Does IGRT ensure target dose coverage of head and neck IMRT patients?

被引:24
作者
Graff, Pierre
Hu, Weigang [2 ]
Yom, Sue S.
Pouliot, Jean [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, Helen Diller Family Comprehens Canc Ctr, San Francisco, CA 94143 USA
[2] Fudan Univ, Dept Radiat Oncol, Shanghai 200433, Peoples R China
关键词
Head and neck IMRT; Target coverage; Dose recalculation; IGRT; MVCBCT; CONE-BEAM CT; RADIATION-THERAPY; CANCER-PATIENTS; RADIOTHERAPY; REGISTRATION; RECALCULATION; STRATEGIES; DOSIMETRY; GUIDANCE; SYSTEM;
D O I
10.1016/j.radonc.2011.09.024
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To determine if image-guided radiotherapy (IGRT) ensures dose coverage to the target, and to assess the dosimetric impact of anatomic changes using megavoltage cone-beam CT (MVCBCT) for patient positioning during head and neck IMRT. Methods and materials: Forty-eight MVCBCT from 10 head and neck IMRT/IGRT patients were analyzed off-line. Target volumes and organs at risk (OARS) contours delineated on CT were transferred and adjusted on MVCBCT images. Each MVCBCT was processed to allow dose recalculation, resulting in 469 dose-volume histograms (DVHs). The concept of dosimetric latitude was introduced to provide a clinical perspective. Results: MVCBCT target DVHs showed a moderate level of difference in D95 (dose to >= 95% of volume), generally less than a 5% difference from the planned dose. Delivered-dose increases to the spinal cord and brainstem showed no apparent time trend. The 4 mm margin around OARs was a useful precaution to prevent exceeding critical dose thresholds. The parotid glands showed progressive increases in mean dose related to shrinkage of the external contours. Conclusion: IGRT repositioning ensured target volume coverage, but significant dose variations were observed for OARs. The dosimetric impact of anatomic changes during radiotherapy was of lesser importance than the effects of IGRT repositioning. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 104 (2012) 83-90
引用
收藏
页码:83 / 90
页数:8
相关论文
共 27 条
  • [1] Aubry J.-F., 2008, IMAGING DECISIONS MR, V12, P3
  • [2] Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region
    Aubrya, Jean-Francois
    Pouliot, Jean
    Beaulieu, Luc
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 900 - 907
  • [3] Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system
    Barker, JL
    Garden, AS
    Ang, KK
    O'Daniel, JC
    Wang, H
    Court, LE
    Morrison, WH
    Rosenthal, DI
    Chao, KSC
    Tucker, SL
    Mohan, R
    Dong, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 960 - 970
  • [4] Analysis of the rigid and deformable component of setup inaccuracies on portal images in head and neck radiotherapy
    Birkner, Mattias
    Thorwarth, Daniela
    Poser, Alexander
    Ammazzalorso, Filippo
    Alber, Markus
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (18) : 5721 - 5733
  • [5] Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors
    Castadot, Pierre
    Lee, John Aldo
    Parraga, Adriane
    Geets, Xavier
    Macq, Benoit
    Gregoire, Vincent
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 89 (01) : 1 - 12
  • [6] Assessment by a deformable registration method of the volumetric and positional changes of target volumes and organs at risk in pharyngo-laryngeal tumors treated with concomitant chemo-radiation
    Castadot, Pierre
    Geets, Xavier
    Lee, John Aldo
    Christian, Nicolas
    Gregoire, Vincent
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 95 (02) : 209 - 217
  • [7] DOSE RECALCULATION AND THE DOSE-GUIDED RADIATION THERAPY (DGRT) PROCESS USING MEGAVOLTAGE CONE-BEAM CT
    Cheung, Joey
    Aubry, Jean-Francois
    Yom, Sue S.
    Gottschalk, Alexander R.
    Celi, Juan Carlos
    Pouliot, Jean
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (02): : 583 - 592
  • [8] Adaptive biological image-guided IMRT with anatomic and functional imaging in pharyngo-laryngeal tumors:: Impact on target volume delineation and dose distribution using helical tomotherapy
    Geets, Xavier
    Tomsej, Milan
    Lee, John A.
    Duprez, Thierry
    Coche, Emmanuel
    Cosnard, Guy
    Lonneux, Max
    Gregoire, Vincent
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 85 (01) : 105 - 115
  • [9] Repeat CT imaging and replanning during the course of IMRT for head-and-neck cancer
    Hansen, EK
    Bucci, MK
    Quivey, JM
    Weinberg, V
    Xia, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (02): : 355 - 362
  • [10] Physics strategies for sparing neural stem cells during whole-brain radiation treatments
    Kirby, Neil
    Chuang, Cynthia
    Pouliot, Jean
    Hwang, Andrew
    Barani, Igor J.
    [J]. MEDICAL PHYSICS, 2011, 38 (10) : 5338 - 5344