Spectra of photons and neutrons generated in a heterogeneous head of a 15 MV LINAC at differents field sizes

被引:1
|
作者
Benites-Rengifo, J. L. [1 ]
Vega-Carrillo, H. R. [2 ]
Velazquez-Fernandez, J. B. [1 ]
机构
[1] Univ Autonoma Nayarit, Posgrado & CBAP, Carretera Tep Compostela Km 9 CP 63780, Xalisco, Nayarit, Mexico
[2] Universidad Autonoma Zacatecas, Unidad Acad Estudios Nucl, Monaco 98000, Monaco
来源
MEDICAL PHYSICS | 2012年 / 1494卷
关键词
LINAC; Neutrons; Monte Carlo;
D O I
10.1063/1.4764585
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Spectra of photons and neutrons were calculated, using the Monte Carlo code MCNP-5 using the e/p/n mode. A heterogeneous model was used to define the linac head where the collimators were modeled to produce five different treatment fields at the isocenter. Photon and neutron spectra were estimated in several points along two directions from the isocenter. The total photon fluence beyond 60 cm behaves according to 1/r(2) rule, while total neutron fluence, beyond 80 cm, can be described by diffusion theory using an infinite plane as a neutron source.
引用
收藏
页码:8 / 11
页数:4
相关论文
共 14 条
  • [1] Energy Spectra for 6 and 15 MV Beams Generated by Truebeam Linac
    Liu, J.
    Cozza, S.
    Chang, J.
    Baker, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E619 - E619
  • [2] Neutron Spectra in a 15 MV LINAC
    Vega-Carrillo, H. R.
    Chu, Wei-Han
    Tung, Chuan-Jong
    Lan, Jen-Hong
    ELEVENTH MEXICAN SYMPOSIUM ON MEDICAL PHYSICS, 2010, 1310 : 158 - +
  • [3] Neutron Spectra and H*(10) in a 15 MV Linac
    Benites, J.
    Vega-Carrillo, H. R.
    Hernandez-Davila, V. M.
    Rivera, T.
    Carrillo, A.
    Mondragon, R.
    ELEVENTH MEXICAN SYMPOSIUM ON MEDICAL PHYSICS, 2010, 1310 : 44 - +
  • [4] Ferrous xylenol gel measurements for 6 and 10 MV photons in small field sizes
    de Oliveiraa, Lucas N.
    Calcina, Carmen S. Guzman
    Parada, Marco A.
    de Almeida, Carlos E.
    de Almeidaa, Adelaide
    BRAZILIAN JOURNAL OF PHYSICS, 2007, 37 (3B) : 1141 - 1146
  • [5] Evaluation of equivalent dose from neutrons and activation products from a 15-MV X-ray LINAC
    Israngkul-Na-Ayuthaya, Isra
    Suriyapee, Sivalee
    Pengvanich, Phongpheath
    JOURNAL OF RADIATION RESEARCH, 2015, 56 (06) : 919 - 926
  • [6] Treatment of Large-Size Prostate Cancer Patients On An O-Ring Linac with 6MV FFF: Plan Quality Comparison with a C-Arm Linac Using 10MV Or 15MV Photons
    Liu, B.
    Cheng, C.
    Wang, X.
    Zhu, J.
    Parikh, R.
    Nie, K.
    Yue, N.
    Zhang, Y.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [7] Estimation of neutron production from accelerator head assembly of 15 MV medical LINAC using FLUKA simulations
    Patil, B. J.
    Chavan, S. T.
    Pethe, S. N.
    Krishnan, R.
    Bhoraskar, V. N.
    Dhole, S. D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2011, 269 (24): : 3261 - 3265
  • [8] Energy spectra in water for the 6 MV X-ray therapeutic beam generated by Clinac-2300 linac
    Konefal, Adam
    Bakoniak, Marzena
    Orlef, Andrzej
    Maniakowski, Zbigniew
    Szewczuk, Marek
    RADIATION MEASUREMENTS, 2015, 72 : 12 - 22
  • [9] MONTE CARLO STUDY OF THE OUT-OF-FIELD SPECTRA FOR 6MV AND 15MV PHOTON BEAMS WITH AND WITHOUT MLC.
    Theodorou, K.
    Paramerits, I.
    Tsiamas, P.
    Kappas, C.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S411 - S411
  • [10] Small Field Dosimetric Characterization of 7MV FFF Photons for Stereotactic MR-Guided Adaptive Radiotherapy Using the Unity MR-Linac
    Hwang, M.
    Chen, X.
    Oh, S.
    Lee, D.
    Pavord, D.
    Coopey, B.
    Sohn, J.
    MEDICAL PHYSICS, 2022, 49 (06) : E952 - E952