Evaluation of a Bi-Planar Robot Navigation System for Insertion of Cannulated Screws in Femoral Neck Fractures

被引:51
作者
He, Meng [1 ]
Han, Wei [1 ]
Zhao, Chun-peng [1 ]
Su, Yong-gang [1 ]
Zhou, Li [1 ]
Wu, Xin-bao [1 ]
Wang, Jun-qiang [1 ]
机构
[1] Beijing Jishuitan Hosp, Dept Trauma Orthopaed, Beijing 100035, Peoples R China
关键词
Femoral neck fractures; Bi-planar robot navigation system; Cannulated screw fixation; COMPUTER-ASSISTED NAVIGATION; INTERNAL-FIXATION; HIP; PLACEMENT;
D O I
10.1111/os.12450
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective To evaluate the bi-planar robot navigation system for insertion of cannulated screws in femoral neck fractures. Method Between January 2016 and December 2016, 60 patients with femoral neck fractures were separately treated using percutaneous cannulated screws assisted by the bi-planar robot navigation system (robot group) and conventional freehand surgery (freehand group). The fluoroscopy time, the number of drilling attempts, and the operation time were recorded during operations; the dispersion and parallelism of the cannulated screws on the posteroanterior and lateral images were measured after operations. Patients were followed up for 12-24 months and the Harris scores and the final results of the two groups were compared. Results During bi-planar robot navigation system-assisted surgery, the fluoroscopy time for acquisition of images was 2.3 seconds on average, and the time for planning screws during the operation was 2.8 min on average. The average fluoroscopy time during the placement of the guide pin was 5.7 seconds and 14.14 seconds (P = 0.00), respectively. The average time of the placement of the cannulated screws was 12.7 min and 19.4 min (P = 0.00), respectively, in the robot group and the freehand group. In the robot group, only one guide pin was replaced during the operation, and the average number of adjustments for each guide pin was 2.39 in the freehand group. The screw parallelism and dispersion measured by postoperative imaging in the robot group were significantly superior to those in the freehand group. From postoperative CT it was evident that there were 5 cases of screws exiting the posterior cortex in both groups. During the follow-up phase, 1 case of femoral head necrosis and 5 cases of femoral neck shortening of more than 10 mm occurred in the robotic navigation group; 3 cases of femoral head necrosis, 1 case of fracture nonunion, and 2 cases of shortening of more than 10 mm occurred in the freehand group. At 18 months after surgery, the average Harris scores of the patients were 85.20 and 83.45, respectively, with no significant difference. Conclusion Using bi-planar robot navigation system-assisted placement of femoral neck cannulated screws can significantly reduce the time of intraoperative fluoroscopy, drilling attempts, and operation time. The placed screws are superior to the screws placed freehand in relation to parallelism and dispersion. However, it is still necessary for surgeons to have a good reduction of the femoral neck fracture before surgery and to be proficient in the operation of the robot navigation system. In summary, the bi-planar robot navigation system is an effective assistant instrument for surgery.
引用
收藏
页码:373 / 379
页数:7
相关论文
共 18 条
[1]  
Booth KC, 1998, ORTHOPEDICS, V21, P1173
[2]   Challenges in the Treatment of Femoral Neck Fractures in the Nonelderly Adult [J].
Davidovitch, Roy I. ;
Jordan, Charles J. ;
Egol, Kenneth A. ;
Vrahas, Mark S. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 68 (01) :236-242
[3]   Unique stability of femoral neck fractures treated with the novel biplane double-supported screw fixation method: A biomechanical cadaver study [J].
Filipov, Orlin ;
Gueorguiev, Boyko .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (02) :218-226
[4]   Hemiarthroplasty or internal fixation for intracapsular displaced femoral neck fractures: randomised controlled trial [J].
Frihagen, Frede ;
Nordsletten, Lars ;
Madsen, Jan Erik .
BMJ-BRITISH MEDICAL JOURNAL, 2007, 335 (7632) :1251-1254
[5]   Anatomy of the medial femoral circumflex artery and its surgical implications [J].
Gautier, E ;
Ganz, K ;
Krügel, N ;
Gill, T ;
Ganz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (05) :679-683
[6]   Safety of computer-assisted surgery for cannulated hip screws [J].
Hamelinck, H. K. M. ;
Haagmans, M. ;
Snoeren, M. M. ;
Biert, J. ;
van Vugt, A. B. ;
Frolke, J. P. M. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (455) :241-245
[7]   Computerized navigation for the internal fixation of femoral neck fractures [J].
Liebergall, Meir ;
Ben-David, Dror ;
Weil, Yoram ;
Peyser, Amos ;
Mosheiff, Rami .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (08) :1748-1754
[8]  
Maurer Stephen G, 2003, Am J Orthop (Belle Mead NJ), V32, P438
[9]   Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study [J].
Mueller, Marcus Christian ;
Belei, Peter ;
Pennekamp, Peter H. ;
Kabir, Koroush ;
Wirtz, Dieter C. ;
Burger, Christof ;
Weber, Oliver .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (07) :1463-1469
[10]   Basic principles of CAOS [J].
Nolte, LP ;
Beutler, T .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 :6-16