Oblique gantry - An alternative for heavy-ion cancer therapy

被引:2
作者
Pavlovic, M [1 ]
机构
[1] Gesell Schwerionenforsch GmbH, D-64220 Darmstadt, Germany
关键词
beam-delivery; conformal therapy; gantry; heavy-ion therapy;
D O I
10.1007/BF03038881
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Problem: To design a gantry-type beam delivery system for heavy-ion therapy beams. In order to allow for installation of the gantry in hospital environment, small overall gantry size and low weight are required. The isocentric configuration is to be preferred. Solution: A new gantry concept with oblique output beam is introduced. Transformations between the treatment angle, angles of gantry and patient table rotations are derived. Results and Conclusions: The isocentric gantry with the overall radius of 2.8 m is feasible using the 60 degrees output beam. The oblique output beam imposes some restrictions on gantry capabilities, which has to be carefully evaluated by radiation oncology community.
引用
收藏
页码:24 / 26
页数:3
相关论文
共 45 条
[21]   A horizontal CT system dedicated to heavy-ion beam treatment [J].
Kamada, T ;
Tsujii, H ;
Mizoe, JE ;
Matsuoka, Y ;
Tsuji, H ;
Osaka, Y ;
Minohara, S ;
Miyahara, N ;
Endo, M ;
Kanai, T .
RADIOTHERAPY AND ONCOLOGY, 1999, 50 (02) :235-237
[22]   Design of a post linac for an energy upgrade of a heavy-ion injector [J].
Iwata, Y. ;
Noda, K. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2014, 331 :10-14
[23]   Total nuclear reaction cross-section database for radiation protection in space and heavy-ion therapy applications [J].
Luoni, F. ;
Horst, F. ;
Reidel, C. A. ;
Quarz, A. ;
Bagnale, L. ;
Sihver, L. ;
Weber, U. ;
Norman, R. B. ;
de Wet, W. ;
Giraudo, M. ;
Santin, G. ;
Norbury, J. W. ;
Durante, M. .
NEW JOURNAL OF PHYSICS, 2021, 23 (10)
[24]   Intra- and inter-fraction relative range verification in heavy-ion therapy using filtered interaction vertex imaging [J].
Hymers, Devin ;
Kasanda, Eva ;
Bildstein, Vinzenz ;
Easter, Joelle ;
Richard, Andrea ;
Spyrou, Artemis ;
Hohr, Cornelia ;
Mucher, Dennis .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (24)
[25]   RADIATION PROTECTION STUDY FOR THE FIRST KOREA HEAVY-ION MEDICAL ACCELERATOR FACILITY [J].
Kum, Oyeon ;
Heo, Seung Uk ;
Choi, Sang Hyoun ;
Song, Yongkeun ;
Cho, Sung-Ho .
NUCLEAR TECHNOLOGY, 2015, 192 (03) :208-214
[26]   Alignment of heavy ion therapy center vertical terminal [J].
Chen, Wenjun ;
Yang, Yaqing ;
Cai, Guozhu ;
Zheng, Yajun ;
Wang, Shaoming ;
Yuan, Jiandong .
RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2019, 174 (9-10) :809-818
[27]   Spot scanning control system for heavy ion therapy [J].
Liu, Wufeng ;
Huang, Jijiang ;
Qiao, Weimin ;
Jing, Lan ;
Liu, Nanbo ;
Zhang, Wei ;
Zhang, Shijie .
Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2014, 26 (01)
[28]   Broad-beam three-dimensional irradiation system for heavy-ion radiotherapy at HIMAC [J].
Futami, Y ;
Kanai, T ;
Fujita, M ;
Tomura, H ;
Higashi, A ;
Matsufuji, N ;
Miyahara, N ;
Endo, M ;
Kawachi, K .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 430 (01) :143-153
[29]   Deflection Analysis of a Robotic Bed on the Applied Loads and Its Postures for a Heavy-Ion Therapeutic System [J].
Choi, Hongseok ;
Park, Jong-Oh ;
Ko, Seong Young ;
Park, Sukho .
9TH INTERNATIONAL CONFERENCE ON ROBOTIC, VISION, SIGNAL PROCESSING AND POWER APPLICATIONS: EMPOWERING RESEARCH AND INNOVATION, 2017, 398 :343-350
[30]   Validation of Geant4 fragmentation for Heavy Ion Therapy [J].
Bolst, David ;
Cirrone, Giuseppe A. P. ;
Cuttone, Giacomo ;
Folger, Gunter ;
Incerti, Sebastien ;
Ivanchenko, Vladimir ;
Koi, Tatsumi ;
Mancusi, Davide ;
Pandola, Luciano ;
Romano, Francesco ;
Rosenfeld, Anatoly B. ;
Guatelli, Susanna .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 869 :68-75