Auscultatory versus oscillometric measurement of blood pressure in octogenarians

被引:3
作者
Rosholm, Jens-Ulrik [1 ]
Arnspang, Sidsel [1 ]
Matzen, Lars [1 ]
Jacobsen, Ib A. [2 ]
机构
[1] Odense Univ Hosp, Dept Geriatr, DK-5000 Odense, Denmark
[2] Odense Univ Hosp, Dept Endocrinol, Hypertens Clin, DK-5000 Odense, Denmark
关键词
auscultatory blood pressure; blood pressure determination; octogenarians; oscillometric blood pressure; MEASURING DEVICES; EUROPEAN-SOCIETY; HYPERTENSION; VALIDATION; PROTOCOL;
D O I
10.3109/08037051.2012.680751
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background. Auscultatory measurement using a sphygmomanometer has been the predominant method for clinical estimation of blood pressure, but it is now rapidly being replaced by oscillometric measurement. Objective. To compare blood pressure by auscultatory and oscillometric measurements in patients >= 80 years. Method. 100 patients had blood pressure measured by auscultation with a sphygmomanometer and by an electronic device using the oscillometric method. For each patient the mean of two blood pressures with each method measured within 15 min were compared. Results. The mean age of participants was 85.8 years; 55.8% were women. The correlation coefficient for systolic blood pressure was 0.88 and for diastolic 0.79. Differences between auscultatory and oscillometric values were less than 10 mmHg in 70.6% of systolic blood pressures and in 83.2% for diastolic. Arrhythmia and hypertension did not influence the results, and there was no correlation between the magnitude of the differences and the level of blood pressure. Conclusion. Agreement between oscillometric and auscultatory measurements of blood pressure in octogenarians was found to be less than required by validation protocols. However, semi-automatic equipment, which is observer-independent, may be used even in the very elderly, particularly if multiple readings are performed.
引用
收藏
页码:269 / 272
页数:4
相关论文
共 14 条
[1]  
[Anonymous], MERC SPHYGM HEALTHC
[2]   ABC of hypertension - Blood pressure measurement Part II - Conventional sphygmomanometry: technique of auscultatory blood pressure measurement [J].
Beevers, G ;
Lip, GYH ;
O'Brien, E .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7293) :1043-1047
[3]   Selected major risk factors and global and regional burden of disease [J].
Ezzati, M ;
Lopez, AD ;
Rodgers, A ;
Vander Hoorn, S ;
Murray, CJL .
LANCET, 2002, 360 (9343) :1347-1360
[4]   DISCREPANCIES IN RECORDING SYSTOLIC BLOOD-PRESSURE OF ELDERLY PERSONS BY AMBULATORY BLOOD-PRESSURE MONITOR [J].
MILLER, ST ;
ELAM, JT ;
GRANEY, MJ ;
APPLEGATE, WB .
AMERICAN JOURNAL OF HYPERTENSION, 1992, 5 (01) :16-21
[5]   European Society of hypertension recommendations for conventional, ambulatory and home blood pressure measurement [J].
O'Brien, E ;
Asmar, R ;
Beilin, L ;
Imai, Y ;
Mallion, JM ;
Mancia, G ;
Mengden, T ;
Myers, M ;
Padfield, P ;
Palatini, P ;
Parati, G ;
Pickering, T ;
Redon, J ;
Staessen, J ;
Stergiou, G ;
Verdecchia, P .
JOURNAL OF HYPERTENSION, 2003, 21 (05) :821-848
[6]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[7]  
O'Brien E, 1993, CARDIOL ELDER, V1, P464
[8]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[9]   Validation of A&D UA-767 device for the self-measurement of blood pressure [J].
Rogoza, AN ;
Pavlova, TS ;
Sergeeva, MV .
BLOOD PRESSURE MONITORING, 2000, 5 (04) :227-231
[10]   A CONSENSUS VIEW ON THE TECHNIQUE OF AMBULATORY BLOOD-PRESSURE MONITORING [J].
STAESSEN, JA ;
FAGARD, R ;
THIJS, L ;
AMERY, A ;
CELLS, H ;
VANHOOF, R ;
ASMAR, R ;
BUNTINX, F ;
DECORT, P ;
CLEMENT, D ;
DUPREZ, D ;
COATS, AJS ;
CONWAY, J ;
DEGAUTE, JP ;
DEPLAEN, JF ;
DELEEUW, PW ;
DEVEREUX, RB ;
PICKERING, TG ;
EFSTRATOPOULOS, AD ;
ELLIOTT, HL ;
GELLERT, J ;
GOSSE, P ;
IMAI, Y ;
IMHOLZ, BPM ;
VANMONTFRANS, GA ;
WESSELING, KH ;
KINGSWOOD, C ;
LIU, LS ;
MALLION, JM ;
OBRIEN, ET ;
OCONPUJADAS, K ;
OTSUKA, K ;
PALATINI, P ;
PARATI, G ;
PERLOFF, D ;
PORCELLATI, C ;
VERDECCHIA, P ;
RAFTERY, EB ;
REDON, J ;
ROSENFELD, J ;
RYNKIEWICZ, A ;
SALVETTI, A ;
TUOMILEHTO, J ;
VANHANEN, H ;
WAEBER, B ;
WHITE, WB ;
WEBSTER, J .
HYPERTENSION, 1995, 26 (06) :912-918