Implant-Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis

被引:87
作者
Ding, Xi [1 ]
Zhu, Xing-Hao [1 ]
Liao, Sheng-Hui [2 ]
Zhang, Xiu-Hua [1 ]
Chen, Hong [1 ]
机构
[1] Wenzhou Med Coll, Affiliated Hosp 1, Dept Stomatol, Wenzhou 32500, Zhejiang, Peoples R China
[2] Zhejiang Univ, State Key Lab CAD & CG, Hangzhou, Peoples R China
来源
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY | 2009年 / 18卷 / 05期
关键词
Finite element; biomechanics; dental implants; immediate loading; stress; mandible;
D O I
10.1111/j.1532-849X.2009.00453.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To establish a 3D finite element model of a mandible with dental implants for immediate loading and to analyze stress distribution in bone around implants of different diameters. Materials and Methods: Three mandible models, embedded with thread implants (ITI, Straumann, Switzerland) with diameters of 3.3, 4.1, and 4.8 mm, respectively, were developed using CT scanning and self-developed Universal Surgical Integration System software. The von Mises stress and strain of the implant-bone interface were calculated with the ANSYS software when implants were loaded with 150 N vertical or buccolingual forces. Results: When the implants were loaded with vertical force, the von Mises stress concentrated on the mesial and distal surfaces of cortical bone around the neck of implants, with peak values of 25.0, 17.6 and 11.6 MPa for 3.3, 4.1, and 4.8 mm diameters, respectively, while the maximum strains (5854, 4903, 4344 mu epsilon) were located on the buccal cancellous bone around the implant bottom and threads of implants. The stress and strain were significantly lower (p < 0.05) with the increased diameter of implant. When the implants were loaded with buccolingual force, the peak von Mises stress values occurred on the buccal surface of cortical bone around the implant neck, with values of 131.1, 78.7, and 68.1 MPa for 3.3, 4.1, and 4.8 mm diameters, respectively, while the maximum strains occurred on the buccal surface of cancellous bone adjacent to the implant neck, with peak values of 14,218, 12,706, and 11,504 mu m, respectively. The stress of the 4.1-mm diameter implants was significantly lower (p < 0.05) than those of 3.3-mm diameter implants, but not statistically different from that of the 4.8 mm implant. Conclusions: With an increase of implant diameter, stress and strain on the implant-bone interfaces significantly decreased, especially when the diameter increased from 3.3 to 4.1 mm. It appears that dental implants of 10 mm in length for immediate loading should be at least 4.1 mm in diameter, and uniaxial loading to dental implants should be avoided or minimized.
引用
收藏
页码:393 / 402
页数:10
相关论文
共 52 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]  
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[3]  
Akca K, 2001, Implant Dent, V10, P238
[4]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[5]   ENDOSSEUS TITANIUM IMPLANTS IN EXTRACTION SOCKETS - AN EXPERIMENTAL-STUDY IN MONKEYS [J].
ANNEROTH, G ;
HEDSTROM, KG ;
KJELLMAN, O ;
KONDELL, PA ;
NORDENRAM, A .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1985, 14 (01) :50-54
[6]   Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog [J].
Barbier, L ;
Vander Sloten, J ;
Krzesinski, G ;
Schepers, E ;
Van Der Perre, G .
JOURNAL OF ORAL REHABILITATION, 1998, 25 (11) :847-858
[7]   Radiographic bone density around immediately loaded oral implants - A case series [J].
Barone, A ;
Covani, U ;
Cornelini, R ;
Gherlone, E .
CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (05) :610-615
[8]   Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis [J].
Bozkaya, D ;
Muftu, S ;
Muftu, A .
JOURNAL OF PROSTHETIC DENTISTRY, 2004, 92 (06) :523-530
[9]  
Brunski J B, 1988, J Dent Educ, V52, P775
[10]  
Buser D, 1991, Int J Oral Maxillofac Implants, V6, P405