Effect of room illuminance on monitor black level luminance and monitor calibration

被引:26
作者
Chakrabarti, K
Kaczmarek, RV
Thomas, JA
Romanyukha, A
机构
[1] US FDA, Radiol Devices Branch,DRARD, ODE, CDRH, Rockville, MD 20850 USA
[2] OHIP, Div Mammog Qual & Radiat Program, CDRH, Rockville, MD 20850 USA
[3] Uniformed Serv Univ Hlth Sci, Dept Radiol, Bethesda, MD 20814 USA
关键词
black level; luminance; illuminance; image contrast; diffuse reflection; image surround;
D O I
10.1007/s10278-003-1720-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this article we demonstrate the effect of room illuminance and surrounding monitor black level luminance on image quality for soft copy interpretation. Luminance values of a 10% central target and image quality evaluations and observer performance using a contrast-detail mammography (CDMAM) phantom demonstrate these effects. Our results indicate that high room illuminance has a more damaging effect on image quality when the surrounding monitor luminance is 0% to 5% of the maximum monitor luminance. The effect of room illuminance is less obvious when the surrounding monitor luminance is 20% of the maximum.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 15 条
[1]  
BIJKERK KR, 1992, RADIOLOGY, V185, P395
[2]   Effect of different background lighting conditions on diagnostic performance of digital and film images [J].
Cederberg, RA ;
Frederiksen, NL ;
Benson, BW ;
Shulman, JD .
DENTOMAXILLOFACIAL RADIOLOGY, 1998, 27 (05) :293-297
[3]   Optimization of viewing conditions and phantom image quality evaluations on GE DMR and full-field digital mammography system [J].
Chakrabarti, K ;
Thomas, JA ;
Kaczmarek, RV ;
Waynant, RW ;
Loscocco, MF .
JOURNAL OF DIGITAL IMAGING, 2000, 13 (02) :226-227
[4]   Image quality degradation by light scattering in display devices [J].
Flynn, MJ ;
Badano, A .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (02) :50-59
[5]   Influence of room lighting on grey-scale perception with a CRT- and a TFT monitor display [J].
Haak, R ;
Wicht, MJ ;
Hellmich, M ;
Nowak, G ;
Noack, MJ .
DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (03) :193-197
[6]   Effect of display luminance on the feature detection rates of masses in mammograms [J].
Hemminger, BM ;
Dillon, AW ;
Johnston, RE ;
Muller, KE ;
Deluca, MC ;
Coffey, CS ;
Pisano, ED .
MEDICAL PHYSICS, 1999, 26 (11) :2266-2272
[7]   Why and how is soft copy reading possible in clinical practice? [J].
Mertelmeier, T .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) :3-11
[8]  
*NAT EL MAN ASS DI, 2000, GRAYSC STAND DISPL F
[9]  
Roehrig H, 2001, J DIGIT IMAGING, V14, P142, DOI 10.1053/jdim.2001.28993
[10]   Flat-panel digital mammography system: Contrast-detail comparison between screen-film radiographs and hard-copy images [J].
Suryanarayanan, S ;
Karellas, A ;
Vedantham, S ;
Ved, H ;
Baker, SP ;
D'Orsi, CJ .
RADIOLOGY, 2002, 225 (03) :801-807