An analysis of tracking error in image-guided neurosurgery

被引:17
作者
Gerard, Ian J. [1 ]
Collins, D. Louis [1 ]
机构
[1] McGill Univ, Montreal Neurol Inst & Hosp, McConnell Brain Imaging Ctr, Montreal, PQ H3A 2B4, Canada
关键词
Image-guided neurosurgery; Tracking error; Registration error; Calibration; FIDUCIAL LOCALIZATION ERROR; POINT-BASED REGISTRATION; ELECTROMAGNETIC TRACKING; BRAIN SHIFT; DISPLACEMENT; DEFORMATION;
D O I
10.1007/s11548-014-1145-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study quantifies some of the technical and physical factors that contribute to error in image-guided interventions. Errors associated with tracking, tool calibration and registration between a physical object and its corresponding image were investigated and compared with theoretical descriptions of these errors. A precision milled linear testing apparatus was constructed to perform the measurements. The tracking error was shown to increase in linear fashion with distance normal to the camera, and the tracking error ranged between 0.15 and 0.6 mm. The tool calibration error increased as a function of distance from the camera and the reference tool (0.2-0.8 mm). The fiducial registration error was shown to improve when more points were used up until a plateau value was reached which corresponded to the total fiducial localization error (0.8 mm). The target registration error distributions followed a distribution with the largest error and variation around fiducial points. To minimize errors, tools should be calibrated as close as possible to the reference tool and camera, and tools should be used as close to the front edge of the camera throughout the intervention, with the camera pointed in the direction where accuracy is least needed during surgery.
引用
收藏
页码:1579 / 1588
页数:10
相关论文
共 18 条
[1]  
Cuchet E, 1995, J Image Guid Surg, V1, P198, DOI 10.1002/(SICI)1522-712X(1995)1:4<198::AID-IGS2>3.0.CO
[2]  
2-5
[3]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[4]  
Fonov VS, 2010, LECT NOTES COMPUT SC, V6326, P324
[5]  
HARTOV A, 1999, NEUROSURG FOCUS, V6
[6]   Measurement of intraoperative brain surface deformation under a craniotomy [J].
Hill, DLG ;
Maurer, CR ;
Maciunas, RJ ;
Barwise, JA ;
Fitzpatrick, JM ;
Wang, MY .
NEUROSURGERY, 1998, 43 (03) :514-526
[7]   Design and application of an assessment protocol for electromagnetic tracking systems [J].
Hummel, JB ;
Bax, MR ;
Figl, ML ;
Kang, Y ;
Maurer, C ;
Birkfellner, WW ;
Bergmann, H ;
Shahidi, R .
MEDICAL PHYSICS, 2005, 32 (07) :2371-2379
[8]  
Khadem R, 2000, Comput Aided Surg, V5, P98, DOI 10.3109/10929080009148876
[9]  
Labadie Robert F, 2005, Curr Opin Otolaryngol Head Neck Surg, V13, P27, DOI 10.1097/00020840-200502000-00008
[10]   The Study of Fiducial Localization Error of Image in Point-based Registration [J].
Liu, Wenbo ;
Ding, Hui ;
Han, Hongyan ;
Xue, Qinghua ;
Sun, Zhaohui ;
Wang, Guangzhi .
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, :5088-+