Comparison of temporal artery, nasopharyngeal, and axillary temperature measurement during anesthesia in children

被引:12
|
作者
Sahin, Sevtap Hekimoglu [1 ]
Duran, Ridvan [2 ]
Sut, Necdet [3 ]
Colak, Alkin [1 ]
Acunas, Betul [2 ]
Aksu, Burhan [4 ]
机构
[1] Trakya Univ, Fac Med, Dept Anaesthesiol & Reanimat, TR-22030 Edirne, Turkey
[2] Trakya Univ, Fac Med, Dept Pediat, TR-22030 Edirne, Turkey
[3] Trakya Univ, Fac Med, Dept Biostat, TR-22030 Edirne, Turkey
[4] Trakya Univ, Fac Med, Dept Pediat Surg, TR-22030 Edirne, Turkey
关键词
Anesthesia; pediatric; Temperature measurement; Axillary nasopharyngeal; Temporal artery; INFANTS; THERMOMETER; RELIABILITY; AGREEMENT; ACCURACY; NEWBORN; SKIN;
D O I
10.1016/j.jclinane.2012.05.003
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Study Objective: To evaluate the accuracy and precision of a new, noninvasive infrared thermometer applied to the temporal artery. Design: Prospective randomized study. Setting: Trakya University Hospital. Patients: 60 ASA physical status 1 and 2 children undergoing surgery. Interventions: During anesthesia, temperature measurements were recorded with three different techniques: temporal artery, nasopharynx, and axillary temperature. Measurements: Temperatures measured from the nasopharynx, temporal artery, and the axilla were recorded at 15-minute intervals for the first hour, then at 30-minute intervals until the completion of surgery. During each measurement, heart rate and midarterial pressure were recorded. Main Results: There were no statistically significant differences between temperatures recorded at the temporal artery and nasopharynx at 15, 30, 45, 60, 90, and 120 minutes, and the completion of surgery. Axillary temperatures were statistically lower than those recorded at the nasopharynx and the temporal artery (P < 0.001). Bland-Altman plots showed a correlation of temperature measurements between the temporal artery and nasopharyngeal methods. The axillary method had a lower correlation with the temporal artery and the nasopharyngeal methods. Conclusions: The temporal artery thermometer is a substitute for the nasopharyngeal thermometer for core temperature measurement during anesthesia in children. (C) 2012 Elsevier Inc. All rights reserved.
引用
收藏
页码:647 / 651
页数:5
相关论文
共 50 条
  • [1] Nasopharyngeal temperature measurement in sheep during general anesthesia
    Tan, Tabita
    Tuke, Jonathon
    Musk, Gabrielle C.
    CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 2017, 81 (01): : 64 - 68
  • [2] Temporal artery and axillary thermometry comparison with rectal thermometry in children presenting to the ED
    Forrest, Adam J.
    Juliano, Michael L.
    Conley, Sean P.
    Cronyn, Patrick D.
    McGlynn, Andrea
    Auten, Jonathan D.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2017, 35 (12): : 1855 - 1858
  • [3] MEASUREMENT OF RETINAL ARTERY BLOOD PRESSURE DURING ANESTHESIA IN CHILDREN
    FARMATI, O
    FREEMAN, A
    MOYA, F
    SOUTHERN MEDICAL JOURNAL, 1966, 59 (11) : 1297 - +
  • [4] Comparison of temporal artery to mercury and digital temperature measurement in pediatrics
    Isler, Aysegul
    Aydin, Resmiye
    Guven, Serife Tutar
    Gunay, Sema
    INTERNATIONAL EMERGENCY NURSING, 2014, 22 (03) : 165 - 168
  • [5] Comparison of Axillary and Temporal Artery Thermometry in Preterm Neonates
    Smith, Sandra
    Keltner, Carlotta
    Stikes, Reetta
    Hayes, Pauline
    Crawford, Timothy N.
    JOGNN-JOURNAL OF OBSTETRIC GYNECOLOGIC AND NEONATAL NURSING, 2018, 47 (03): : 352 - 361
  • [6] Comparison of Temporal Artery and Axillary Temperatures in Healthy Newborns
    Haddad, Lisa
    Smith, Sarah
    Phillips, Kenneth D.
    Heidel, R. Eric
    JOGNN-JOURNAL OF OBSTETRIC GYNECOLOGIC AND NEONATAL NURSING, 2012, 41 (03): : 383 - 388
  • [7] Temporal Artery Temperature Measurement in the Neonate
    Syrkin-Nikolau, Mashette E.
    Johnson, Karen J.
    Colaizy, Tarah T.
    Schrock, Ruthann
    Bell, Edward F.
    AMERICAN JOURNAL OF PERINATOLOGY, 2017, 34 (10) : 1026 - 1031
  • [8] Comparison of axillary with rectal and oral temperature in children
    Sohail, Shagufta
    Shama, Qurat-ul-ain
    RAWAL MEDICAL JOURNAL, 2016, 41 (01): : 57 - 60
  • [9] Comparison of Temporal Artery and Rectal Temperature Measurement During Cooling and Rewarming in Neonates Treated for Hypoxic Ischemic Encephalopathy
    Labella, Michelle
    Kelly, Karen
    Carlin, Kristen
    Walsh, Elaine
    ADVANCES IN NEONATAL CARE, 2025, 25 (01) : 92 - 98
  • [10] Temperature Measurement in Pediatrics: A Comparison of the Rectal Method Versus the Temporal Artery Method
    Bahorski, Jessica
    Repasky, Terri
    Ranner, Donna
    Fields, Ally
    Jackson, Michelle
    Moultry, Lucy
    Pierce, Karen
    Sandell, Mary
    JOURNAL OF PEDIATRIC NURSING-NURSING CARE OF CHILDREN & FAMILIES, 2012, 27 (03): : 243 - 247