A study of respiration-correlated cone-beam CT scans to correct target positioning errors in radiotherapy of thoracic cancer

被引:7
作者
Santoro, J. P. [1 ]
McNamara, J. [1 ]
Yorke, E. [1 ]
Pham, H. [1 ]
Rimner, A. [2 ]
Rosenzweig, K. E. [2 ]
Mageras, G. S. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10065 USA
关键词
cone-beam computed tomography; image-guided radiation treatment; organ motion; lung cancer; IMAGE-GUIDED RADIOTHERAPY; STEREOTACTIC BODY RADIOTHERAPY; COMPUTED-TOMOGRAPHY; LUNG-CANCER; MOTION; LOCALIZATION; ACCURACY; VOLUME; SETUP;
D O I
10.1118/1.4748503
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: There is increasingly widespread usage of cone-beam CT (CBCT) for guiding radiation treatment in advanced-stage lung tumors, but difficulties associated with daily CBCT in conventionally fractionated treatments include imaging dose to the patient, increased workload and longer treatment times. Respiration-correlated cone-beam CT (RC-CBCT) can improve localization accuracy in mobile lung tumors, but further increases the time and workload for conventionally fractionated treatments. This study investigates whether RC-CBCT-guided correction of systematic tumor deviations in standard fractionated lung tumor radiation treatments is more effective than 2D image-based correction of skeletal deviations alone. A second study goal compares respiration-correlated vs respiration-averaged images for determining tumor deviations. Methods: Eleven stage II-IV nonsmall cell lung cancer patients are enrolled in an IRB-approved prospective off-line protocol using RC-CBCT guidance to correct for systematic errors in GTV position. Patients receive a respiration-correlated planning CT (RCCT) at simulation, daily kilovoltage RC-CBCT scans during the first week of treatment and weekly scans thereafter. Four types of correction methods are compared: (1) systematic error in gross tumor volume (GTV) position, (2) systematic error in skeletal anatomy, (3) daily skeletal corrections, and (4) weekly skeletal corrections. The comparison is in terms of weighted average of the residual GTV deviations measured from the RC-CBCT scans and representing the estimated residual deviation over the treatment course. In the second study goal, GTV deviations computed from matching RCCT and RC-CBCT are compared to deviations computed from matching respiration-averaged images consisting of a CBCT reconstructed using all projections and an average-intensity-projection CT computed from the RCCT. Results: Of the eleven patients in the GTV-based systematic correction protocol, two required no correction, seven required a single correction, one required two corrections, and one required three corrections. Mean residual GTV deviation (3D distance) following GTV-based systematic correction (mean 1 standard deviation 4.8 +/- 1.5 mm) is significantly lower than for systematic skeletal-based (6.5 +/- 2.9 mm, p = 0.015), and weekly skeletal-based correction (7.2 +/- 3.0 mm, p = 0.001), but is not significantly lower than daily skeletal-based correction (5.4 +/- 2.6 mm, p = 0.34). In two cases, first-day CBCT images reveal tumor changes one showing tumor growth, the other showing large tumor displacement that are not readily observed in radiographs. Differences in computed GTV deviations between respiration-correlated and respiration-averaged images are 0.2 +/- 1.8 mm in the superior-inferior direction and are of similar magnitude in the other directions. Conclusions: An off-line protocol to correct GTV-based systematic error in locally advanced lung tumor cases can be effective at reducing tumor deviations, although the findings need confirmation with larger patient statistics. In some cases, a single cone-beam CT can be useful for assessing tumor changes early in treatment, if more than a few days elapse between simulation and the start of treatment. Tumor deviations measured with respiration-averaged CT and CBCT images are consistent with those measured with respiration-correlated images; the respiration-averaged method is more easily implemented in the clinic. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4748503]
引用
收藏
页码:5825 / 5834
页数:10
相关论文
共 26 条
  • [1] QUANTIFYING INTERFRACTION AND INTRAFRACTION TUMOR MOTION IN LUNG STEREOTACTIC BODY RADIOTHERAPY USING RESPIRATION-CORRELATED CONE BEAM COMPUTED TOMOGRAPHY
    Bissonnette, Jean-Pierre
    Franks, Kevin N.
    Purdie, Thomas G.
    Moseley, Douglas J.
    Sonke, Jan-Jakob
    Jaffray, David A.
    Dawson, Laura A.
    Bezjak, Andrea
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : 688 - 695
  • [2] CONE-BEAM COMPUTED TOMOGRAPHIC IMAGE GUIDANCE FOR LUNG CANCER RADIATION THERAPY
    Bissonnette, Jean-Pierre
    Purdie, Thomas G.
    Higgins, Jane A.
    Li, Winnie
    Bezjak, Andrea
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (03): : 927 - 934
  • [3] Observation of interfractional variations in lung tumor position using respiratory gated and ungated megavoltage cone-beam computed tomography
    Chang, Jenghwa
    Mageras, Gig S.
    Yorke, Ellen
    De Arruda, Fernando
    Sillanpaa, Jussi
    Rosenzweig, Kenneth E.
    Hertanto, Agung
    Pham, Hai
    Seppi, Edward
    Pevsner, Alex
    Ling, C. Clifton
    Amols, Howard
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05): : 1548 - 1558
  • [4] RADIATION DOSE FROM KILOVOLTAGE CONE BEAM COMPUTED TOMOGRAPHY IN AN IMAGE-GUIDED RADIOTHERAPY PROCEDURE
    Ding, George X.
    Coffey, Charles W.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 610 - 617
  • [5] Respiration-correlated spiral CT: A method of measuring respiratory-induced anatomic motion for radiation treatment planning
    Ford, EC
    Mageras, GS
    Yorke, E
    Ling, CC
    [J]. MEDICAL PHYSICS, 2003, 30 (01) : 88 - 97
  • [6] GinsbergRJ, 2001, Cancer Principles and Practice of Oncology, V6th, P925
  • [7] Image-guided radiotherapy via daily online cone-beam CT substantially reduces margin requirements for stereotactic lung radiotherapy
    Grills, Inga S.
    Hugo, Geoffrey
    Kestfn, Larry L.
    Galerani, Ana Paula
    Chao, K. Kenneth
    Wloch, Jennifer
    Yan, Di
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (04): : 1045 - 1056
  • [8] Dosimetric advantages of four-dimensional adaptive image-guided radiotherapy for lung tumors using online cone-beam computed tomography
    Harsolia, Asif
    Hugo, Geoffrey D.
    Kestin, Larry L.
    Grills, Inga S.
    Yan, Di
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 582 - 589
  • [9] EFFECT OF IMAGE-GUIDANCE FREQUENCY ON GEOMETRIC ACCURACY AND SETUP MARGINS IN RADIOTHERAPY FOR LOCALLY ADVANCED LUNG CANCER
    Higgins, Jane
    Bezjak, Andrea
    Hope, Andrew
    Panzarella, Tony
    Li, Winnie
    Cho, John B. C.
    Craig, Tim
    Brade, Anthony
    Sun, Alexander
    Bissonnette, Jean-Pierre
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (05): : 1330 - 1337
  • [10] On-line target position localization in the presence of respiration: A comparison of two methods
    Hugo, Geoffrey D.
    Liang, Jian
    Campbell, Jonathan
    Yan, Di
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (05): : 1634 - 1641