Recovery coefficients determination for partial volume effect correction in oncological PET/CT images considering the effect of activity outside the field of view

被引:4
作者
Krempser, Alexandre R. [1 ]
Ichinose, Roberto M. [1 ]
Miranda de Sa, Antonio M. F. L. [1 ]
Velasques de Oliveira, Silvia M. [2 ]
Carneiro, Michel P. [3 ]
机构
[1] Univ Fed Rio de Janeiro, Biomed Engn Program, COPPE, Ilha Fundao, BR-21941914 Rio de Janeiro, RJ, Brazil
[2] Inst Radioprotecao & Dosimetria, CNEN, Inst Radiat Protect & Dosimetry, Rio de Janeiro, RJ, Brazil
[3] PET CT Clin, Fleury Labs, Rio De Janeiro, RJ, Brazil
关键词
PET/CT; Partial volume effect correction; Recovery coefficient; Activity quantification; CANCER; THERAPY; SCANNER; IMPACT;
D O I
10.1007/s12149-013-0773-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The partial volume effect (PVE) has a great impact in quantitative PET/CT imaging. Correction methods have been recently proposed by many authors to make the image quantification more accurate. This work presents a methodology for determining the recovery coefficients (RCs) for PVE correction in PET/CT images. It was taken into account the radioactivity outside the field of view (FOV), which is expected in a patient image acquisition. The NEMA image quality phantom and the NEMA scatter phantom were used. The phantoms were filled with F-18-FDG for different sphere-to-background ratios. The RCs have been determined from image acquisitions in a Siemens Biograph 16 Hi-Rez PET/CT scanner with and without the scatter phantom. The RC values that ranged from 0.38 to 1.00 without the scatter phantom exhibited a wider variation when this latter was taken into account (from 0.27 to 1.02). This more realistic estimation must be considered if one takes into account that an incorrect SUV measure in tumors leads to errors in the evaluation of the response to therapy based on PET/CT images. The activity outside the FOV should be considered in RCs determination to improve the RC-based PVE correction method.
引用
收藏
页码:924 / 930
页数:7
相关论文
共 21 条
[1]   A Systematic Review of the Factors Affecting Accuracy of SUV Measurements [J].
Adams, Michael C. ;
Turkington, Timothy G. ;
Wilson, Joshua M. ;
Wong, Terence Z. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (02) :310-320
[2]  
BJOERN W, 2009, IEEE T NUCL SCI, V56, P633
[3]   Standards for PET Image Acquisition and Quantitative Data Analysis [J].
Boellaard, Ronald .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 :11S-20S
[4]  
Erdi YE, 2004, J NUCL MED, V45, P813
[5]   PVE Correction in PET-CT Whole-Body Oncological Studies From PVE-Affected Images [J].
Gallivanone, Francesca ;
Stefano, Alessandro ;
Grosso, Eleonora ;
Canevari, Carla ;
Gianolli, Luigi ;
Messa, Cristina ;
Gilardi, Maria Carla ;
Castiglioni, Isabella .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (03) :736-747
[6]   Impact of Partial-Volume Effect Correction on the Predictive and Prognostic Value of Baseline 18F-FDG PET Images in Esophageal Cancer [J].
Hatt, Mathieu ;
Le Pogam, Adrien ;
Visvikis, Dimitris ;
Pradier, Olivier ;
Le Rest, Catherine Cheze .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (01) :12-20
[7]   Partial volume correction strategies for quantitative FDG PET in oncology [J].
Hoetjes, Nikie J. ;
van Velden, Floris H. P. ;
Hoekstra, Otto S. ;
Hoekstra, Corneline J. ;
Krak, Nanda C. ;
Lammertsma, Adriaan A. ;
Boellaard, Ronald .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (09) :1679-1687
[8]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .1. EFFECT OF OBJECT SIZE [J].
HOFFMAN, EJ ;
HUANG, SC ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (03) :299-308
[9]   Iodine-124 PET dosimetry in differentiated thyroid cancer:: recovery coefficient in 2D and 3D modes for PET(/CT) systems [J].
Jentzen, Walter ;
Weise, Reiner ;
Kupferschlaeger, Juergen ;
Freudenberg, Lutz ;
Brandau, Wolfgang ;
Bares, Ronald ;
Burchert, Wolfgang ;
Bockisch, Andreas .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (03) :611-623
[10]   ANALYSIS OF EMISSION TOMOGRAPHIC SCAN DATA - LIMITATIONS IMPOSED BY RESOLUTION AND BACKGROUND [J].
KESSLER, RM ;
ELLIS, JR ;
EDEN, M .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (03) :514-522