Repeatability of the measurement of aortic pulse wave velocity (aPWV) in the clinical assessment of arterial stiffness in community-dwelling older patients using the Vicorder® device

被引:21
作者
McGreevy, Cora [1 ]
Barry, Miriam [1 ]
Bennett, Kathleen [2 ]
Williams, David [1 ]
机构
[1] Beaumont Hosp, RCSI, Dept Geriatr & Stroke Med, Dublin 9, Ireland
[2] St James Hosp, Trinity Ctr Hlth Sci, Dept Pharmacol & Therapeut, Dublin, Ireland
关键词
Pulse wave velocity; reproducibility; Vicorder (R); arterial stiffness; elderly; AUGMENTATION INDEX; SPHYGMOCOR; MORTALITY; DISEASE;
D O I
10.3109/00365513.2013.770162
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. Pulse wave velocity measured using the Vicorder (R) apparatus is a non-invasive indicator of arterial stiffness. The objective of this study was to assess its repeatability in older patients when used by medical professionals with limited experience of the technique. Methods. Aortic pulse wave velocity (aPWV) was measured using the Vicorder (R) system four times using the conventional pathway and twice by adapted pathway in 25 consecutive ambulant patients (15 male) after they rested supine for 15 min. A nurse and a doctor independently and alternately measured PWV using the same equipment and were blinded to their colleague's PWV readings. 'Within' and 'between' observer differences were assessed using intra-class correlation coefficients (rI) and 95% limits of agreement (95% LoA) derived from Bland-Altman plots. Results. Mean age was 79.8 (mean blood pressure [BP] = 133/69, Mean heart rate [HR]: 70.9). Mean PWV was 11.73 (Standard Deviation [SD] 2.6-3.6). 'Between' and 'within' observer repeatability was high, with rIs ranging from 0.8-0.93. The repeatability index for 'between' nurse and doctor measures was slightly lower (rI = 0.88) when an adapted cuff measure was used in assessing PWV compared to conventional cuff measures (rI = 0.93). Mean PWV readings 'between' observers differed by only 0.094 (95% CI 0.24 to 1.59). Conclusion. When undertaken by operators with limited previous technical experience, both 'within' and 'between' observer repeatability of PWV measurement was high. This method has the potential to be included in the clinical assessment of arterial stiffness in older ambulant patients.
引用
收藏
页码:269 / 273
页数:5
相关论文
共 11 条
  • [1] Aortic pulse wave velocity index and mortality in end-stage renal disease
    Blacher, J
    Safar, ME
    Guerin, AP
    Pannier, B
    Marchais, SJ
    London, GM
    [J]. KIDNEY INTERNATIONAL, 2003, 63 (05) : 1852 - 1860
  • [2] Repeatability of the measurement of augmentation index in the clinical assessment of arterial stiffness using radial applanation tonometry
    Crilly, M.
    Coch, C.
    Clark, H.
    Bruce, M.
    Williams, D.
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2007, 67 (04) : 413 - 422
  • [3] Validity and repeatability of the Vicorder apparatus: a comparison with the SphygmoCor device
    Hickson, Stacey S.
    Butlin, Mark
    Broad, Jeremy
    Avolio, Alberto P.
    Wilkinson, Ian B.
    McEniery, Carmel M.
    [J]. HYPERTENSION RESEARCH, 2009, 32 (12) : 1079 - 1085
  • [4] PRESSURE WAVE-PROPAGATION IN A MULTIBRANCHED MODEL OF THE HUMAN UPPER-LIMB
    KARAMANOGLU, M
    GALLAGHER, DE
    AVOLIO, AP
    OROURKE, MF
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 269 (04): : H1363 - H1369
  • [5] Validating a New Oscillometric Device for Aortic Pulse Wave Velocity Measurements in Children and Adolescents
    Kracht, Daniela
    Shroff, Rukshana
    Baig, Sabrina
    Doyon, Anke
    Jacobi, Christoph
    Zeller, Rene
    Querfeld, Uwe
    Schaefer, Franz
    Wuehl, Elke
    Schmidt, Bernhard M. W.
    Melk, Anette
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2011, 24 (12) : 1294 - 1299
  • [6] Aortic stiffness is an independent predictor of all-cause and cardiovascular mortality in hypertensive patients
    Laurent, S
    Boutouyrie, P
    Asmar, R
    Gautier, I
    Laloux, B
    Guize, L
    Ducimetiere, P
    Benetos, A
    [J]. HYPERTENSION, 2001, 37 (05) : 1236 - 1241
  • [7] Expert consensus document on arterial stiffness: methodological issues and clinical applications
    Laurent, Stephane
    Cockcroft, John
    Van Bortel, Luc
    Boutouyrie, Pierre
    Giannattasio, Cristina
    Hayoz, Daniel
    Pannier, Bruno
    Vlachopoulos, Charalambos
    Wilkinson, Ian
    Struijker-Boudier, Harry
    [J]. EUROPEAN HEART JOURNAL, 2006, 27 (21) : 2588 - 2605
  • [8] The Vicorder device compared with SphygmoCor in the assessment of carotid-femoral pulse wave velocity in patients with peripheral arterial disease
    Shahin, Yousef
    Barakat, Hashem
    Barnes, Rachel
    Chetter, Ian
    [J]. HYPERTENSION RESEARCH, 2013, 36 (03) : 208 - 212
  • [9] Carotid-femoral pulse wave velocity: Impact of different arterial path length measurements
    Sugawara J.
    Hayashi K.
    Yokoi T.
    Tanaka H.
    [J]. Artery Research, 2010, 4 (1) : 27 - 31
  • [10] Comparison of two instruments measuring carotid-femoral pulse wave velocity: Vicorder versus SphygmoCor
    van Leeuwen-Segarceanu, Elena M.
    Tromp, Wilma F.
    Bos, Willem-Jan W.
    Vogels, Oscar J. M.
    Groothoff, Jaap W.
    van der Lee, Johanna H.
    [J]. JOURNAL OF HYPERTENSION, 2010, 28 (08) : 1687 - 1691