A dosimetric comparison of MammoSite® and ClearPath high-dose-rate breast brachytherapy devices

被引:15
作者
Dickler, Adam [1 ]
Seif, Nell [1 ]
Kirk, Michael C. [1 ]
Pate, Mita B. [2 ]
Bernard, Damian [1 ]
Coon, Alan [1 ]
Dowlatshahi, Kambiz [3 ]
Das, Rupak K. [4 ]
Patel, Rakesh R. [4 ]
机构
[1] Rush Univ, Med Ctr, Dept Radiat Oncol, Chicago, IL 60612 USA
[2] Rush Univ, Med Ctr, Coll Med, Chicago, IL 60612 USA
[3] Rush Univ, Med Ctr, Dept Gen Surg, Chicago, IL 60612 USA
[4] Univ Wisconsin, Dept Human Oncol, Madison, WI USA
关键词
Breast Brachytherapy; Breast Cancer; MammoSite (R); ClearPath; CONSERVING SURGERY; RADIATION-THERAPY; IRRADIATION;
D O I
10.1016/j.brachy.2008.07.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: A new form of partial breast irradiation (PBI), ClearPath (CP) breast brachytherapy, has been introduced. We present our results of a dosimetric comparison of MammoSite (R) (MS) and CP PBI. METHODS AND MATERIALS: The dimensions of the CP device were reconstructed onto the MS planning CT scans for 15 previously treated patients. The mean %V-100, %V-150, %V-200 (percent of the PTV that received 100%, 150%, and 200% of the prescription dose, respectively), ipsilateral breast %V-50 (percent of the ipsilateral normal breast that received 50% of the prescription dose), ipsilateral lung %V-30 (percent of the ipsilateral lung that received 30% of the prescription dose), the heart %V-30 (percent of the heart that received 5% of the prescription dose), and the maximum skin point dose per fraction were then determined for each patient using the two methods of balloon-based PBI. RESULTS: The mean %V100 was 96.5% vs. 96.5%, the mean %V150 was 42.1% vs. 42.9% (p = ns), and the mean V-200 was 11.4% vs. 15.2% (p < .05) for the MS and CP methods, respectively. The mean ipsilateral breast %V-50 was 19.8% vs. 18.0% (p < .05). the mean ipsilateral lung %V-30 was 3.7% vs. 2.8% (p < .05), the mean heart %V-5 was 57.0% vs. 54.3% (p <.05), and the maximum skin point dose per fraction was 312.2 and 273.6 cGy (p <.05) for the MS and CP methods, respectively. CONCLUSIONS: The MS and CP methods of PBI offer comparable target volume coverage: however. the CP device achieves increased normal tissue sparing. (c) 2009 Published by Elsevier Inc on behalf of American Brachytherapy Society.
引用
收藏
页码:14 / 18
页数:5
相关论文
共 14 条
[1]  
*AM SOC BREAST SUR, 2005, CONS STAT ACC PART B
[2]  
Arthur Douglas W, 2002, Brachytherapy, V1, P184, DOI 10.1016/S1538-4721(02)00099-5
[3]   Cosmetic outcome and incidence of infection with the MammoSite breast brachytherapy applicator [J].
Dickler, A ;
Kirk, MC ;
Choo, J ;
Hsi, WC ;
Chu, J ;
Dowlatshahi, K ;
Francescatti, D ;
Shott, S ;
Nguyen, C .
BREAST JOURNAL, 2005, 11 (05) :306-310
[4]   Treatment volume and dose optimization of MammoSite breast brachytherapy applicator [J].
Dickler, A ;
Kirk, M ;
Choo, J ;
Hsi, WC ;
Chu, J ;
Dowlatshahi, K ;
Francescatti, D ;
Nguyen, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (02) :469-474
[5]   A dosimetric comparison of MammoSite high-dose-rate brachytherapy and Xoft Axxent electronic brachytherapy [J].
Dickler, Adam ;
Kirk, Michael C. ;
Seif, Neil ;
Griem, Katherine ;
Dowlatshahi, Kambiz ;
Francescatti, Darius ;
Abrams, Ross A. .
BRACHYTHERAPY, 2007, 6 (02) :164-168
[6]   Breast conserving surgery and accelerated partial breast irradiation using the MammoSite system - Initial clinical experience [J].
DiFronzo, LA ;
Tsai, PI ;
Hwang, JM ;
Lee, JJ ;
Ryoo, MC ;
Rahimian, J ;
Tome, M ;
Takasugi, JK ;
Haigh, PI .
ARCHIVES OF SURGERY, 2005, 140 (08) :787-792
[7]   Dosimetric characteristics of the MammoSite RTS, a new breast brachytherapy applicator [J].
Edmundson, GK ;
Vicini, FA ;
Chen, PY ;
Mitchell, C ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (04) :1132-1139
[8]   Thirty-nine month results with the MammoSite brachytherapy applicator: Details regarding cosmesis, toxicity and local control in partial breast irradiation [J].
Keisch, M ;
Vicini, F ;
Scroggins, T ;
Hebert, M ;
White, J ;
Kuske, R ;
Quiet, C ;
Arthur, D ;
Streeter, O .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02) :S6-S6
[9]   A dosimetric comparison of three-dimensional conformal, intensity-modulated radiation therapy, and MammoSite partial-breast irradiation [J].
Khan, Atif J. ;
Kirk, Michael C. ;
Mehta, Par S. ;
Seif, Neil S. ;
Griem, Katherine L. ;
Bernard, Damian A. ;
Chu, James C. H. ;
Dickler, Adam .
BRACHYTHERAPY, 2006, 5 (03) :183-188
[10]   Long-term results of wide field brachytherapy as the sole method of radiation therapy after segmental mastectomy for Tis,1,2 breast cancer [J].
King, TA ;
Bolton, JS ;
Kuske, RR ;
Fuhrman, GM ;
Scroggins, TG ;
Jiang, XZ .