.2 +/- 26.8 N). Specimens typically reached a steady state of compression by 120-150s after final tightening. Conclusion: Peak interfragmentary compression observed during screw installation was similar for both screw systems. However, the mean interfragmentary compression generated by the Acutrak 2 Standard was significantly greater. Our study demonstrates that the Synthes headless compression screw experienced a greater loss of interfragmentary compressive force from the time of installation to the final steady state compression level. The higher post-installation compression of the Acutrak 2 Standard may be attributable to the greater number of threads throughout the entire length of the screw. The clinical significance of these results, are, at this point uncertain. We do demonstrate that a fully threaded design offers a more reliable compression that may translate to more predictable bony union.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Comparison of Operative and Non-Operative Treatment of Acute Undisplaced or Minimally-Displaced Scaphoid Fractures: A Meta-Analysis of Randomized Controlled Trials
    Shen, Longxiang
    Tang, Jianfei
    Luo, Congfeng
    Xie, Xuetao
    An, Zhiquan
    Zhang, Changqing
    PLOS ONE, 2015, 10 (05):
  • [32] Non-operative treatment of non-displaced scaphoid fractures may be preferred
    Rhemrev, Steven J.
    van Leerdam, Roderick H.
    Ootes, Daan
    Beeres, Frank J. P.
    Meylaerts, Sven A. G.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (06): : 638 - 641
  • [33] Percutaneous Retrograde Technique Using Intramedullary Headless Compression Screws for Metacarpal Fractures Under Wide-Awake Local Anaesthesia No Tourniquet
    Ahmed, Naufal
    Norris, Rory
    Faiaz, Aghna
    Sharma, Anirudh
    Bindumadhavan, Santosh
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (11)
  • [34] Interfragmentary compression across a simulated scaphoid fracture - Analysis of 3 screws
    Beadel, GP
    Ferreira, L
    Johnson, JA
    King, GJW
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2004, 29A (02): : 273 - 278
  • [35] Outcomes after arthroscopic treatment for scaphoid nonunion using ipsilateral radius bone graft and a headless compression screw: a comparison between the patients with and without avascular necrosis
    Lin, Ting-Yi
    Fu, Chun-Wei
    Tsai, Pei-Shan
    Huang, Chang-Hung
    Wei, Shun-Hua
    Chuang, Min-Yao
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (04) : 1041 - 1049
  • [36] Biomechanical Comparison of Cortical Lag Screws and Cortical Position Screws for Their Generation of Interfragmentary Compression and Area of Compression in Simulated Lateral Humeral Condylar Fractures
    Fuchter, Lucy M.
    Wills, Dan
    Walsh, William R.
    Newman, Mark
    Tan, Chris
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2023, 36 (05) : 250 - 256
  • [37] Conjoint Analysis of Treatment Preferences for Nondisplaced Scaphoid Fractures
    Shammas, Ronnie L.
    Mela, Nathan
    Wallace, Scott
    Tong, Betty C.
    Huber, Joel
    Mithani, Suhail K.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2018, 43 (07): : 678.e1 - 678.e9
  • [38] Comparison of two different screw trajectories in the treatment of oblique scaphoid fractures: A mechanical study on composite bone models
    Kilic, Bulent
    Caliskan, Mustafa
    Agar, Anil
    Uzun, Bora
    Ertem, Fatih
    Gulabi, Deniz
    Erturk, Cemil
    JOINT DISEASES AND RELATED SURGERY, 2021, 32 (02): : 377 - 382
  • [39] Comparison of Two Percutaneous Volar Approaches for Screw Fixation of Scaphoid Waist Fractures
    Meermans, Geert
    Van Glabbeek, Francis
    Braem, Marc J.
    van Riet, Roger P.
    Hubens, Guy
    Verstreken, Frederik
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (16): : 1369 - 1376
  • [40] Effects of full-threaded headless cannulated compression screws and anatomical plates on the efficacy, safety, and prognosis of patients with triplane fractures of the distal tibia
    Ma, Zhilin
    Feng, Wenfang
    Duan, Xiaowei
    Chen, Xinzhi
    Qiao, Guoyong
    Liu, Zhiping
    AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2022, 14 (03): : 1714 - 1720