A reservoir-type oxygen sensor with 2 x 3 array for measuring cellular respiration levels

被引:12
作者
Park, Jungil [1 ]
Nam, Hyunwook [1 ]
Ahn, Sun Young [2 ]
Pak, Youngmi Kim [2 ]
Pak, James Jungho [1 ]
机构
[1] Korea Univ, Sch Elect Engn, Seoul 136713, South Korea
[2] Kyung Hee Univ, Coll Med, Dept Physiol, Seoul 130701, South Korea
基金
新加坡国家研究基金会;
关键词
Electrochemistry; Array type; Oxygen sensor; Cellular respiration; Oxygen consumption rate; PO(2); PCO(2); MICROELECTRODE; CONSUMPTION; MICROSENSOR; ELECTRODE; MODULE; PH;
D O I
10.1016/j.snb.2012.09.037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes a 2 x 3 reservoir-type oxygen sensor array integrated with Clark-type sensors, heaters, and temperature sensors on a glass substrate. The measured 90% response time of the fabricated sensors from an air-saturated state to a zero-oxygen state was approximately 9.35 s. These oxygen sensors showed good repeatability during 6 consecutive measurements at 300-s intervals, with an output of 108.2 +/- 3.15 nA (mean +/- standard deviation) for the air-saturated state and 1.33 +/- 0.14 nA for the zero-oxygen state. In addition, these sensors showed good linearity, with a correlation coefficient of 0.995 and a sensitivity averaging 2.04 A/cm(2) M. A LabVIEW-based measurement system was developed for simultaneously measuring the multiple cellular respiration levels from this oxygen sensor array using a voltage supply, an amplifying module, a DMM (digital multimeter), a switch module, and a temperature-sensing module. The measured cellular respiration ratio of the L6 (rat skeletal muscle cells) between the uncoupled oxygen consumption rate and the coupled oxygen consumption rate for this oxygen sensor array was 2.06, which is similar to the 2.23 value obtained from a commercially available system. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:913 / 920
页数:8
相关论文
共 30 条
[1]  
[Anonymous], 2001, ELECTROCHEMICAL METH
[2]   Thickness of electrical double layer. Effect of ion size [J].
Bohinc, K ;
Kralj-Iglic, V ;
Iglic, A .
ELECTROCHIMICA ACTA, 2001, 46 (19) :3033-3040
[3]   Chip-based time-continuous monitoring of toxic effects on stem cell differentiation [J].
Cho, Sungbo ;
Gorjup, Erwin ;
Thielecke, Hagen .
ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2009, 191 (01) :145-152
[4]  
Clark LC, 1959, United States Patent, Patent No. 2913386
[5]   Quantitative Microplate-Based Respirometry with Correction for Oxygen Diffusion [J].
Gerencser, Akos A. ;
Neilson, Andy ;
Choi, Sung W. ;
Edman, Ursula ;
Yadava, Nagendra ;
Oh, Richard J. ;
Ferrick, David A. ;
Nicholls, David G. ;
Brand, Martin D. .
ANALYTICAL CHEMISTRY, 2009, 81 (16) :6868-6878
[6]   The dynamic response of optical oxygen sensors and voltammetric electrodes to temporal changes in dissolved oxygen concentrations [J].
Glazer, BT ;
Marsh, AG ;
Stierhoff, K ;
Luther, GW .
ANALYTICA CHIMICA ACTA, 2004, 518 (1-2) :93-100
[7]  
GNAIGER E, 2001, MITOCHONDRIAL PHYSL, V6, P1
[8]   Detection of the osteogenic differentiation of mesenchymal stem cells in 2D and 3D cultures by electrochemical impedance spectroscopy [J].
Hildebrandt, Cornelia ;
Bueth, Heiko ;
Cho, Sungbo ;
Impidjati ;
Thielecke, Hagen .
JOURNAL OF BIOTECHNOLOGY, 2010, 148 (01) :83-90
[9]  
Hitchman M.L., 1978, MEASUREMENT DISSOLVE
[10]   THIN-FILM CLARK-TYPE OXYGEN SENSOR-BASED ON NOVEL POLYMER MEMBRANE SYSTEMS FOR IN-VIVO AND BIOSENSOR APPLICATIONS [J].
JOBST, G ;
URBAN, G ;
JACHIMOWICZ, A ;
KOHL, F ;
TILADO, O ;
LETTENBICHLER, I ;
NAUER, G .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (3-4) :123-128