Effect of Stent-Graft Compliance on Endotension After EVAR

被引:12
作者
Bosman, Willem-Maarten P. F. [1 ]
Hinnen, Jan-Willem [1 ]
Rixen, Daniel J. [2 ]
Hamming, Jaap F. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Surg, Div Vasc Surg, NL-2300 RC Leiden, Netherlands
[2] Delft Univ Technol, Fac Mech Maritime & Mat Engn Engn Dynam, Delft, Netherlands
关键词
abdominal aortic aneurysm; endovascular aneurysm repair; aneurysm sac pressure; pressure sensor; in-vitro study; endotension; pressure transmission; ANEURYSM SAC PRESSURE; ENDOVASCULAR REPAIR; TRANSMISSION;
D O I
10.1583/08-2505.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To investigate whether, and to which extent, the type of graft has influence on the pressure in the aneurysm sac. Methods: The study consisted of a dynamic and a static experiment. In the dynamic experiment, a latex aneurysm was inserted into an in-vitro circulation model, with variable (80-180 mmHg) systemic systolic pressure and a pulse pressure of 40 mmHg. Seven types of devices were used to exclude the aneurysm: a latex reference graft, 3 stentless grafts made of different fabrics, and 3 different types of commercially produced stented grafts used for endovascular aneurysm repair (EVAR). The systolic and diastolic intra-aneurysm pressures OAP) were measured, along with the pulse pressure. The mean IAP and pulse pressures were compared for each category of graft (stented/stentless) and for each graft. In the static experiments, the compliance of each graft and the latex aneurysm Were determined by measuring the change in volume necessary to obtain a change in pressure. Furthermore, the IAP with each graft was measured in a nonpulsatile situation under systolic pressures varying between 60 and 140 mmHg. Results: The experiments showed that, with increasing systemic pressures, there was a small pressure (<5 mmHg) increase in the aneurysm, but there was no significant difference among grafts in the dynamic or the static experiments (p=0.12 and p=0.17, respectively). With the 3 stented EVAR grafts, there was less pressure transmission than with the 3 stentless grafts, but this difference was not significant (p>0.05). Conclusion: This in-vitro study demonstrated that there is a clinically irrelevant pressure transmission through the graft wall and no significant difference in the pressure transmission between stented and stentless grafts.
引用
收藏
页码:105 / 113
页数:9
相关论文
共 20 条
[1]   Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, RP ;
Cope, C ;
Golden, MA ;
Velazquez, OC ;
Neschis, DG ;
Mitchell, ME ;
Barker, CF ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :32-40
[2]   Reporting standards for endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Blankensteijn, JD ;
Harris, PL ;
White, GH ;
Zarins, CK ;
Bernhard, VM ;
Matsumura, JS ;
May, J ;
Veith, FJ ;
Fillinger, MF ;
Rutherford, RB ;
Kent, KC .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1048-1060
[3]   Direct intra-aneurysm sac pressure measurement using tip-pressure sensors: In vivo and in vitro evaluation [J].
Dias, NV ;
Ivancev, K ;
Malina, M ;
Hinnen, JW ;
Visser, M ;
Lindblad, B ;
Sonesson, B .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (04) :711-716
[4]  
Gawenda M, 2003, J ENDOVASC THER, V10, P1091, DOI 10.1583/1545-1550(2003)010<1091:EIIBAV>2.0.CO
[5]  
2
[6]   Endotension is influenced by wall compliance in a latex aneurysm model [J].
Gawenda, M ;
Knez, P ;
Winter, S ;
Jaschke, G ;
Wassmer, G ;
Schmitz-Rixen, T ;
Brunkwall, J .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2004, 27 (01) :45-50
[7]   Endotension as a result of pressure transmission through the graft following endovascular aneurysm repair - An in vitro study [J].
Gawenda, M ;
Jaschke, G ;
Winter, S ;
Wassmer, G ;
Brunkwall, J .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2003, 26 (05) :501-505
[8]   Intra-aneurysm sac pressure - The holy grail of endoluminal grafting of AAA [J].
Gawenda, M ;
Heckenkamp, J ;
Zaehringer, M ;
Brunkwall, J .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2002, 24 (02) :139-145
[9]  
Gilling-Smith G, 1999, J ENDOVASC SURG, V6, P305, DOI 10.1583/1074-6218(1999)006<0305:EAEARD>2.0.CO
[10]  
2