In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: Role of implant geometry

被引:50
作者
Calcagnini, Giovanni [1 ]
Triventi, Michele [1 ]
Censi, Federica [1 ]
Mattei, Eugenio [1 ,2 ]
Bartolini, Pietro [1 ]
Kainz, Wolfgang [3 ]
Bassen, Howard I. [3 ]
机构
[1] Italian Natl Inst Hlth, Dept Technol & Hlth, Rome, Italy
[2] Univ Rome, Dept Informat & Syst Sci, Rome, Italy
[3] US FDA, Ctr Devices & Radiol Hlth, Silver Spring, MD USA
关键词
pacemaker; lead heating; fluoroptic measurements; implant geometry; radiofrequency;
D O I
10.1002/jmri.21536
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the effect of the geometry of implantable pacemakers (PMs) on lead heating induced by magnetic resonance imaging (MRI). Materials and Methods: In vitro experiments were conducted with two different setups, using fluoroptic probes to measure the temperature increase. The first experiment consisted of a rectangular box filled with a gelled saline and a pacemaker with its leads. This box was exposed in an MRI birdcage coil to a sinusoidal 64-MHz field with a calibrated whole-body specific absorption rate (WB-SAR) of 1 W/kg. The highest SAR and temperature increase (3000 W/kg. 12 degrees C) occurred for the implant configuration having the largest area. The second experimental setup consisted of a human-shaped torso filled with gelled saline. In this setup the PM and its lead were exposed to a real MRI scanner, using clinical sequences with WB-SAR up to 2 W/kg. Results: We found that higher heating occurs for configurations with longer exposed lead lengths and that right chest PMs showed the highest temperature and local SAR (11.9 degrees C, 2345 W/kg), whereas the left chest PMs were less heated (6.3 degrees C, 1362 W/kg). Implant geometry, exposed lead length, and lead area must be considered in the wide variation of temperature increases induced by MRI. Conclusions: The amount of MRI-induced lead tip heating depends strongly on implant geometry, particularly the lead area, exposed lead length, and position of the implant in the phantom. Critical lead tip heating was found for the longer leads. Therefore, to minimize MRI-induced lead tip heating, the PM lead should be as short as possible.
引用
收藏
页码:879 / 886
页数:8
相关论文
共 24 条
[1]   Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]  
[Anonymous], 2002, C9532002 IEEE
[3]  
*ASTM, 2004, F218202A ASTM
[4]   Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating [J].
Baker, KB ;
Tkach, JA ;
Nyenhuis, JA ;
Phillips, M ;
Shellock, FG ;
Gonzalez-Martinez, J ;
Rezai, AR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (02) :315-320
[5]   MRI-induced heating of selected thin wire metallic implants - laboratory and computational studies - findings and new questions raised [J].
Bassen, H ;
Kainz, W ;
Mendoza, G ;
Kellom, T .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (02) :76-84
[6]   Pacing in magnetic resonance imaging environment:: Clinical and technical considerations on compatibility [J].
Duru, F ;
Luechinger, R ;
Scheidegger, MB ;
Lüscher, TF ;
Boesiger, P ;
Candinas, R .
EUROPEAN HEART JOURNAL, 2001, 22 (02) :113-124
[7]   EFFECT OF MAGNETIC-RESONANCE-IMAGING ON DDD PACEMAKERS [J].
ERLEBACHER, JA ;
CAHILL, PT ;
PANNIZZO, F ;
KNOWLES, RJR .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :437-440
[8]   THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE IMAGERS ON EXTERNAL AND IMPLANTABLE PULSE GENERATORS [J].
FETTER, J ;
ARAM, G ;
HOLMES, DR ;
GRAY, JE ;
HAYES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (04) :720-727
[9]   Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers [J].
Gimbel, JR ;
Johnson, D ;
Levine, PA ;
Wilkoff, BL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (06) :913-919
[10]   EFFECT OF 1.5 TESLA NUCLEAR-MAGNETIC-RESONANCE IMAGING SCANNER ON IMPLANTED PERMANENT PACEMAKERS [J].
HAYES, DL ;
HOLMES, DR ;
GRAY, JE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :782-786