A portable system for in-situ re-calibration of force platforms:: Theoretical validation

被引:10
作者
Cedraro, Andrea [1 ]
Cappello, Angelo [1 ]
Chiari, Lorenzo [1 ]
机构
[1] Univ Bologna, Biomed Engn Grp, Dept Elect Comp Sci & Syst, Bologna, Italy
关键词
force platform; re-calibration matrix; optimization; movement analysis; error propagation;
D O I
10.1016/j.gaitpost.2008.03.006
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
The periodic calibration of measurement devices, Such as force platforms (FPs), guarantees and optimizes the quality of the acquired data. In this Study, the theoretical validation of a portable system for the in-situ calibration of six-component FPs is presented. A least-squares algorithm, that estimates the FP re-calibration matrix, was designed and tested using a simulation approach. The algorithm is the core of a portable system, described in a separate paper, and represents a refinement of an algorithm previously presented by Cappello et al. [Cappello A, Lenzi D, Chiari L. Periodical in-situ re-calibration of force platforms: a new method for the robust estimation of the calibration matrix. Med Biol Eng Comput 2004;42:350-5]. The new algorithm assumes that the calibration inputs are known, 3-D, time-varying loads, applied to the FP at known coordinates, and not constrained in their direction of application. Simulation results confirmed the a priori identifiability of the re-calibration matrix and some of the algorithm features were optimized in the perspective of an actual building of the system. With the aid of simple sinusoidal loads, applied in at least five different points, we proved that the algorithm can ensure errors less than 0.2 N and 0.4 Nm when calculating force and moment components of an applied load. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:488 / 494
页数:7
相关论文
共 16 条
[1]   Data management in gait analysis for clinical applications [J].
Benedetti, MG ;
Catani, F ;
Leardini, A ;
Pignotti, E ;
Giannini, S .
CLINICAL BIOMECHANICS, 1998, 13 (03) :204-215
[2]   CALCULATION OF VERTICAL GROUND REACTION FORCE ESTIMATES DURING RUNNING FROM POSITIONAL DATA [J].
BOBBERT, MF ;
SCHAMHARDT, HC ;
NIGG, BM .
JOURNAL OF BIOMECHANICS, 1991, 24 (12) :1095-1105
[3]   A quality control procedure for force platforms [J].
Browne, J ;
O'Hare, N .
PHYSIOLOGICAL MEASUREMENT, 2000, 21 (04) :515-524
[4]   Periodical in-situ re-calibration of force platforms:: a new method for the robust estimation of the calibration matrix [J].
Cappello, A ;
Lenzi, D ;
Chiari, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (03) :350-355
[5]   GAIT ANALYSIS METHODOLOGY [J].
CAPPOZZO, A .
HUMAN MOVEMENT SCIENCE, 1984, 3 (1-2) :27-50
[6]  
CEDRARO A, GAIT POSTURE UNPUB
[7]   Do strain gauge force platforms need in situ correction? [J].
Chockalingam, N ;
Giakas, G ;
Iossifidou, A .
GAIT & POSTURE, 2002, 16 (03) :233-237
[8]  
Draper NR., 1966, APPL REGRESSION ANAL
[9]  
HARRIS GF, 1996, HUMAN MOTION ANAL
[10]   Influence of body segment parameters and modeling assumptions on the estimate of center of mass trajectory [J].
Lenzi, D ;
Cappello, A ;
Chiari, L .
JOURNAL OF BIOMECHANICS, 2003, 36 (09) :1335-1341