Effect of room illuminance on monitor black level luminance and monitor calibration

被引:26
作者
Chakrabarti, K
Kaczmarek, RV
Thomas, JA
Romanyukha, A
机构
[1] US FDA, Radiol Devices Branch,DRARD, ODE, CDRH, Rockville, MD 20850 USA
[2] OHIP, Div Mammog Qual & Radiat Program, CDRH, Rockville, MD 20850 USA
[3] Uniformed Serv Univ Hlth Sci, Dept Radiol, Bethesda, MD 20814 USA
关键词
black level; luminance; illuminance; image contrast; diffuse reflection; image surround;
D O I
10.1007/s10278-003-1720-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this article we demonstrate the effect of room illuminance and surrounding monitor black level luminance on image quality for soft copy interpretation. Luminance values of a 10% central target and image quality evaluations and observer performance using a contrast-detail mammography (CDMAM) phantom demonstrate these effects. Our results indicate that high room illuminance has a more damaging effect on image quality when the surrounding monitor luminance is 0% to 5% of the maximum monitor luminance. The effect of room illuminance is less obvious when the surrounding monitor luminance is 20% of the maximum.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 15 条
  • [1] BIJKERK KR, 1992, RADIOLOGY, V185, P395
  • [2] Effect of different background lighting conditions on diagnostic performance of digital and film images
    Cederberg, RA
    Frederiksen, NL
    Benson, BW
    Shulman, JD
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 1998, 27 (05) : 293 - 297
  • [3] Optimization of viewing conditions and phantom image quality evaluations on GE DMR and full-field digital mammography system
    Chakrabarti, K
    Thomas, JA
    Kaczmarek, RV
    Waynant, RW
    Loscocco, MF
    [J]. JOURNAL OF DIGITAL IMAGING, 2000, 13 (02) : 226 - 227
  • [4] Image quality degradation by light scattering in display devices
    Flynn, MJ
    Badano, A
    [J]. JOURNAL OF DIGITAL IMAGING, 1999, 12 (02) : 50 - 59
  • [5] Influence of room lighting on grey-scale perception with a CRT- and a TFT monitor display
    Haak, R
    Wicht, MJ
    Hellmich, M
    Nowak, G
    Noack, MJ
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (03) : 193 - 197
  • [6] Effect of display luminance on the feature detection rates of masses in mammograms
    Hemminger, BM
    Dillon, AW
    Johnston, RE
    Muller, KE
    Deluca, MC
    Coffey, CS
    Pisano, ED
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2266 - 2272
  • [7] Why and how is soft copy reading possible in clinical practice?
    Mertelmeier, T
    [J]. JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) : 3 - 11
  • [8] *NAT EL MAN ASS DI, 2000, GRAYSC STAND DISPL F
  • [9] Roehrig H, 2001, J DIGIT IMAGING, V14, P142, DOI 10.1053/jdim.2001.28993
  • [10] Flat-panel digital mammography system: Contrast-detail comparison between screen-film radiographs and hard-copy images
    Suryanarayanan, S
    Karellas, A
    Vedantham, S
    Ved, H
    Baker, SP
    D'Orsi, CJ
    [J]. RADIOLOGY, 2002, 225 (03) : 801 - 807